(Chest. 2001;119:1123-1130.)
© 2001
American College of Chest Physicians
Effect of Size and Disease on Estimated Deposition of Drugs Administered Using Jet Nebulization in Children With Cystic Fibrosis*
Allan L. Coates, MD;
Paul D. Allen, M Eng;
Clair F. MacNeish, BSc, RRT;
Sharon L. Ho, BSc, RRT and
Larry C. Lands, MD, PhD
*
From the Division of Respiratory Medicine (Dr. Coates and Ms. Ho), Hospital for Sick Children, University of Toronto, Toronto; and Division of Respiratory Medicine (Mr. Allen, Ms. MacNeish, and Dr. Lands), Montreal Childrens Hospital McGill University, Montreal, Canada.
Correspondence to: Allan L. Coates, MD, Division of Respiratory Medicine, Hospital for Sick Children, 555 University Ave, Toronto, Ontario, Canada M5G 1X8; e-mail: allan.coates{at}sickkids.on.ca
 |
Abstract
|
|---|
Study objectives: To develop a model that quantified
the nebulizer output that was inhaled by subjects with cystic fibrosis
(CF) in order to predict the amount of drug likely to enter the upper
airway contained in particles small enough to be deposited in the lower
respiratory tract of individual patients.
Design:
Forty-three patients (age, 6 to 18 years) with CF, with
FEV1 of 26 to 124% of predicted, breathed through a
nebulizer circuit with a pneumotachograph in place at the distal end.
Algorithms were developed from the measured flows through the
pneumotachograph, allowing partitioning of inspiration into undiluted
aerosol and fresh gas. In order to validate the algorithms, argon was
added to the nebulizing gas flow and then its concentration was
analyzed at the mouth by mass spectrometry.
Results:
Predictions of the concentration of argon at the mouth were concordant
with that measured by mass spectrometry, thus validating the model.
Combining data from the model with in vitro nebulizer
performance data, predictions for estimates for lung deposition for
individuals were possible. Total estimate was independent of patient
size or FEV1. The respiratory duty cycle was 0.44 ± 0.05
(mean ± SD) and correlated (r = 0.91, p < 0.001)
with estimated deposition and minute ventilation
(r = 0.60, p < 0.01). However, when expressed in
milligrams per kilogram of body weight, the estimated deposition in
smaller children was fourfold higher than in larger children.
Conclusions: If the effect of patient size and pattern of
breathing on estimated drug deposition are not considered when
prescribing drugs given by nebulization, the result may be overdosing
younger children, underdosing older children, or
both.
Key Words: cystic fibrosis inhaled antibiotics nebulization
 |
Introduction
|
|---|
The
administration of nebulized medication for the treatment of lung
disease in children has a number of attractions. It requires minimal
cooperation, and it delivers medication directly to the organ of
concern. Furthermore, medications that may not be formulated
specifically for inhalation therapy can be administered this way in an
"off-label" usage. Of particular interest has been the use of
inhaled antibiotics for the treatment of cystic fibrosis
(CF).1
2
3
4
5
However, there are disadvantages of using this
method. Less than 10% of the initial dose of medication that is placed
in the nebulizer is likely to be deposited in the lung; of that amount,
the fraction deposited is highly variable.6
7
Fortunately,
the inhaled medications currently used are relatively nontoxic, so that
huge doses can be placed in a nebulizer in order to ensure adequate
deposition to achieve a therapeutic effect.1
However, for
economic reasons as well as the concern that future medications may not
have this margin of safety, better methods of ensuring the delivery of
an appropriate dose that will provide clinical efficacy are desirable.
Scientific data to support the choice of pediatric drug dosages for
nebulization are frequently lacking. Doses for parenterally or
enterally administered medications are commonly based on the patients
body weight (milligrams per kilogram), and sometimes aerosolized
medications are also prescribed in this manner.8
9
At
other times, the same dose is given to all patients.1
2
10
In both these situations, the amount of drug that actually reaches the
targeted area of the lung will remain highly variable from patient to
patient. During aerosol administration using a conventional jet
nebulizer, the concentration of the aerosol contained in the inspired
gas will be determined by the output of the nebulizer. The aerosol
contained in each breath may be diluted by ambient air if the
patients inspiratory flow exceeds the flow driving the nebulizer
(
N). The aerosol will remain undiluted only if a
small childs peak inspiratory flow is less than the
N. Uniquely, under these circumstances alone, the
amount of medication that will be received for a given nebulizer will
be directly proportional to the minute ventilation
(
E) of the child, which is dependent on size or,
more specifically, metabolic demand.
