Abstract
Objective: To determine the accuracy with which spirometric measurements of FVC and expiratory flow rates can diagnose the presence of a restrictive impairment.
Design: The pulmonary function tests of 1,831 consecutive white adult patients who had undergone both spirometry and lung volume measurements on the same visit over a 2-year period were analyzed. The probability of restrictive pulmonary impairment, defined as a reduced total lung capacity (TLC) below the lower limit of the 95% confidence interval, was determined for each of several categoric classifications of the spirometric data, and additionally for each of several interval levels of the FVC and the FEV1/FVC ratio.
Setting: A large clinical laboratory in a university teaching hospital using quality-assured and standardized spirometry and lung volume measurement techniques according to American Thoracic Society standards. Results: Two hundred twenty-five of 1,831 patients (12.3%) had a restrictive defect. The positive predictive value of spirometry for predicting restriction was relatively low; of 470 patients with a low FVC on spirometry, only 41% had restriction confirmed on lung volume measurements. When the analysis was confined to the 264 patients with a restrictive pattern on spirometry (ie, low FVC and normal or above normal FEV1 /FVC ratio), the positive predictive value was 58%. Conversely, spirometry had a very favorable negative predictive value; only 2.4% of patients (32 of 1,361) with a normal vital capacity (VC) on spirometry had a restrictive defect by TLC measurement. The probability of a restrictive defect was directly and linearly related to the degree of reduction of FVC when the FVC was < 80% of predicted (p = 0.002). Combining the FVC and the FEV1/FVC ratio improved the predictive ability of spirometry; for all values of FVC < 80% of the predicted amount, the likelihood of restrictive disease increased as the FEV1/FVC ratio increased. Conclusions: Spirometry is very useful at excluding a restrictive defect. When the VC is within the normal range, the probability of a restrictive defect is < 3%, and unless restrictive lung disease is suspected a priori, measurement of lung volumes can be avoided. However, spirometry is not able to accurately predict lung restriction; < 60% of patients with a classical spirometric restrictive pattern had pulmonary restriction confirmed on lung volume measurements. For these patients, measurement of the TLC is needed to confirm a true restrictive defect.
Key Words: diagnosis lung volumes respiratory function tests restrictive lung disease spirometry vital capacity
Lung diseases are broadly classified as those leading to airflow obstruction, volume restriction, or a combination of obstructive and restrictive defects.1 Airflow obstruction can be diagnosed by a low FEV1/FVC ratio on spirometry. Spirometry is also often used as a screening tool to diagnose or rule out restrictive pulmonary impairment; however, its reliability and accuracy at predicting a restrictive process is unknown. Currently, the gold-standard diagnosis of restrictive pulmonary impairment requires measurement of the total lung capacity (TLC) through helium dilution or plethysmograph techniques.2
A low FVC can result from true restrictive pulmonary disease, or from airflow obstruction with excessive air trapping. In the absence of airflow obstruction, it has been suggested that a restrictive defect can be diagnosed by spirometry alone.3 Therefore, a low spirometric FVC together with a normal or high FEV1/FVC ratio has traditionally been classified as a restrictive abnormality.3 However, studies verifying the accuracy of this interpretation are lacking.
Similarly, there are no studies verifying the accuracy of spirometry at ruling out restrictive impairment. Although it has been suggested that lung volume measurements may be useful in providing evidence of a diagnosis of restrictive lung disease not clearly evident from spirometric testing,1 this would be true only if spirometry had a poor negative predictive value for ruling out restrictive disease.
It would be convenient to be able to diagnose, or rule out, the possibility of a restrictive defect by spirometry alone, without measurement of TLC, because measurement of lung volumes is time-consuming and expensive. Avoidance of unnecessary lung volume testing can help to achieve a reduction in health care costs. The purpose of this study was to determine the predictive value of spirometry for diagnosing, and for ruling out, a restrictive pulmonary defect in a population of consecutive patients referred to a tertiary care pulmonary function laboratory.
Materials and Methods
Test results from all white patients who had undergone both spirometry and lung volume measurements on the same visit from January 1, 1991 to January 1, 1993 were included. Non-whites, and patients with technically inadequate tests (because of poor patient effort or inability to perform spirometry) were excluded. Pulmonary function tests were ordered at the discretion of the patients' physicians. If the patient had undergone multiple lung volume measurements over this 2-year period, we used only the measurements from their first visit. Lung volumes were always performed after spirometry. For patients who had undergone spirometry before and after administration of a bronchodilator (salbutamol, 200 µg, via metered-dose inhaler with a spacer device), the postbronchodilator spirometry was used as the reference and was compared to the postbronchodilator lung volume measurements.
Spirometric testing was done by experienced pulmonary technologists
certified and licensed by The Canadian College of Respiratory
Therapists and The Canadian Association of Cardio-Pulmonary
Technologists. Spirometry was performed according to American Thoracic
Society criteria.4
Testing was done with the subjects
seated, and a maximum forced exhalation was carried out for a minimum
of 6 s. After 6 s, the test continued until zero flow was
achieved (
50-mL flow extrapolated over 30 s). Testing was
repeated until a minimum of three acceptable flow volume loops with
FEV1 and FVC within 5% were obtained. Helium
dilution and plethysmograph techniques were similarly performed
according to published guidelines.5
For helium dilution,
patients were asked to perform an inspiratory capacity maneuver every
30 s to speed equilibration, and attempt to open closed lung
areas. Mixing time was prolonged until the stability of helium dilution
remained constant (< 0.02% variation) for a minimum of 30 s.
For plethysmography, a minimum of two attempts with the FRC
reproducible within 5% were made for each patient. Spirometric and
helium lung volume measurements were made using a Gould 2450 Complete
PF System (Gould; Dayton, OH) or a PK Morgan Complete PF System (PK
Morgan; Chatham, England). Spirometric and plethysmographic lung volume
measurements were made using a body plethysmograph (2800 AutoBox;
Gould). All machines were calibrated with a 3-L volumetric syringe
daily.
Lung volume testing was performed by the method of helium dilution in 77% of cases, and by plethysmograph in 22% of patients; 1% of patients had both methods used to determine lung volumes.
Measurements of FVC, FEV1,
FEV1/FVC, and TLC were expressed as a percentage
of predicted to control for the influence of age, gender, and height.
We used published predicted equations taken from healthy, nonsmoking
white patients. Spirometric reference values used were from Knudson et
al6
for those patients < 65 years old and from Enright
et al7
for patients
65. Lung volume reference values
used were from Crapo et al.8
Test values for the FVC,
FEV1/FVC ratio, and TLC that fell below the lower
limit of the 95% confidence interval (CI) of the predicted value were
classified as abnormal.
Spirometry data were classified categorically as being consistent with a normal pattern (normal FVC and FEV1/FVC ratio), an obstructive pattern with a normal FVC (reduced FEV1/FVC ratio, normal FVC), an obstructive pattern with a reduced FVC (reduced FEV1/FVC ratio, reduced FVC), or a restrictive pattern (reduced FVC, normal or above normal FEV1/FVC ratio). The probability of restrictive pulmonary impairment, defined as a reduced TLC below the lower limit of the 95% CI, was determined for each of the categoric classifications of the spirometric data, and additionally for each of several interval levels of the FVC and the FEV1/FVC ratio.
CIs around estimates of probability were calculated using the binomial
distribution, except when point estimates for probability were < 0.05
in which case the Poisson distribution was used. The relationship
between individual spirometric variables and the probability of
restrictive disease was assessed in a logistic model with
2 testing for linear trend. p Values < 0.05
were accepted as significant.
Results
A total of 1,879 pulmonary function tests were performed on white patients during the 2-year study period. 1,831 of these tests were technically adequate and therefore were included in data analysis.
The mean (± SD) age of the patients was 50.8 ± 17.6 years, and 49.8% of the patients were males. Four hundred seventy patients (25.7%) had an FVC on spirometry that was below the 95% predicted CI. Two hundred twenty-five patients (12.3%) had a restrictive defect confirmed on lung volume testing. Ten percent of those patients who were tested with plethysmography were found to be restricted vs 13.0% of those patients tested using the helium dilution method (p = 0.10).
A simple spirometry classification system is pictured in Table 1 . Patients were grouped as having a normal spirometric FVC or a low FVC (FVC below the lower limit of the predicted 95% CI). The sensitivity and specificity of a low FVC for determining true restrictive pulmonary impairment were 86% and 83%, respectively. However, the positive predictive value of spirometry for predicting restriction in this population was relatively low; of 470 patients with a low FVC on spirometry, only 41% had restriction confirmed on lung volume measurements. Conversely, the negative predictive value was quite high; only 2.4% of 1,361 patients with a normal VC on spirometry had a restrictive defect by TLC measurement.
|
We then reclassified the spirometric data into four physiologic groups (see Table 2 ). Spirometry data were classified categorically as being consistent with a normal pattern, an obstructive pattern with a normal FVC, an obstructive pattern with a reduced FVC, or a restrictive pattern, and the prevalence of pulmonary restriction among the four spirometric physiologic groups was determined (Table 2 ). When the definition of a restrictive spirometric pattern was thus narrowed to include only those patients with a low FVC and a normal or above normal FEV1/FVC ratio, the sensitivity of the classification scheme for determining restrictive impairment fell to 68% with a specificity of 93% (Table 3 ). The positive predictive value was 58%.
|
|
|
|
This study has shown that spirometry is very useful at excluding a restrictive defect. However, the classical spirometric pattern of restriction can be misleading, because it represents a true restrictive defect in < 60% of cases.
In this study, the probability of a restrictive pulmonary defect was
only 2.4% for the 1,361 patients with a normal FVC on spirometry. This
very low false-negative rate is reassuring, because this rate falls
within the expected
error of the test, ie, we would have
expected as many as 2.5% of normal patients to show a low TLC on lung
volume measurements.
The very high negative predictive value of spirometry has enormous cost-saving implications. Lung volumes are, for the most part, only useful for diagnosing pulmonary restriction. Therefore, given the high negative predictive value of spirometry, 1,361 of 1,831 lung volume tests could have been avoided in our patient population. In the province of Ontario, the addition of lung volume measurements to spirometry adds $32.15 (Canadian) to the cost of pulmonary function testing.9 If we consider that measurement of lung volumes yielded new diagnostic information in only 32 of 1,361 patients with a normal spirometric FVC, then 42.5 tests would have been required to obtain each single test providing new diagnostic information. Thus, for patients with a normal FVC, the cost of one lung volume test providing new diagnostic information was $32.15 x 42.5 = $1,366.38 Canadian. These calculated cost savings might be expected to be even greater in centers in the United States where the charge for performing and interpreting lung volume measurements is on the order of $130.00 US per test.
Conversely, for those patients whose spirometry demonstrated the classical restrictive pattern, only 58% of patients were found to have true restriction when lung volumes were measured. This suggests that measurement of lung volumes is necessary to confirm a restrictive impairment in this subset of patients.
One potential criticism of our study is that lung volume measurements were performed using helium dilution techniques for 77% of patients and by using plethysmography for the remainder. In healthy subjects, helium dilution and plethysmography yield very similar results,10 and the coefficient of variation of repeated measurements for both methods, on the same subject, is < 10%.11 However, in patients with severe airflow obstruction, the true lung volume can be underestimated by the dilution method, unless mixing time is prolonged.12 To ensure that our method of lung volume measurement did not introduce systematic bias into our data, we repeated analyses for those patients who had lung volumes measured by plethysmography separately from those who had lung volumes measured by helium dilution. Results were similar for both groups of patients. In addition, we calculated the percent variation by which TLC measures of helium dilution differed from plethysmography in a sample of 16 patients from our data set. These 16 patients, all of whom had spirometric evidence of moderate to severe expiratory airflow obstruction (mean FEV1, 1.61 ± 0.76 L; mean FEV1/FVC, 0.55 ± 0.15), had lung volume measurements done by both methods on the same visit to the pulmonary function laboratory during the study time period. The mean percent variation by which TLC measures of helium dilution differed from plethysmograph in these obstructed patients was only 9.7 ± 8.1%.
Finally, because helium dilution techniques can only underestimate lung volumes in patients with severe airflow obstruction, the effect, if any, of this potential bias would have been to overdiagnose restrictive impairment in those patients with airflow obstruction. This phenomenon was not seen in our patient population; of 260 patients with obstruction to airflow and a normal FVC, only 0.8% had evidence of restriction on lung volume testing.
The patients were selected for this study if spirometry and lung volumes had been measured on the same visit to the pulmonary function laboratory during the 2-year study period. This may have introduced the possibility of selection bias, ie, only selected patients were sent for both tests. Although selection bias may have been present, the effects of this bias would not have changed the conclusions of this study. Presumably, patients selected by their physicians for lung volume measurements had a greater prior probability of restrictive disease; however, despite this, we found that patients with a normal FVC on spirometry had a very low probability of having restrictive pulmonary impairment on measurement of lung volumes.
Recently, Lefante et al13
published a study in which they
attempted to derive an equation to account for the effects of airflow
obstruction on the observed FVC. They found that in patients with an
FEV1/FVC ratio < 0.70, obstruction explained 15
to 17% of the variability in FVC percent predicted. They found that in
these obstructed patients, the predicted FVC could be adjusted for the
effects of obstruction according to the following equation:
![]() |
In summary, this study has served to determine the accuracy with which spirometric measurements of FVC and expiratory flow rates can diagnose the presence of a restrictive impairment. Patients whose FVC fall above the 95% CI of the predicted value are very unlikely to have a restrictive impairment, and in these patients, unless restrictive disease is suspected a priori, measurements of lung volumes can be avoided.
