Chest ACCP Education Calendar
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Guest Access | Sign In via User Name/Password
This Article
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Berlinski, A.
Right arrow Articles by Waldrep, J. C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Berlinski, A.
Right arrow Articles by Waldrep, J. C.
(Chest. 1998;114:847-853.)
© 1998 American College of Chest Physicians

Four Hours of Continuous Albuterol Nebulization

Ariel Berlinski MD1 and J. Clifford Waldrep PhD2

1 From the Department of Pediatrics, Pulmonology Section, Baylor College of Medicine, Houston, TX.
2 From the Department of Molecular Physiology and Biophysics, Baylor College of Medicine, Houston, TX.

J. Clifford Waldrep, PhD, Department of Molecular Physiology and Biophysics, Baylor College of Medicine, One Baylor Plaza, Houston, TX 77030

Background and objectives: Continuous albuterol nebulization (CAN) is a therapeutic modality available to treat status asthmaticus. Currently, CAN may be administered using a large-volume nebulizer (LVN) or a small-volume nebulizer attached to an infusion pump or refilled as needed. Few data are available regarding the reproducibility of aerosol characteristics during CAN. In this study, we determined the aerodynamic profile, drug output (DO), DO in respirable range (RD), solution output (SO), and changes in reservoir's albuterol concentration (AR) hourly during 4 hours of CAN.

Design: A modified Puritan-Bennett 1600 jet nebulizer was tested with a large reservoir (LR; 250 mL), medium reservoir (MR; 45 mL), and small reservoir with infusion pump (SRP; 18 mL). We used 100minus, 40minus, and 4-mL initial fill volumes (with 10-mL/h infusion for SRP) of 1 mg/mL albuterol solution for the LR, MR, and SRP, respectively. Particle size distribution and DO consistency were determined by impaction and spectrophotometric analysis (275 nm). We also determined albuterol mass output. The SO was determined by gravimetric technique.

Results: The PBsj produced a heterodisperse aerosol with a median mass aerodynamic diameter range of 1.8 to 2.2 µm. DO and RD paralleled SO. The LR had the highest SO, DO, and RD (8.03±2.36 vs 5.73±2.48 and 5.85±0.51 mg/h for MR and SRP, respectively). The AR showed no statistically significant changes.

Conclusions: The PBsj demonstrated consistent and adequate aerosol production during 4 hours of CAN. These bench data support the widespread use of a LVN for CAN.

Key Words: aerosol • albuterol • continuous nebulization • drug aerosol therapy • inhalation devices • nebulizers • status asthmaticus

Submitted on November 25, 1997
Accepted on March 5, 1998







HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1998 by the American College of Chest Physicians.