Chest Email Content Delivery
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Guest Access | Sign In via User Name/Password
This Article
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Snashall, P.
Right arrow Articles by Noble, M.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Snashall, P.
Right arrow Articles by Noble, M.

Chest, Vol 99, 1183-1185, Copyright © 1991 by American College of Chest Physicians


ARTICLES

Use of an impedance meter for measuring airways responsiveness to histamine

PD Snashall, S Parker, P Ten Haave, D Simmons and MI Noble
Department of Medicine, Charing Cross Hospital, London, England.

We compared spirometry (FEV1) and airway impedance (z) in the assessment of airway responsiveness to histamine. Airway impedance was measured by the oscillator technique during quiet breathing; both measurements were made twice after each increment of histamine during the challenge. Percentage change in impedance was related to percentage change in FEV1 according to: -delta z = 1.09-2.66 (delta FEV1), r = - 0.73, p less than 0.01, ie, impedance increased on average 2.7 times as much as FEV1 fell. The cut-off point for the standard test is the histamine concentration giving a 20 percent fall of FEV1 (PC20(FEV1)). The corresponding cut-off value chosen for impedance was a 30 percent increase (PC30(z)). (PC30(z)) = 0.74 (PC20(FEV1))-0.48, rs = 0.88, where rs is the Spearman rank correlation coefficient. Thus, impedance is a more sensitive index than FEV1 because a smaller dose of histamine gave a diagnostic result. Impedance is a practical alternative to FEV1, being less arduous for the patient and requiring little cooperation.


This article has been cited by other articles:


Home page
Eur Respir JHome page
E. Oostveen, D. MacLeod, H. Lorino, R. Farre, Z. Hantos, K. Desager, and F. Marchal
The forced oscillation technique in clinical practice: methodology, recommendations and future developments
Eur. Respir. J., December 1, 2003; 22(6): 1026 - 1041.
[Abstract] [Full Text] [PDF]


Home page
Eur Respir JHome page
J. Rigau, F. Burgos, C. Hernandez, J. Roca, D. Navajas, and R. Farre
Unsupervised self-testing of airway obstruction by forced oscillation at the patient's home
Eur. Respir. J., October 1, 2003; 22(4): 668 - 671.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1991 by the American College of Chest Physicians.