Chest ACCP Member Benefits
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Guest Access | Sign In via User Name/Password
This Article
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Stawicki, J.
Right arrow Articles by Josephson, M.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Stawicki, J.
Right arrow Articles by Josephson, M.

Chest, Vol 76, 193-197, Copyright © 1979 by American College of Chest Physicians


ARTICLES

Multiple cardiac output measurements in man. Evaluation of a new closed- system thermodilution method

JJ Stawicki, FD Holford, EL Michelson and ME Josephson

Cardiac output estimates by a new closed-system automatic injection thermodilution method (COI-TD) were compared serially with the direct Fick technique (COFICK) and the standard open-system manual injection thermodilution method (COTD). Comparison with cardiac outputs determined simultaneously by the direct Fick technique in 100 measurements involving ten patients showed close agreement with the new closed system method using both 25 degrees C and 3 degrees C injectates. The cardiac output range was between 1.9 to 11.6 L/min. The open-system manual injection thermodilution method under identical conditions produced a wide scatter of measurements when compared to the direct Fick technique. Reproducibility of simultaneous COI-TD and COTD measurements was examined in quadruplicate. The reproducibility of measurements was within 1.9 percent with COI-TD and 5.9 percent with the COTD method using both 25 degrees C and 3 degrees C injectates. The COI-TD method eliminates the technical problems of recirculation, unstable indicator baseline changes, thermal continuity, and reproducibility encountered with the current COTD method. Analysis of cardiac output by the closed-system automatic injection thermodilution method provides a simple, rapid, reproducible, and highly accurate method for multiple cardiac output measurements suitable for use at the bedside.


This article has been cited by other articles:


Home page
Crit Care NurseHome page
A. Gawlinski
Measuring Cardiac Output: Intermittent Bolus Thermodilution Method
Crit. Care Nurse, October 1, 2004; 24(5): 74 - 78.
[Full Text] [PDF]


Home page
J Am Coll CardiolHome page
K. G. Lehmann and M. S. Platt
Improved accuracy and precision of thermodilution cardiac output measurement using a dual thermistor catheter system
J. Am. Coll. Cardiol., March 1, 1999; 33(3): 883 - 891.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1979 by the American College of Chest Physicians.