Chest ACCP Education Calendar
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Guest Access | Sign In via User Name/Password
This Article
Right arrow Full Text Free
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (16)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Noppen, M.
Right arrow Articles by Vinken, W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Noppen, M.
Right arrow Articles by Vinken, W.
(Chest. 2005;127:482-487.)
© 2005 American College of Chest Physicians

Removal of Covered Self-Expandable Metallic Airway Stents in Benign Disorders*

Indications, Technique, and Outcomes

Marc Noppen, MD, PhD, FCCP; Grigoris Stratakos, MD; Jan D’Haese, MD; Marc Meysman, MD, FCCP and Walter Vinken, MD, PhD

* From Interventional Endoscopy Clinic, Respiratory Division, University Hospital AZ-VUB, Free University of Brussels, Brussels, Belgium.

Correspondence to: M. Noppen MD, PhD, Head, Interventional Endoscopy Clinic, Respiratory Division, University Hospital AZ-VUB, Laarbeeklaan 101, B 1090 Brussels, Belgium; e-mail: marc.noppen{at}az.vub.ac.be

Study objectives: To describe the technique and outcome of removal of self-expandable metallic airway stents (SEMAS) in a series of patients who underwent stenting for benign airway stenoses.

Design and setting: A retrospective cohort analysis of patients with benign airway disorders treated with SEMAS between 1997 and 2003, who presented with an indication for stent removal in a tertiary-care hospital, and referral interventional pulmonology clinic.

Patients and methods: During the study period, 49 SEMAS were inserted in 39 patients for treatment of benign airway disorders. Ten of these 39 patients (25.6%), bearing 12 covered stents, presented with an indication for stent removal. Data of these cases were extracted from electronic files kept in our institution archive.

Results: Indications for stent removal included excessive or recurrent granuloma formation (five cases), recurrence of stenosis after stent failure (one case), stent fracture (two cases), and accomplishment of treatment (two cases). In all of these cases, covered versions of SEMAS had been placed either in the trachea or in a main bronchus. In contrast to many previous reports, these stents proved to be absolutely retrievable even if some difficulties were encountered. In all cases, however, removal was successful without major complications. All patients resumed normal spontaneous ventilation postoperatively, and follow-up was uneventful. Average duration of stenting before removal was 16.2 ± 17.5 months (± SD) [range, 1 to 60 months].

Conclusions: We conclude that although placement of SEMAS is assumed to be permanent in patients with benign airway disorders, an indication for stent removal is often observed (25.6% in our series). The covered SEMAS can be effectively and safely removed if needed without major sequelae. Nevertheless, new technical improvements in metallic stent design and materials may help reinforce the concept of a retrievable metallic airway stent, which may offer significant clinical advantages.

Key Words: benign tracheobronchial strictures • central airways obstruction • interventional pulmonology • rigid bronchoscopy • self-expandable stents




This article has been cited by other articles:


Home page
Am. J. Roentgenol.Home page
J. H. Kim, J. H. Shin, H.-Y. Song, T. S. Shim, C. J. Yoon, and G.-Y. Ko
Benign Tracheobronchial Strictures: Long-Term Results and Factors Affecting Airway Patency After Temporary Stent Placement
Am. J. Roentgenol., April 1, 2007; 188(4): 1033 - 1038.
[Abstract] [Full Text] [PDF]


Home page
Arch Otolaryngol Head Neck SurgHome page
A. M. Rampey, G. A. Silvestri, and M. B. Gillespie
Combined Endoscopic and Open Approach to the Removal of Expandable Metallic Tracheal Stents
Arch Otolaryngol Head Neck Surg, January 1, 2007; 133(1): 37 - 41.
[Abstract] [Full Text] [PDF]


Home page
Journal of the American Animal Hospital AssociationHome page
M. Ouellet, M. E. Dunn, B. Lussier, N. Chailleux, and P. Helie
Noninvasive correction of a fractured endoluminal nitinol tracheal stent in a dog.
J. Am. Anim. Hosp. Assoc., November 1, 2006; 42(6): 467 - 471.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 2005 by the American College of Chest Physicians.