Chest ACCP Education Calendar
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Guest Access | Sign In via User Name/Password
This Article
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Olivenstein, R.
Right arrow Articles by Kreisman, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Olivenstein, R.
Right arrow Articles by Kreisman, H.

Chest, Vol 90, 392-395, Copyright © 1986 by American College of Chest Physicians


ARTICLES

A comparison of responses to albuterol delivered by two aerosol devices

R Olivenstein, N Wolkove, C Cohen, H Frank and H Kreisman

Nineteen outpatients with stable obstructive pulmonary disease (mean forced expiratory volume in one second [FEV1], 1.00 + 0.10 L) were evaluated for airway response to albuterol (salbutamol) administered by metered-dose inhaler and Bosch ultrasonic nebulizer (BUSN). Albuterol administered by metered-dose inhaler but not by nebulizer caused a significant increase in FEV1 and the mean forced expiratory flow over the middle half of the forced vital capacity (FEF25-75%) (p less than 0.02). Absolute increase from baseline of FEV1 and FEF25-75% was significantly greater for metered-dose inhaler (0.21 +/- 0.05 L; 0.32 +/- 0.13 L/sec) compared to ultrasonic nebulizer (0.07 +/- 0.03 L; 0.03 +/- 0.04 L/sec) (p less than 0.02). In 11 subjects (mean FEV1, 1.08 + 0.14 L), the placebo effect of inhalation of the diluent from the metered-dose inhaler (Freon) and the ultrasonic nebulizer (isotonic saline solution) was determined. Freon produced the mean increase of 1.5 percent, whereas the ultrasonic aerosol of isotonic saline solution resulted in a mean decrease of 8 percent in FEV1. Therefore, the inferior response to albuterol administered by ultrasonic nebulizer was at least in part due to the superimposed broncho-constriction occurring with ultrasonically administered saline solution. The metered-dose inhaler was more effective than the ultrasonic nebulizer for administration of albuterol in stable obstructive pulmonary disease, and the latter device is not recommended. A specific ultrasonic nebulizer should be prescribed only if its superiority to a metered- dose inhaler can be objectively documented.


This article has been cited by other articles:


Home page
Am J Health Syst PharmHome page
L. Hendeles, R. C. Hatton, T. J. Coons, and L. Carlson
Automatic replacement of albuterol nebulizer therapy by metered-dose inhaler and valved holding chamber
Am. J. Health Syst. Pharm., May 15, 2005; 62(10): 1053 - 1061.