Guest Access | Sign In via User Name/Password |
|||||||||
Chest, Vol 88, 519-526, Copyright © 1985 by American College of Chest Physicians
ARTICLES |
JA Katz, RW Kraemer and GE Gjerde
To determine work-of breathing with continuous positive airway pressure (CPAP) delivery systems, we used a lung model to simulate spontaneous breathing. "Additional work" during tidal breathing was derived by comparing change in airway pressure with change in tidal volume. Seven demand-flow CPAP delivery systems were compared with one continuous- flow, 5-L reservoir-bag system (flow of 60 L/min to maintain positive airway pressure). It was concluded that demand-flow CPAP delivery systems vary widely in the amount of additional work required of a patient. When a lung model is used, some demand-flow systems perform as well as, or better than, a continuous-flow reservoir-bag system.
This article has been cited by other articles:
![]() |
R Agarwal, A N Aggarwal, D Gupta, and S K Jindal Non-invasive ventilation in acute cardiogenic pulmonary oedema Postgrad. Med. J., October 1, 2005; 81(960): 637 - 643. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. M. Kacmarek Ventilatory Muscle Function Seminars in Cardiothoracic and Vascular Anesthesia, September 1, 2002; 6(3): 189 - 201. [Abstract] [PDF] |
||||
![]() |
R. H. Kallet, J. A. Alonso, J. M. Luce, and M. A. Matthay Exacerbation of Acute Pulmonary Edema During Assisted Mechanical Ventilation Using a Low-Tidal Volume, Lung-Protective Ventilator Strategy* Chest, December 1, 1999; 116(6): 1826 - 1832. [Abstract] [Full Text] [PDF] |
||||