Chest Email Content Delivery
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Guest Access | Sign In via User Name/Password
This Article
Right arrow Full Text Free
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (8)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Scherer, T. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Scherer, T. A.
(Chest. 2000;117:1449-1454.)
© 2000 American College of Chest Physicians

Nd-YAG Laser Ignition of Silicone Endobronchial Stents*

Thomas A. Scherer, MD, FCCP{dagger}

* From the Pulmonary Division, Department of Internal Medicine, Triemli Hospital, Birmensdorferstrasse 497, 8063 Zurich, Switzerland. {dagger} Currently at LungenZentrum Hirslanden, Witellikerstrasse 36, 8008 Zuerich.

Correspondence to: Thomas A. Scherer, MD, FCCP, LungenZentrum Hirslanden, Witellikerstrasse 36, 8008 Zurich, Switzerland, e-mail: thsche{at}swissonline.ch

Study objectives: To test the incendiary characteristics of various silicone endobronchial stents under the impact of the Nd-YAG laser.

Design: In vitro study in the laser laboratory of a university-affiliated city hospital.

Setting: The experiments were performed in a reaction chamber under controlled oxygen concentrations. The radiolucent and radiopaque Dumon silicone stent (Novatech; Aubagne, France) and the tracheal part of the Dynamic stent (Ruesch AG; Kernen, Germany) were tested. The Dumon stents were either clean, covered with a thin layer of blood, or mounted on fresh pig tracheal wall. The laser was aimed on them perpendicularly from distances of 1.0 cm and 0.5 cm.

Interventions: Minimal oxygen concentration to allow ignition and impact time for power outputs (POs) between 10 W and 80 W were determined.

Measurements and results: The lowest oxygen concentration that allowed ignition of some stents was 40%. The clean radiolucent stent could not be ignited with up to 100% ambient oxygen concentration. Radiopacity, presence of blood, tracheal wall, and metal, as well as higher PO and shorter distance of the laser probe decreased impact time to ignition. The radiopaque blood-covered stent was most easily ignited. For this stent, at a PO of 40 W, impact time to ignition was 1.5 ± 0.2 s, and at 30 W was 2.6 ± 0.3 s.

Conclusions: At ambient oxygen concentrations >= 40%, silicone stents can catch fire. Depending on the condition of the stent, the distance of the laser probe, and PO, ignition can occur after short impact times. To prevent stent ignition, oxygen concentration should be kept < 40%. When unusual circumstances require working with higher oxygen concentrations, pulse duration needs to be limited or stent removal might be considered before firing the laser.

Key Words: adverse effects • burns • endoscopy • laser surgery • oxygen • stents




This article has been cited by other articles:


Home page
Br J AnaesthHome page
I. D. Conacher
Anaesthesia and tracheobronchial stenting for central airway obstruction in adults
Br. J. Anaesth., March 1, 2003; 90(3): 367 - 374.
[Abstract] [Full Text] [PDF]


Home page
ChestHome page
J. J. Dalupang, T. G. Shanks, and H. G. Colt
Nd-YAG Laser Damage to Metal and Silicone Endobronchial Stents : Delineation of Margins of Safety Using an In Vitro Experimental Model
Chest, September 1, 2001; 120(3): 934 - 940.