Study objective: To determine the feasibility of using real-time bronchoscope position technology coupled with previously acquired three-dimensional CT data to enhance transbronchial needle aspiration (TBNA).
Design: Eight swine were given percutaneously created target lesions for TBNA. A miniature position sensor was placed at the tip of a bronchoscope, and real-time position information during bronchoscopy was presented on a monitor simultaneously displaying previously acquired three-dimensional CT data.
Interventions: TBNA of target lesions and submucosal ink-spot injection of computer-generated targets.
Measurements and results: TBNA specimens revealed successful aspiration of target material. Distances between ink marks made at computer-generated tracheal targets varied, on average (±SD), 4.2 mm ± 2.6 mm from predetermined computer-distance coordinates.
Conclusion: Real-time bronchoscope position technology coupled with previously acquired CT images may aid with TBNA of nonvisible extrabronchial lesions.
Key Words: animal model biopsy bronchoscopy CT bronchoscopy image-guided surgery transbronchial needle aspiration virtual bronchoscopy
Submitted on January 13, 1998
Accepted on June 22, 1998
This article has been cited by other articles:
![]() |
Y. Schwarz, J. Greif, H. D. Becker, A. Ernst, and A. Mehta Real-Time Electromagnetic Navigation Bronchoscopy to Peripheral Lung Lesions Using Overlaid CT Images: The First Human Study. Chest, April 1, 2006; 129(4): 988 - 994. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Hautmann, A. Schneider, T. Pinkau, F. Peltz, and H. Feussner Electromagnetic Catheter Navigation During Bronchoscopy: Validation of a Novel Method by Conventional Fluoroscopy Chest, July 1, 2005; 128(1): 382 - 387. [Abstract] [Full Text] [PDF] |
||||
![]() |
L.-H. Hsu, C.-C. Liu, and J.-S. Ko Education and Experience Improve the Performance of Transbronchial Needle Aspiration: A Learning Curve at a Cancer Center Chest, February 1, 2004; 125(2): 532 - 540. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. S. White, E. A. Weiner, P. Patel, and E. J. Britt Transbronchial Needle Aspiration : Guidance With CT Fluoroscopy Chest, December 1, 2000; 118(6): 1630 - 1638. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. B. Solomon, P. White Jr., C. M. Wiener, J. B. Orens, and K. P. Wang Three-dimensional CT-Guided Bronchoscopy With a Real-Time Electromagnetic Position Sensor : A Comparison of Two Image Registration Methods Chest, December 1, 2000; 118(6): 1783 - 1787. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. M. HARROW, W. ABI-SALEH, J. BLUM, T. HARKIN, S. GASPARINI, D. J. ADDRIZZO-HARRIS, A. C. ARROLIGA, G. WIGHT, and A. C. MEHTA The Utility of Transbronchial Needle Aspiration in the Staging of Bronchogenic Carcinoma Am. J. Respir. Crit. Care Med., February 1, 2000; 161(2): 601 - 607. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |