|
|
||||||||
Guest Access | Sign In via User Name/Password |
|||||||||
* From the Intensive Care Unit (Drs. Miyoshi, Fujino, Mashimo, and Nishimura) and Department of Anesthesiology (Dr. Mashimo), Osaka University Hospital, Osaka Japan.
Correspondence to: Yuji Fujino, MD, Intensive Care Unit, Osaka University Hospital, 215, Yamadaoka, Suita, Osaka, Japan 565-0871; e-mail: fujino{at}hp-icu.med.osaka-u.ac.jp
| Abstract |
|---|
|
|
|---|
Study design: Laboratory study with a mechanical lung model.
Methods: We compared the performance of four transport
ventilators (model 740, Mallinckrodt, Pleasanton, CA; TBird, Bird
Products Corp, Palm Springs, CA; LTV1000, Pulmonetic Systems, Colton,
CA; Esprit, Respironics, Vista, CA) with a standard ICU ventilator
(model 7200ae; Mallinckrodt) using a test lung that simulated
spontaneous breathing (compliance, 46.8 mL/cm H2O;
resistance, 5 cm H2O/L/s). The settings of ventilators were
positive end-expiratory pressure (PEEP) of 0 or 5 cm H2O,
and pressure support (PS) of 0 or 10 cm H2O. The settings
of the test lung were inspiratory time of 1 s, respiratory rate of
10/min, peak inspiratory flow of 40, 60, and 80 L/min. To evaluate
inspiratory function at each setting, we measured the inspiratory delay
time (DT), inspiratory trigger pressure (P-I), and the time for airway
pressure to rise from the baseline pressure to 90% of the
end-inspiratory pressure (T90%); for expiratory function,
supraplateau expiratory pressure (P-E) and the time constant (
e) for
pressure decrease during exhalation were evaluated. Oxygen requirement
was assessed as the time required to empty a 3.5-L oxygen tank.
Results: For inspiratory triggering, four transport
ventilators had DT < 100 ms, which is considered clinically
satisfactory, in all the settings except for PS 0 cm H2O,
PEEP 0 cm H2O, and inspiratory flow of 80 L/min with
LTV1000. P-I increased only in LTV1000 when PEEP was increased from 0
to 5 cm H2O.
e for the transport ventilators was
> 50% shorter than for the ICU ventilator except for PS 0 cm
H2O and PEEP 5 cm H2O with TBird. Oxygen
requirement was lowest for the Esprit, followed by the 740, LTV1000,
and TBird.
Conclusion: The newer Food and Drug Administrationapproved transport ventilators have performance indexes comparable to the ventilator currently used in ICUs and can probably be recommended for clinical use.
Key Words: ARDS transport ventilator performance work of breathing
| Introduction |
|---|
|
|
|---|
Although a number of studies of ventilator performance have been published, they have generally been concerned with ICU ventilators5 6 and home-care ventilators.7 8 Since the advent of pressure-support ventilation, ICU ventilators have provided improved synchrony with patient breathing and increased patient comfort during mechanical ventilation. The basic requirements for transport ventilators include physical ease of portability, an independent power source, and low oxygen utilization; these design considerations have technically limited the performance of transport ventilators. The transport ventilators that provide inspiratory triggering functions and pressure supportcontrol modes to preserve spontaneous breathing have recently become commercially available. In this study, we evaluated these ventilators in comparison with a standard ICU ventilator, using a lung model simulating spontaneous breathing.
| Materials and Methods |
|---|
|
|
|---|
|
|
Experimental Protocol and Data Analysis
Experiment 1:
Each ventilator was set at positive
end-expiratory pressure (PEEP) 0 or 5 cm H2O and
pressure support (PS) of 0 or 10 cm H2O, and used
to ventilate the model lung, which was set to require inspiratory flow
rates of 40, 60, and 80 L/min. Except for the 7200ae (-1 cm
H2O), flow triggering was used in all transport
ventilators. Sensitivity was set at 1 L/min except for the Esprit (0.5
L/min) as the most sensitive setting without self-triggering.
Figure 2
shows the variables that we analyzed to evaluate ventilator
performance. The time between the start of inspiration to the point of
minimum airway pressure was recorded as inspiratory delay time (DT).
Inspiratory trigger pressure (P-I) records the difference between the
baseline pressure and the maximum subbaseline pressure established
during triggering of inspiration. T90% values
represent the time for airway pressure to rise from the baseline
pressure to 90% of the end-inspiratory pressure. Supraplateau
expiratory pressure change (P-E) expresses the pressure change from the
end-inspiratory value after the onset of exhalation. The rate of airway
pressure decrease during exhalation was evaluated from the time taken
for airway pressure to decrease from the peak value to 37% of peak
value (time constant [
e]).
|
| Results |
|---|
|
|
|---|
|
|
|
|
e was not affected
by increasing flow, and all four transport ventilators showed > 50%
shorter values than the 7200ae except for PS 0 cm
H2O and PEEP 5 cm H2O
(Table 2)
. The time taken for the tested ventilators to empty the O2 tank was 25 min 18 s for the TBird, 44 min 40 s for the 740, 57 min 19 s for the Esprit, and 31 min 8 s for the LTV1000.
| Discussion |
|---|
|
|
|---|
Mechanically ventilated critically ill patients often require transport to perform diagnostic or therapeutic procedures that cannot be conducted in the ICU. Outside of the ICU, patients are ventilated manually or mechanically with a transport ventilator. Braman et al1 have reported that clinically significant hemodynamic and blood gas changes occurred in 16 of 20 patients receiving manual ventilation during transport. They concluded that patients should be ventilated with a volume ventilator during transport. Weg and Haas,4 however, could not discern any hemodynamic and blood gas change in 20 patients during transport and concluded that mechanical portable transport ventilators should not replace well-trained personnel. Hurst et al2 and Gervais et al3 have reported respiratory alkalosis in patients ventilated manually during transport. Gervais et al3 also showed that the unintended respiratory alkalosis during manual ventilation could be avoided when minute ventilation was monitored using a spirometer. There has been no evidence that transport ventilators outperform manual ventilation. In previous reports, the transport ventilators studied were volume or time-cycled ventilators lacking the ability to synchronize with the breathing of patients.
By contrast, the ventilators that we evaluated in this study feature sophisticated pressure supportcontrol modes and are able to synchronize with the patients breathing. Using a model lung, we have found that they performed as well as the modern ICU ventilator. In this model lung study, we followed the methodology of a previously published report.7 As the inspiratory function, we evaluated DT as the representative value of inspiratory triggering and T90% as the representative value of flow delivery. The DTs of all ventilators were < 100 ms, which is considered satisfactory9 except for in one setting (PS, 0 cm H2O; PEEP, 0 cm H2O, and inspiratory flow of 80 L/min) with the LTV1000. The performance of inspiratory triggering of the LTV 1000 declined with increasing PEEP (Table 2 and Fig 4 ). This was probably because the exhalation valve attached to the circuit did not work as well as those in the other ventilators. Small T90% values in the four transport ventilators show their capacity to meet the inspiratory flow demand in critically ill patients. The Esprit that we tested was still a prototype at the time of evaluation, and we could not apply some of the test lung settings because of intractable self-triggering. This problem may have been corrected in the commercial version.
Ventilator performance during expiration also correlates to the work of
breathing and comfort in mechanically ventilated patients. As an
expiratory factor, we evaluated P-E as the representative value of
expiratory triggering function. The resistance of the expiratory system
is reflected as the combination of P-E and
e. With the LTV 1000, the
P-E values were greater than those with the other ventilators. In this
study we set the expiratory trigger sensitivity of each transport
ventilator at 25%. P-E can be minimized by adjusting expiratory
trigger sensitivity in the LTV1000 and the Esprit (Table 1)
. Proper
adjustment is not easy, however, because none of the transport
ventilators in this study comes with a display to monitor waveform in
the standard configuration. All four transport ventilators outperformed
the 7200ae with regard to
e. We do not conclude, however, that these
transport ventilators have better expiratory performance than other ICU
ventilators, because the 7200ae is known to have a high expiratory
resistance.7
Recent recommendations for ARDS patients include keeping the lungs open by adding sufficient levels of PEEP, with limitations for peak alveolar pressure to avoid additional lung injury by mechanical ventilation.10 11 It is difficult to accomplish this sophisticated ventilation technique with manual ventilation. Furthermore, Zakynthinos et al12 reported that early ARDS patients ventilated with PS showed better oxygenation than with controlled ventilation probably owing to alveolar recruitment augmented by active diaphragmatic contraction. These new transport ventilators not only save the labor of the person who would otherwise have to administer manual ventilation, but they also provide the same quality of mechanical ventilation during transport as ventilators in the ICU.
The basic practical requirements, in addition to the ability to synchronize with the breathing of patients, for transport ventilators are: (1) compact form, (2) low oxygen requirement, and (3) a built-in independent power source. The most compact ventilator that we tested in this study was the LTV 1000 (Table 1) . In our findings, the order of oxygen requirement grossly followed that of constant flow rate during the expiratory phase. Transport ventilators that more sensitively detected inspiration (740 and Esprit) are bulkier than the other two, possibly because of more complex detection systems. The oxygen requirement of the Esprit can be changed by adjusting the flow sensitivity setting. We set flow sensitivity of the Esprit at 0.5 L/min in this study, which corresponds to a total base flow of 3.5 L/min (Table 1) . Reducing the flow sensitivity of the Esprit will cause an increase in the oxygen requirement. To evaluate the oxygen requirement, we used an FIO2 of 1.0 to avoid the effects of inaccurate FIO2 values. We did not test the capacity of the batteries used in these units. According to the manufacturers information, however, each of these transport ventilators provide sufficient battery capacity.
The use of a model lung limited the usefulness of this study. First, it is impossible to evaluate gas exchange or the comfort of patients. Second, performance may be dependent on the model lung. At least it can be concluded from our results that the transport ventilators tested in this study can provide the same quality of mechanical ventilation during transport. The performance of these new transport ventilators requires further evaluation in a clinical setting before being recommended for critical care patients. Judging from our findings, we hypothesize that they may be able to do away with the need for manual ventilation during transport.
In conclusion, the newer US Food and Drug Administrationapproved transport ventilators available for clinical use have specific performance indexes comparable to ventilators currently used in ICUs. Toward that end, they can probably be recommended for clinical use, specifically for the transport of critically ill patients.
| Footnotes |
|---|
e = time constant of
expiration Supported by departmental funding. Ventilators were provided by Fuji Respiratory Care Co., Ltd. and Mallinckrodt Japan Co., Ltd.
Received for publication September 16, 1999. Accepted for publication March 29, 2000.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
E. Miyoshi, Y. Fujino, A. Uchiyama, T. Mashimo, and M. Nishimura Effects of Gas Leak on Triggering Function, Humidification, and Inspiratory Oxygen Fraction During Noninvasive Positive Airway Pressure Ventilation Chest, November 1, 2005; 128(5): 3691 - 3698. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Farre, S. J. Lloyd-Owen, N. Ambrosino, G. Donaldson, J. Escarrabill, B. Fauroux, D. Robert, B. Schoenhofer, A. Simonds, and J. A. Wedzicha Quality control of equipment in home mechanical ventilation: a European survey Eur. Respir. J., July 1, 2005; 26(1): 86 - 94. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Nakamura, Y. Fujino, A. Uchiyama, T. Mashimo, and M. Nishimura Intrahospital Transport of Critically Ill Patients Using Ventilator With Patient-Triggering Function Chest, January 1, 2003; 123(1): 159 - 164. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |