|
|
||||||||
Guest Access | Sign In via User Name/Password |
|||||||||
* From the Barnett Medical Aerosol Research Laboratory (Drs. Newhouse and Nantel and Mss Chambers and Pratt), St. Joseph's Hospital-McMaster University, Hamilton, Ontario, Canada, and ML Laboratories PLC (Dr. Parry-Billings), St. Albans, Hertfordshire, UK.
| Abstract |
|---|
|
|
|---|
Design: Randomized, double-blind, placebo-controlled comparison of the bronchodilator response to albuterol DPI (200 µg) at inspiratory flow rates of approximately 15, 30, and 60 L/min in patients with stable asthma with demonstrated reversibility to albuterol. Active (albuterol via pMDI inhaled at 30 L/min) and placebo controls were included.
Setting: Single center study at the chest/allergy unit of a teaching hospital in Canada.
Patients: Sixteen patients with moderate to moderately severe stable asthma.
Measurements and results: Efficacy end points were FEV1, FVC, FEV1/FVC, maximum expiratory flow, and forced expiratory flow between 25% and 75% of vital capacity. Safety end points included heart rate, BP, and tremor. There was no significant difference between the bronchodilator response to albuterol via the CH at 15, 30, and 60 L/min inspiratory flow rate and, at all flow rates, no significant difference was found comparing albuterol CH with the pMDI. All of the techniques for delivering albuterol provided significantly better bronchodilatation than placebo. Adverse events were minimal and did not differ between CH and pMDI or between the various flow rates inhaled through the CH.
Conclusion: A novel passive albuterol DPI (CH) provides a similar bronchodilator response at 15, 30, and 60 L/min inspiratory flow rates compared with a pMDI used optimally.
Key Words: albuterol bronchodilatation dry powder inhaler inspiratory flow rate
| Introduction |
|---|
|
|
|---|
All currently available (so-called "passive") DPIs require the patient's vigorous inspiratory effort to both disperse and deliver the powder formulation. It has also been shown that, when using a passive DPI, inspiratory flow rate is correlated with drug deposition,3 ,4 with low inspiratory flow rates leading to reduced lower respiratory tract fine particle drug delivery and therapeutic response.5 The Clickhaler (CH) (ML Laboratories PLC; St. Albans, UK) is a novel passive DPI; therefore, it was considered important to evaluate the bronchodilator efficacy of the albuterol sulfate CH over a range of inspiratory flow rates, particularly the low inspiratory flow rates that might be encountered under conditions of severe airflow obstruction common to severe exacerbations of asthma or COPD and in children < 6 years.6
The study was designed to compare the bronchodilator response of a novel DPI (CH) at inspiratory flow rates of 15, 30, and 60 L/min with albuterol pMDI used optimally and placebo in patients with asthma.
| Materials and Methods |
|---|
|
|
|---|
18 years of
age and nonsmokers, or asthmatics with < 10 pack-years of smoking
history who had not smoked for at least 15 years. Patients had a
resting FEV1 of 40 to 80% of predicted normal and a
minimum 15% increase of FEV1 to 200 µg of albuterol
inhaled under the supervision of the study coordinator, using a pMDI
(Ventolin; Allen & Hanburys; Uxbridge, UK) with the closed mouth
method. Over the course of the study, patients continued their use of steroids, cromolyn sodium, or nedocromil sodium for asthma control. Prior to each study visit, use of asthma medication, caffeine, and alcohol was discontinued for at least 8 h. Patients with concomitant respiratory diseases such as cystic fibrosis, tuberculosis, or COPD were excluded from the study. The protocol was approved by the Institutional Review Board of St. Joseph's Hospital, Hamilton, Ontario, and by Health and Welfare, Canada (Health Protection Branch). Written informed consent was obtained from all patients.
Study Design
This was a single-center, randomized, double-blind, five-way
crossover study, that compared the effects of albuterol CH at
inspiratory flow rates of 15, 30, and 60 L/min, with an albuterol pMDI
used with lips closed about the mouthpiece at a monitored inspiratory
flow rate of 30 L/min and a placebo inhalation (pMDI at 30 L/min). The
study involved a screening visit and five randomized treatment visits.
Inspiratory flows through the CH (Fig 1 ) were determined by measuring the pressure at the mouthpiece using a small unobtrusive sampling tube leading to a pressure transducer.8 The resistance of the DPI had been calculated previously from its characteristic pressure-flow curve. Inspiratory flows through the metered-dose inhaler (MDI) were recorded with a pneumotachometer and displayed on an oscilloscope and chart recorder. The pneumotachometer was connected to the MDI boot by a plastic bulb approximately 7 cm in diameter and sealed around both end fittings. To allow MDI actuation, a hole was made into the side of the bulb and fitted with a sealed finger cot. A laboratory-grade electronic manometer was employed to calibrate both sensors to within an error of 0.05 cm H2O.
|
Study Assessments
At the screening visit, a flow volume loop was obtained using a
spirometer (Compact; Vitalograph Ltd; Buckingham, UK) before drug
administration to provide the FEV1, maximum expiratory flow
(MEF), FVC, forced expired flow between 25% and 75% of vital capacity
(FEF2575) and FEV1/FVC ratio. At the
screening visit, FEV1 was also measured postalbuterol via
pMDI. At each treatment visit, spirometry was performed pretreatment
and at 15, 30, 45, 60, 90, 120, 180, and 240 min posttreatment. At each
study visit, the baseline FEV1 was within ±15% of that
obtained at the screening visit.
Safety Assessments
BP, heart rate (by three-lead ECG rhythm strip),
tremor,9
and respiratory rate were recorded at each visit,
before treatment and at 30, 60, 120, and 240 min posttreatment. All
adverse events occurring throughout the study period were also
recorded.
Inspiratory Flow Rate Targeting and Monitoring
Following appropriate training, patients were able to provide
mean peak inspiratory flow rates (±25%) of 15, 30, and 60 L/min. The
inspiratory flows were measured and displayed in real time using a
computerized DPI flow targeting system (Clinical Flow Technologies;
Hamilton, Ontario, Canada).8
During inhalation, the
instrument measured mouth pressure via a small, unobtrusive tube at the
CH mouthpiece. Using the predetermined resistance of the CH, pressure
measurement values were converted to flow values. Patients observed
their inspiratory flows in real-time on a computer monitor and, after
appropriate training, were thus able to adjust their efforts to provide
inspiratory flow rates within the target range.
Data Analysis
Data are presented as mean ± SD unless otherwise indicated.
All analyses were compared between treatment groups by analysis of
variance.
The primary efficacy analysis was the magnitude of the change in FEV1. Secondary efficacy analyses were also performed on measurements of MEF, FEF2575, FVC, and the ratio of FEV1/FVC. The duration of the response was investigated using the area under the FEV1 response vs time curve between 0 and 240 min posttreatment. Differences were considered significant at the 5% level.
Safety analyses involved comparison of the mean changes in heart rate, BP, respiratory rate, and tremor between the treatment groups, compared with baseline pretreatment and placebo response over time.
| Results |
|---|
|
|
|---|
|
|
|
|
Vital Signs and Tremor
There were no clinically significant changes in systolic/diastolic
BP, heart rate, respiratory rate, or tremor following any of the study
treatments.
| Discussion |
|---|
|
|
|---|
Previous in vitro work to characterize the aerosol output from the CH has demonstrated that both the emitted dose and the particle size distribution of the emitted dose are relatively constant over the range of flow rates used in the present study.11 To our knowledge, the correlation between these in vitro measurements and in vivo measures of clinical efficacy, however, has not been studied previously with this device. In the present study, no significant differences were found in bronchodilator response between the CH over the clinically relevant range of flow rates studied (15 to 60 L/min) and the pMDI inhaled under near-optimum conditions. This finding was consistent for all spirometric parameters and for the duration of the response.
Studies have confirmed that most patients are able to achieve an inspiratory flow rate of at least 30 L/min through the CH. In pediatric patients with asthma (mean age, 12.9 years; range, 7 to 14 years), the mean peak inspiratory flow rate through the CH was 74.4 L/min and the lowest was 62.1 L/min.12 In the present study, all patients were able to achieve the flow rates required. Therefore, the range of flow rates evaluated represents those that are actually achieved by children > 6 years and adults with asthma in routine clinical practice.
Furthermore, the finding in this study of the clinical equivalence of albuterol pMDI and CH in asthmatics confirms the results of a previous study in 62 patients with reversible airflow obstruction.13 Another study with the albuterol CH used a pharmacokinetic method.14 Urinary excretion of albuterol and metabolites at 30 min and 24 h was measured as an indication of the relative lung deposition and total body bioavailability of albuterol, respectively, and showed that the relative lung bioavailability delivered via the CH was similar at relatively slow and fast inspiratory flow rates (30.1 ± 0.91 L/min and 58.9 ± 2.6 L/min, respectively).15 Further studies measuring lung deposition using gamma scintigraphy have demonstrated that there is no correlation between inspiratory flow rate and lung deposition.16
| Conclusion |
|---|
|
|
|---|
| Acknowledgements |
|---|
| Footnotes |
|---|
Correspondence to: M. Newhouse, MD, FCCP, Inhale Therapeutic Systems, 150 Industrial Road, San Carlos, CA 94070; e-mail: mikenewhouse@inhale.com
Abbreviations: CH = Clickhaler; DPI = dry powder inhaler; FEF2575 = forced expired flow between 25% and 75% of vital capacity; MDI = metered-dose inhaler; MEF = maximum expiratory flow; pMDI = pressurized metered-dose inhaler
Received for publication March 10, 1998. Accepted for publication September 4, 1998.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M.T. Newhouse, P. Patel, and M. Parry-Billings Protection against methacholine-induced bronchospasm: salbutamol pMDI versus Clickhaler(R) DPI Eur. Respir. J., May 1, 2003; 21(5): 816 - 820. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. S F Ram, J. Wright, D. Brocklebank, and J. E S White Systematic review of clinical effectiveness of pressurised metered dose inhalers versus other hand held inhaler devices for delivering beta 2 agonists bronchodilators in asthma BMJ, October 20, 2001; 323(7318): 901 - 901. [Abstract] [Full Text] |
||||
![]() |
P. J. Anderson Delivery Options and Devices for Aerosolized Therapeutics Chest, September 1, 2001; 120(3_suppl): 89S - 93S. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |