|
|
||||||||
Guest Access | Sign In via User Name/Password |
|||||||||
*
From the Service de Réanimation Médicale, Hôpital
Louis Mourier (Assistance Publique-H
pitaux de
Paris), Colombes, and Unité de Recherches Inserm U82, Faculté de
Médecine Xavier Bichat, Université Paris VII, Paris, France.
| Abstract |
|---|
|
|
|---|
Design: Prospective randomized clinical trial.
Setting: Medical ICU of Louis Mourier Hospital, Colombes, France, a university-affiliated teaching hospital.
Patients: Forty-five consecutive mechanically ventilated critically ill patients.
Interventions: Patients undergoing mechanical ventilation were randomly assigned to receive humidification with one of the four heat and moisture exchangers (HMEs) tested or with a conventional heated humidifier.
Measurements: The hygrometric performances of four HMEs (BB2215, BB50, and BB100 from Pall Biomedical, Saint-Germaine-en-Laye, France; and Hygrobac-Dar from Mallinckrodt, Mirandola, Italy) and a heated humidifier (Fisher & Paykel; Auckland, New Zealand) were studied after 3 h and also after 48 h of use for the Hygrobac-Dar and correlated to a clinical visual inspection rating the amount of condensation in the flex-tube of the endotracheal tube.
Results: A total of 95 measurements in 45 patients were performed. The best hygrometric parameters were obtained with the heated humidifier (p < 0.001). The Hygrobac-Dar yielded significantly higher values for both humidities and tracheal temperature than the other three HMEs (p < 0.001). The performance of Hygrobac-Dar was unchanged after 48 h of use. There was a significant correlation between the condensation seen in the flex-tube and the hygrometric parameters measured by psychrometry (absolute humidity, rho = 0.7; relative humidity, rho = 0.7; tracheal temperature, rho = 0.5, p < 0.0001).
Conclusion: In mechanically ventilated ICU patients, visual evaluation of the condensation in the flex-tube provides an estimation of the heating and humidifying efficacy of the heating and humidifying device used, thus allowing the clinician bedside monitoring of airway humidification.
Key Words: acute respiratory failure airway humidification endotracheal tube heat and moisture exchangers mechanical ventilation psychrometry
| Introduction |
|---|
|
|
|---|
The clinician wishing to use an HME must identify the ones that perform poorly without having to conduct hygrometric measurements. This may make him/her reluctant to use HMEs for all ICU patients. Recommendation for using or not using an HME lack evidence-based guidelines.12 Furthermore, indications for when to use an HME during mechanical ventilation (including the type of patients appropriate for HME use and the duration of ventilation with an HME) are still a matter of debate. Indeed, Branson and colleagues13 14 15 tend to limit their use of HMEs to surgical patients and for a maximum of 5 days of mechanical ventilation, while others, ourselves included, use HMEs for all patients, surgical and medical, including patients with COPD, whatever the duration of mechanical ventilation.8 9 10 11 16 However, as pointed out in a recent editorial,17 a simple method for evaluating HMEs is needed to allow comparison of the various brands of HMEs available and selection of the appropriate one for mechanical ventilation. The lack of evidence-based guidelines is undoubtedly due to the few data available from independent laboratories on the performance of HMEs. Indeed, the characteristics of the HMEs indicated by the manufacturer have not necessarily been established by an independent laboratory. Objective measurements are time consuming and require equipment that is not routinely available. This study was therefore carried out to provide an objective assessment of the performance of HMEs and to determine whether a clinician can assess this performance at the bedside by observing simple visual parameters. We recently reported that visual inspection of the amount of moisture in the flex-tube correlated adequately with the performance of the HME given by manufacturers.18 However, we did not check our visual scale against objective hygrometric measurements performed at the bedside. This study compares the amounts of moisture seen in the flex-tube with the hygrometric parameters measured psychometrically, thus enabling the clinician to evaluate the efficacy of a given HME without hygrometric measurements.
| Materials and Methods |
|---|
|
|
|---|
Materials
Four HMEs and one HH were tested. Two HMEs were purely
hydrophobic (Pall Ultipor Filter BB2215 and BB50; Pall Biomedical;
Saint-Germaine-en-Laye, France); the two others had both hygroscopic
and hydrophobic properties (Pall Ultipor Filter BB100; Pall Biomedical;
and HD). The conventional HH was the Fisher & Paykel MR 450; Auckland,
New Zealand (FP450). The ventilators used were either a Servo 900D
ventilator (Siemens-Elema; Solna, Sweden) or a Bird 8400 (Bird Products
Corp; Palm Springs, CA).
Hygrometric Measurements
Hygrometric measurements (absolute humidity [AH], relative
humidity [RH]) and tracheal temperature measurement were performed
after 3 h of mechanical ventilation for all devices and also after
48 h for the HD, which is routinely used for this time in our
unit.11
The number of times devices delivered AH below 24
mg H2O/L was recorded. AH and RH were obtained using the
psychrometric method.21
22
23
24
A device to separate
inspiratory and expiratory gas flow by the means of two unidirectional
valves was inserted between the endotracheal tube and the HME (or
between the endotracheal tube and the Y-piece when the HH was used).
Two thermal probesa dry one and a wet onewere placed in the
inspiratory part of the device. The temperatures recorded by the two
probes were measured and displayed on a chart recorder (Yokogawa;
Tokyo, Japan). Tracheal temperature was measured with another thermal
probe inserted in the endotracheal tube and also displayed on the chart
recorder. Mean temperatures were recorded from both probes after a
30-min period allowing for optimal thermal equilibrium. The
psychrometric method is based on comparing the temperatures obtained
with the two probes placed on the inspiratory part of the separating
device. The dry probe is placed upstream and measures the actual gas
temperature. The downstream probe is coated with sterile cotton wetted
with sterile water. Evaporation in the inspiratory part is proportional
to the dryness of the gas. The temperature gradient between the two
probes varies inversely as the humidity of the inspired gas. There is
no thermal gradient when the inspired gas is fully saturated with water
(100% RH). RH was calculated by reference to a psoriometric diagram
taking into account the temperature difference between the two
probes.25
AH at saturation point (100% of RH) (AHs) was
calculated with the following formula:
![]() |
![]() |
Visual Evaluation
The amount of moisture in the flex-tube connecting the HME (or
the Y-piece when the HH was used) to the endotracheal tube was
rated at the same time by an independent observer unaware of the
hygrometric results as follows: dry, moisture only, moisture + few
water droplets, moisture + several water droplets,
moisture + numerous water droplets, and dripping wet. These items
were assigned values from 1 (dry) to 6 (dripping wet). The first three
items of the scale were counted as "dry" and the last three as
"wet" in some analyses to make comparisons between devices easier.
This protocol was approved by the institutional review board for human studies of the French Intensive Care Society. Informed consent was not requested since all procedures were considered routine practice, and none was invasive.
Statistical Analysis
All results are means ± SD. Hygrometrics parameters were
compared by analysis of variance. When analysis of variance indicated
differences among groups, the groups were compared using protected
least significant difference. Kruskal-Wallis analysis was used to
compare the clinical rating scores of each humidifying device.
Contingency table analysis was used to compare the number of times AH
was < 24 mg H2O/L. The correlation between the clinical
rating of humidifying efficiency and the measured values was performed
using Spearman's rank correlation coefficient. A p value < 0.05 was
considered significant. Statistical analysis was done with a commercial
statistical package (Statview; Abacus Concepts Inc; Berkeley, CA).
| Results |
|---|
|
|
|---|
|
|
|
Correlations Between Hygrometric Measurements and Visual Evaluation
There was a significant correlation between the visual rating of
the condensation in the flex-tube and the AH measured by psychrometry
(rho = 0.7, p < 0.0001) (Fig 1)
.
The correlation was also significant with the other two parameters (RH
and tracheal temperature), though less powerful (rho = 0.7, and
rho = 0.5, respectively, both with p < 0.0001).
|
| Discussion |
|---|
|
|
|---|
The gas inspired during mechanical ventilation must be adequately heated and humidified. HMEs are increasingly employed because they are easy to use and cost-effective.8 9 11 Though the clinical importance of sufficient humidification has been often debated, there is no simple, visual method for evaluating the efficacy of HMEs during long-term mechanical ventilation. Indirect parameters such as the occurrence of nosocomial pneumonia, number of endotracheal tube occlusions, number of tracheal suctionings, and instillations required8 9 11 have been reported, but they do not always reflect accurately the hygrometric performance of a given HME.16 Indeed, in this previous report,16 the clinical evaluation of three different HMEs found no differences in humidifying properties, whereas hygrometric measurements revealed great differences between them. The only direct parameters suitable for evaluation are the hygrometric measurements. They may be expressed as an amount of water loss26 or expressed as humidity outputs obtained by psychrometric method. This method is the one most often used, but it requires a flow divider, three thermal probes, and > 30 to 60 min to ensure thermal equilibrium.4 5 6 These measurements are not available in all ICUs and are not feasible in routine practice. This is why a suitable visual evaluation method is worthwhile. We found that the humidity in the flex-tube was correlated with values provided by the manufacturers for AH.18 We have now evaluated each heating and humidifying device using a visual rating and precise hygrometric measurements to determine whether there was a correlation between the two means of evaluation on the same patient.
A group of four commonly used HMEs was tested to validate our visual rating of the condensation in the flex-tube. As the room temperature was constant in our unit, the condensation in the flex-tube reflected the humidity delivered by the HME, rather than thermal variations. The hygrometric parameters measured were consistent with those measured in other studies for each HME tested, and confirm the superiority of the HH FP450 and of the HD.21 22 23 24 The humidifying properties of this hygroscopic and hygrophobic filter have been studied and it consistently gives better hygrometric parameters than many other HMEs. These data were also obtained after use of the HME for 24 h or 48 h.16
We selected HMEs having very different efficacies to emphasize the differences in the visual assessment. It would have been difficult to detect hygrometric differences with a visual scale if all the HMEs had produced similar humidities. Thus, the visual rating provides an accurate, satisfactory evaluation of both the effective HMEs (HD and BB100), and the less effective ones (BB2215 and BB50).
The AH delivered by the HME to the patient via the inspired gas is a decisive determinant of the humidifying performance of an HME. The lower the AH, the higher total amount of water lost by the patient during mechanical ventilation. The ideal gas is 37°C with 100% RH (44 mg H2O/L water vapor) because it does not remove water from the patient.27 This is clinically important since endotracheal tube occlusions occurred when the AH delivered was insufficient.3 4 5 6 These occlusions occurred after < 12 h of use3 and some were lethal.4 Endotracheal tube occlusions occur after gradual reduction of the tube's diameter, and depend partly on the type of humidification device used, purely hydrophobic HMEs having a significantly greater reduction of their tracheal tube diameter.7 The precise AH an HME should deliver to avoid endotracheal tube occlusion is not clear, but some recommendations suggest that 24 mg H2O/L is the minimum an HME should deliver when used for long-term ventilation.28 However, International Standards Organization recommends that a minimum of 30 mg H2O/L should be provided by devices during mechanical ventilation. Interestingly, a number of studies evaluating different HMEs at the bedside report, however, much lower humidity output measured with commonly used HMEs. Indeed, measured values of absolute humidity are well below 30 mg H2O/L: 20.6 mg H2O/L for the Pall BB2215,22 22.9 mg H2O/L for the Intersurgical Filtatherm (Intersurgical; Wokingham, England),21 25.8 mg H2O/L for the Clear-Thermal (Intersurgical),16 26 mg H2O/L for the Pall BB50,23 and 28.8 mg H2O/L for the Hygrobac filter.22 It also appears that no endotracheal occlusion has been reported with AH > 24 mg H2O/L. This is the reason why 24 mg H2O/L was set as the lowest acceptable inspired gas AH. In this setting, the scale described herein enables satisfactory rating of the poorly performing HMEs, since most of those delivering < 24 mg H2O/L AH were rated "dry." Similarly, most of those delivering > 24 mg H2O/L AH were rated "humid." Our findings are consistent with those of Miyao and colleagues27 who reported that the condensation in the breathing circuit is a good way of monitoring the humidity delivered by a humidifying device. They also speculated that the condensation helps to prevent the secretions inside the endotracheal tube from drying out.
Both HMEs and HHs present several drawbacks. Placing an HME in the respiratory circuit generates gasflow resistance. This increase in the circuit's resistance has been studied previously; it reaches approximately 2.5 cm H2O/Ls and is related to the water load of the HME.29 30 The clinical impact of this resistance has not been studied extensively. Conti and coworkers31 found that in COPD patients receiving volume-controlled mechanical ventilation, the presence of an HME in the circuit did not modify intrinsic positive end-expiratory pressure. Moreover, the additional resistance occasioned by the HME is not higher than the one caused by an HH.32 Additional dead space may also be caused by an HME, and increased PaCO2 and minute ventilation during weaning from mechanical ventilation in comparison with HHs.33 However, a possible increase in total weaning time caused by HMEs has not been evaluated (to our knowledge) and the problem of increased dead space is easily overcome by increasing pressure support level.34 It has been shown clearly in the literature either by in vitro or in vivo studies that increase in tidal volume or minute ventilation affected the performances of some HMEs. However, only the purely hydrophobic HMEs (Pall BB2215 and BB50) suffered from this drop in performance. Indeed, Martin and coworkers22 have shown that the efficacy of the HD was not affected by minute ventilation > 10 L/min. In our study, humidifying devices were used with minute ventilation > 10 L/min. In fact, minute ventilation ranged from 7 to 15 L/min. Therefore, our scale can assess the performances of a humidifying device whatever the minute ventilation used. The use of purely hydrophobic HMEs cannot be recommended during mechanical ventilation with minute ventilation > 10 L/min for which clinicians may use either hydrophobic and hygroscopic HMEs or HHs.22 23 Excessive humidification (which can be caused only by a HH) can lead to epithelial and mucociliary damage.35 However, insufficient humidification still remains the major problem that faces a clinician wishing to choose a humidifying device. Evaluation of the efficacy of this device whether it be an HH or an HME can be achieved with this simple inspection.
This inspection must be performed several times during the use of the HME. A flex-tube that is always dry indicates a HME delivering insufficient humidity, and therefore at risk of endotracheal tube occlusion. In such instances, we believe it is wise either to change the HME used for a high-performing HME or to switch the patient to an HH. In our unit, we use only high-performing HMEs and such instances have never occurred whatever the type of patient.9 11 16 However, a flex-tube that is constantly covered in water droplets indicates that the HME is appropriate for long-term mechanical ventilation without fear of endotracheal tube occlusion. Our observations indicate that clinicians can evaluate the performance of an HME at the bedside. This may result in considerable cost containment since mechanical ventilation with HMEs is much less expensive than with HH,8 9 11 and could lead to the widespread use of HMEs with further substantial savings as long as contraindications to the use of HMEs be carefully respected; these include profound hypothermia, bronchopleural fistulas, and breath-eliminated drug poisoning. It is thus possible to evaluate the efficacy of HMEs visually by rating the amount of moisture in the flex-tube connecting the HME to the endotracheal tube.
| Footnotes |
|---|
Correspondence to: Didier Dreyfuss, MD, Service de Réanimation Médicale, Hôpital Louis Mourier, 178, rue des Renouillers, 92700 Colombes, France; e-mail: didier.dreyfuss@lmr.ap-hop-paris.fr
Abbreviations: AH = absolute humidity; HD = Hygrobac-Dar; HH = heated humidifier; HME = heat and moisture exchanger; RH = relative humidity; SAPS = Simplified Acute Physiology Score
Received for publication July 8, 1998. Accepted for publication December 1, 1998.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
F. Lellouche, S. Taille, S. M. Maggiore, S. Qader, E. L'Her, N. Deye, and L. Brochard Influence of Ambient and Ventilator Output Temperatures on Performance of Heated-Wire Humidifiers Am. J. Respir. Crit. Care Med., November 15, 2004; 170(10): 1073 - 1079. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Paluch, J.-D. Ricard, D. Dreyfuss, W. R. Clarke, L. P. Nielsen, and R. Dahl EFFICIENCY AND SAFETY OF MECHANICAL VENTILATION WITH A HEAT AND MOISTURE EXCHANGER CHANGED ONCE A WEEK Am. J. Respir. Crit. Care Med., November 15, 2001; 164(10): 1999 - 2000. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |