Chest ACCP Career Connection
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Guest Access | Sign In via User Name/Password
This Article
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Standaert, T. A.
Right arrow Articles by Ramsey, B. W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Standaert, T. A.
Right arrow Articles by Ramsey, B. W.
(Chest. 1998;114:577-586.)
© 1998 American College of Chest Physicians

Effects of Repetitive Use and Cleaning Techniques of Disposable Jet Nebulizers on Aerosol Generation

Thomas A. Standaert PhD1; Gregory L. Morlin BS2; Judy Williams-Warren RN1; Pamella Joy RN, ARNP1; Margaret S. Pepe PhD3; Allan Weber BS2; and Bonnie W. Ramsey MD1

1 From the Cystic Fibrosis Research Center, Children's Hospital and Medical Center, Seattle
2 From the Department of Molecular Microbiology and Immunology, Columbia School of Medicine, University of Missouri, Columbia
3 From the Department of Biostatistics, Fred Hutchinson Cancer Research Center, Seattle

Study objective: Patients with cystic fibrosis use disposable jet nebulizers for the self-administration of antibiotics, DNase, and bronchodilators several times per day. Most patients elect to reuse their disposable nebulizers. The purpose of this study was to determine if significant changes in particle size distribution or output (mL/min) occurred with reuse.

Design: In vitro studies were performed using four disposable models and one durable jet nebulizer for up to 100 runs; measurements of particle size and output were obtained at 10 run intervals, using saline solution alone, tobramycin, gentamicin, or a mixture of albuterol and cromolyn. Particle size determinations were made with a laser diffraction analyzer.

Results: There was no significant difference between the baseline performance of the four disposable models and the durable Pari LC, when measuring particle size distribution of the aerosol; the Pari LC had an output rate two to three times higher than the four disposable models. For each of the four solutes tested, there was no clinically significant change in performance for up to 100 cycles, when the nebulizers were properly cleaned between uses. Unwashed units containing tobramycin started to fail by 40 runs.

Conclusions: When properly maintained, there was no trend of deterioration of performance with repeated use of disposable nebulizers. Microbial contamination was not addressed in this study and must be considered prior to recommendations for the reuse of disposable nebulizers.

Key Words: aerosols • aminoglycosides • cleaning • efficiency • nebulizers • tobramycin

Submitted on August 12, 1997
Accepted on January 8, 1998




This article has been cited by other articles:


Home page
Am J Health Syst PharmHome page
M. B. Shirk, K. R. Donahue, and J. Shirvani
Unlabeled uses of nebulized medications.
Am. J. Health Syst. Pharm., September 15, 2006; 63(18): 1704 - 1716.