Chest ACCP Member Benefits
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Guest Access | Sign In via User Name/Password
This Article
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hannhart, B
Right arrow Articles by Polu, J.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Hannhart, B
Right arrow Articles by Polu, J.

Chest, Vol 99, 842-846, Copyright © 1991 by American College of Chest Physicians


ARTICLES

Reliability of six pulse oximeters in chronic obstructive pulmonary disease

B Hannhart, H Michalski, N Delorme, G Chapparo and JM Polu
INSERM, Unite 14, CO No. 10, Vandoeuvre-les-Nancy, France.

Six pulse oximeters with finger probes were studied in three groups of 17 hypoxemic patients with COPD aged 50 to 75 years. Transcutaneous arterial oxygen saturation (SpO2) was measured with the Nellcor N101 (oximeter 1a), the Ohmeda Biox III (oximeter lb), the Nellcor N200 (oximeter 2a), the Critikon Oxyshuttle (oximeter 2b), the Radiometer Oxi100 (oximeter 3a), and the Ohmeda Biox 3700 (oximeter 3b). The SpO2 was compared with SaO2 measured in simultaneously withdrawn samples of arterial blood (Radiometer OSM2) at three 20-minute steady-state levels of FIO2 ranging from 0.21 to 0.40 (SaO2, 62 to 100 percent). The bias (mean SpO2-SaO2 difference) and the error in precision (SD of the differences) were both below 4 percent for instruments 1a and 1b and remained below 1.2 and 3 percent, respectively, for the others. A good agreement between SpO2 and SaO2, as reflected by the Bartko intraclass correlation coefficient, was observed in instruments 2a, 3a, and 3b. The individual relationships between SpO2-SaO2 differences and SaO2 appeared to be linear and parallel. With four instruments (1a, 1b, 2a, and 2b), the mean slope of this relationship was negative, showing a systematic instrumental error: the lower the SaO2, the larger the overestimation of SaO2. The scattering of the data (precision) principally reflects a subject source of error. In most instruments a technical adjustment could greatly improve instrumental errors and accuracy. The correction of the errors due to between-subject variation would require a system of calibration adjustable by the users to each individual.


This article has been cited by other articles:


Home page
CirculationHome page
P. Ponikowski, T. P. Chua, M. Piepoli, D. Ondusova, K. Webb-Peploe, D. Harrington, S. D. Anker, M. Volterrani, R. Colombo, G. Mazzuero, et al.
Augmented Peripheral Chemosensitivity as a Potential Input to Baroreflex Impairment and Autonomic Imbalance in Chronic Heart Failure
Circulation, October 21, 1997; 96(8): 2586 - 2594.
[Abstract] [Full Text]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1991 by the American College of Chest Physicians.