This study has also demonstrated that, contrary to popular opinion, a restrictive pattern on spirometry does not accurately predict true restrictive disease. Even in those patients with a classic spirometric restrictive pattern, the probability of true restrictive disease was < 60%. For any patient with a low VC, measurement of lung volumes is necessary before the patient is diagnosed as having a restrictive pulmonary impairment.
Acknowledgements
ACKNOWLEDGMENT: The authors thank Ms. Mary Lunan for her assistance with data gathering.
Footnotes
Supported by a grant from The Ontario Thoracic Society.
Correspondence to: Dr. S. Aaron, The Ottawa General Hospital, Room LM-17, 501 Smyth Rd, Ottawa, Ontario, Canada K1H 8L6; e-mail: saaron@ogh.on.ca
Abbreviations: CI = confidence interval; FRC = functional residual capacity; TLC = total lung capacity; VC = vital capacity
Received for publication March 30, 1998. Accepted for publication August 13, 1998.
References
This article has been cited by other articles:
![]() |
J. G. Pasipanodya, T. L. Miller, M. Vecino, G. Munguia, S. Bae, G. Drewyer, and S. E. Weis Using the St. George Respiratory Questionnaire To Ascertain Health Quality in Persons With Treated Pulmonary Tuberculosis Chest, November 1, 2007; 132(5): 1591 - 1598. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Pasipanodya, T. L. Miller, M. Vecino, G. Munguia, R. Garmon, S. Bae, G. Drewyer, and S. E. Weis Pulmonary Impairment After Tuberculosis Chest, June 1, 2007; 131(6): 1817 - 1824. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. F. Pedersen FEV6: a shortcut in spirometry? Eur. Respir. J., February 1, 2006; 27(2): 245 - 247. [Full Text] [PDF] |
||||
![]() |
M. Akpinar-Elci, K. B. Fedan, and P. L. Enright FEV6 as a surrogate for FVC in detecting airways obstruction and restriction in the workplace Eur. Respir. J., February 1, 2006; 27(2): 374 - 377. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Pellegrino, G. Viegi, V. Brusasco, R. O. Crapo, F. Burgos, R. Casaburi, A. Coates, C. P. M. van der Grinten, P. Gustafsson, J. Hankinson, et al. Interpretative strategies for lung function tests Eur. Respir. J., November 1, 2005; 26(5): 948 - 968. [Full Text] [PDF] |
||||
![]() |
J. Vandevoorde, S. Verbanck, D. Schuermans, J. Kartounian, and W. Vincken FEV1/FEV6 and FEV6 as an Alternative for FEV1/FVC and FVC in the Spirometric Detection of Airway Obstruction and Restriction Chest, May 1, 2005; 127(5): 1560 - 1564. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. P. Swanney, L. E. Beckert, C. M. Frampton, L. A. Wallace, R. L. Jensen, and R. O. Crapo Validity of the American Thoracic Society and Other Spirometric Algorithms Using FVC and Forced Expiratory Volume at 6 s for Predicting a Reduced Total Lung Capacity Chest, December 1, 2004; 126(6): 1861 - 1866. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Glady, S. D. Aaron, M. Lunau, J. Clinch, and R. E. Dales A Spirometry-Based Algorithm To Direct Lung Function Testing in the Pulmonary Function Laboratory Chest, June 1, 2003; 123(6): 1939 - 1946. [Abstract] [Full Text] [PDF] |
||||
![]() |
D M Mannino, A S Buist, T L Petty, P L Enright, and S C Redd Lung function and mortality in the United States: data from the First National Health and Nutrition Examination Survey follow up study Thorax, May 1, 2003; 58(5): 388 - 393. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Reich and S. Wagers Supranormal Expiratory Flow Rates in Patients With Interstitial Lung Disease Chest, December 1, 2000; 118(6): 1836 - 1836. [Full Text] [PDF] |
||||
![]() |
M. P. SWANNEY, R. L. JENSEN, D. A. CRICHTON, L. E. BECKERT, L. A. CARDNO, and R. O. CRAPO FEV6 Is an Acceptable Surrogate for FVC in the Spirometric Diagnosis of Airway Obstruction and Restriction Am. J. Respir. Crit. Care Med., September 1, 2000; 162(3): 917 - 919. [Abstract] [Full Text] |
||||
![]() |
A. F. Shorr and O. W. Hnatiuk Circulating D Dimer in Patients With Sarcoidosis Chest, April 1, 2000; 117(4): 1012 - 1016. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Stanescu Small Airways Obstruction Syndrome Chest, July 1, 1999; 116(1): 231 - 233. [Abstract] [Full Text] [PDF] |
||||
Read all eLetters
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |