Guest Access | Sign In via User Name/Password |
|||||||||
* From the Istituto di Fisiopatologia Respiratoria del Consiglio Nazionale delle Ricerche, Palermo, Italy.
Correspondence to: Oreste Marrone, MD, Istituto di Fisiopatologia Respiratoria del C. N. R., Via Ugo La Malfa, 153, 90146 Palermo, Italy; e-mail: marrone{at}ifr.pa.cnr.it
Study objectives: To evaluate the relationship between sleep structure and continuous positive airway pressure (CPAP) delivered by an automatic CPAP (auto-CPAP) machine in patients with obstructive sleep apnea syndrome (OSAS).
Design: Nocturnal polysomnography was performed during CPAP administration by an auto-CPAP machine (Autoset Clinical 1; ResMed; Sydney, Australia).
Setting: Sleep-disorders center in a research institute.
Patients: Fifteen subjects with newly diagnosed OSAS deserving home CPAP treatment.
Measurements and results: During the night, in most cases, the lowest CPAP level was recorded during a prolonged nonrapid eye movement (NREM) sleep period uninterrupted by arousals, whereas the highest level during wake-sleep transitions or NREM sleep fragmented by arousals. In four subjects, rapid eye movement sleep was always associated with increasing CPAP. Sleep efficiency was negatively correlated with CPAP variability, evaluated as the SD of the mean nocturnal CPAP level averaged epoch by epoch (r = 0.63, p < 0.02). Eighty-eight percent of rapid CPAP augmentations (increases by at least 2 cm H2O in
2 min) were observed during sleep-wake transitions or after arousals/awakenings (Ar/Aw); 63% of such Ar/Aw were not preceded by any detectable respiratory abnormality.
Conclusions: CPAP levels and variations during auto-CPAP application may be mainly related to sleep continuity and efficiency. The recording of a highly variable pressure during auto-CPAP administration in an unattended environment must raise the question whether the patients sleep quality was acceptable. A poor sleep quality during an autotitration night could lead to an undesirable overestimation of the CPAP level needed for use with fixed-level CPAP machines.
Key Words: continuous positive airway pressure obstructive sleep apnea syndrome titration treatment
This article has been cited by other articles:
![]() |
D. Rodenstein Determination of Therapeutic Continuous Positive Airway Pressure for Obstructive Sleep Apnea Using Automatic Titration: Promises Not Fulfilled Chest, March 1, 2008; 133(3): 595 - 597. [Full Text] [PDF] |
||||
![]() |
O. Marrone, A. Salvaggio, S. Romano, and G. Insalaco Automatic Titration and Calculation by Predictive Equations for the Determination of Therapeutic Continuous Positive Airway Pressure for Obstructive Sleep Apnea Chest, March 1, 2008; 133(3): 670 - 676. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Rigau, J. M. Montserrat, H. Wohrle, D. Plattner, M. Schwaibold, D. Navajas, and R. Farre Bench model to simulate upper airway obstruction for analyzing automatic continuous positive airway pressure devices. Chest, August 1, 2006; 130(2): 350 - 361. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. S. Doherty, P. Nolan, and W. T. McNicholas Effects of topical anesthesia on upper airway resistance during wake-sleep transitions J Appl Physiol, August 1, 2005; 99(2): 549 - 555. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Noseda, C. Kempenaers, M. Kerkhofs, S. Braun, P. Linkowski, and E. Jann Constant vs Auto-Continuous Positive Airway Pressure in Patients With Sleep Apnea Hypopnea Syndrome and a High Variability in Pressure Requirement Chest, July 1, 2004; 126(1): 31 - 37. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Senn, T. Brack, F. Matthews, E. W. Russi, and K. E. Bloch Randomized Short-term Trial of Two AutoCPAP Devices versus Fixed Continuous Positive Airway Pressure for the Treatment of Sleep Apnea Am. J. Respir. Crit. Care Med., December 15, 2003; 168(12): 1506 - 1511. [Abstract] [Full Text] [PDF] |
||||