Chest ACCP Education Calendar
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Guest Access | Sign In via User Name/Password
This Article
Right arrow Full Text Free
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (10)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Vitacca, M.
Right arrow Articles by Ambrosino, N.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Vitacca, M.
Right arrow Articles by Ambrosino, N.
(Chest. 2002;122:2105-2114.)
© 2002 American College of Chest Physicians

Comparison of Five Bilevel Pressure Ventilators in Patients With Chronic Ventilatory Failure*

A Physiologic Study

Michele Vitacca, MD; Luca Barbano, MD; Silvestro D’Anna, MD; Roberto Porta, MD; Luca Bianchi, MD and Nicolino Ambrosino, MD, FCCP

* From the Fondazione Salvatore Maugeri IRCCS, Pulmonary Division, Scientific Institute, Gussago, Italy.

Correspondence to: Nicolino Ambrosino, MD, FCCP, Fondazione S. Maugeri Pulmonary Division Istituto Scientifico di Gussago I-25064 Gussago (BS), Italy; e-mail: nambrosino{at}fsm.it

Objective: To compare patient-ventilator interaction and comfort in patients with chronic ventilatory failure (CVF) who are undergoing noninvasive positive-pressure ventilation with five different commercial bilevel pressure home ventilators. Also, we wanted to evaluate the short-term effects of the five ventilators on physiologic variables, namely, breathing patterns and inspiratory muscles.

Design: Randomized, controlled physiologic study.

Setting: Pulmonary division of a rehabilitation institution.

Patients: Twenty-eight patients with CVF due to COPD (17 patients) and restrictive chest wall diseases (11 patients).

Measurements: Sensation of comfort, breathing patterns and minute ventilation (E), respiratory muscles and mechanics, and patient-ventilator interaction during both unassisted and assisted ventilation with the five ventilators applied randomly.

Results: The five ventilators showed different flow and pressure waveforms. The level of comfort was somehow different among the studied ventilators. When compared to unassisted ventilation, all ventilators induced a significant increase in E (p < 0.01) without any significant difference among ventilators. Use of the five ventilators resulted in significant differences in peak airway opening pressure (Pao,peak) but not in mean airway opening pressure computed over a period of 1 min (PTPao,min), and in a duty cycle. Ineffective efforts (IEs) were similar among the studied ventilators. In comparison with unassisted ventilation, all ventilators induced significant reductions in inspiratory muscle effort (p < 0.001). No significant relationship was found between level of comfort and PTPao,min, Pao,peak, or the number of IEs.

Conclusions: In stable, awake patients with CVF, all of the studied ventilators were well-tolerated, although with a great intersubject variability in comfort, and performed well in terms of improvement in E and inspiratory muscle unloading, thus fulfilling the aims of mechanical ventilation. This effect was obtained with similar levels of PTPao,min, despite the fact that Pao,peak was different among some ventilators. The number of IEs was similar among the studied ventilators.

Key Words: breathing pattern • COPD • hypercapnia • noninvasive mechanical ventilation • respiratory failure • respiratory muscles • restrictive chest wall disease




This article has been cited by other articles:


Home page
Eur Respir JHome page
R. Farre, S. J. Lloyd-Owen, N. Ambrosino, G. Donaldson, J. Escarrabill, B. Fauroux, D. Robert, B. Schoenhofer, A. Simonds, and J. A. Wedzicha
Quality control of equipment in home mechanical ventilation: a European survey
Eur. Respir. J., July 1, 2005; 26(1): 86 - 94.
[Abstract] [Full Text] [PDF]


Home page
ChestHome page
A. Battisti, D. Tassaux, J.-P. Janssens, J.-B. Michotte, S. Jaber, and P. Jolliet
Performance Characteristics of 10 Home Mechanical Ventilators in Pressure-Support Mode: A Comparative Bench Study
Chest, May 1, 2005; 127(5): 1784 - 1792.
[Abstract] [Full Text] [PDF]


Home page
ChestHome page
M. Vitacca, L. Bianchi, E. Zanotti, A. Vianello, L. Barbano, R. Porta, and E. Clini
Assessment of Physiologic Variables and Subjective Comfort Under Different Levels of Pressure Support Ventilation
Chest, September 1, 2004; 126(3): 851 - 859.
[Abstract] [Full Text] [PDF]