Older (larger) children will have an inspiratory flow that well exceeds
the
N and will significantly dilute the aerosol with
ambient air. This phenomenon of less dilution in infants may account
for their seemingly increased sensitivity to histamine
inhalation.11
An eloquent description of the theory of
this process was put forward by Collis et al.12
Based on
radionuclear inhalational studies, the same group of
investigators13
suggested that the nebulizer drug dose be
dramatically reduced for children less than a year and a half of age
compared to that for older children. However, while the concept is
known, to date there has never been a quantification of the influence
of size, breathing pattern, and the degree of lung disease on estimated
pulmonary deposition of nebulized medications. The purposes of this
study were as follows: (1) to develop a model that would allow the
prediction of the inhaled mass (the mass of aerosol that would enter
the airway14
15
); and (2) to evaluate the effect of the
individual pattern of breathing on the quantity of tobramycin carried
in particles small enough to deposit in the lower respiratory tract of
children breathing through a mouthpiece, using an unvented, good
quality, disposable jet nebulizer.
 |
Theory
|
|---|
When the Bain anesthetic circuit was first
introduced,16
there was concern that rebreathing expired
gas would lead to hypercapnia. To investigate this, in 1983,
Meakin and Coates17
developed a model that allowed
partitioning of each inspiration into the volume of rebreathed gas, and
the volume of fresh anesthetic gas. They justified their model by using
the respiratory pattern, measured by a pneumotachograph placed at the
distal end of the circuit, to predict the concentration of
CO2 that would appear at the endotracheal tube.
Concordance between the predicted and the directly measured values
validated the model. For jet nebulization, this model was adapted to
determine the fraction of each breath that would contain undiluted
aerosol, which, coupled with the output of the nebulizer, could be used
to calculate inhaled mass. The fraction of the inhaled mass expected to
deposit in the lungs of that specific CF patient could then be
calculated from the particle size distribution.
The model was based on the mechanics of a patient breathing through a
mouthpiece/T-piece connection to an unvented jet nebulizer (Hudson 1730
Updraft II; Hudson; Temecula, CA). A pneumotachograph was placed on the
distal end of the T-piece (Fig 1
). The nebulizer was run without solution (to prevent wetting the
pneumotachograph) with a gas containing 5% Ar. A respiratory mass
spectrometer (MS) [model MGA-1100; Marquette; Milwaukee, WI] was used
to sample the concentration of Ar at the mouth. In order to align the
signals from the pneumotachograph with those from the MS accurately,
with respect to time, the time of each digitized flow data point was
shifted to account for the delay between the point where gas was
sampled at the mouthpiece, and the time of digitization of the
electrical output from the MS. If the nebulizer output entering the
mouth was undiluted, then the measured Ar concentration would be 5%.
It would be less with dilution by entrained room air. Because the
N was known from the flow through the
pneumotachograph (
pnt[t]), the flow to the patient
(
p[t]) could be calculated:
P(t) =
PNT(t) +
N.
The
pnt(t) was calibrated, digitized, and sampled at
200 Hz so that each 5-ms increment of
p(t)
represented a tiny volume made up of a fraction of aerosol and a
fraction of ambient air.
Figure 2
shows the sources of
p(t). The first part of
inspiration (area 1, t0 to t1) was undiluted
aerosol when
p(t) was less than
N, and the second part (area 2) was the undiluted
aerosol that filled the dead space of the apparatus when
P(t) was less than
N. After this
(area 3, t1 to t2), each part of the
inspiratory flow multiplied by the 5-ms increment of time, represented
an increment of volume that contained a fraction of undiluted aerosol
(
N x 5 ms) and ambient air
(
p[t] -
N) x 5 ms) until
such time (area 4, t2 to inspiratory time
[TI]) when
p(t) was again less than
N and inspiration contained undiluted aerosol. The
tiny areas of aerosol and ambient air in each 5-ms epoch could be
summed (digitally integrated) to give the respective volumes of
undiluted aerosol and ambient air per breath. If the equations (see Appendix) used to calculate the fraction of aerosol in each breath are
correct, then it would be possible to calculate the concentration of Ar
that would appear at the mouth. Agreement with the predicted and the
measured concentration would justify the accuracy of the model.

View larger version (24K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 2. Schematic diagram of an inspiration. The stippled
area is the amount of flow of undiluted aerosol that enters the mouth
and the white area under the curve is the ambient air that is inspired.
Area 1 from t0 to t1 is the undiluted aerosol
that enters when nebulizing flow is greater than inspiratory flow. Area
2 is the apparatus dead space that contains the undiluted aerosol that
filled the space between t0 and t1. Area 3 is
the nebulizer output between t1 and t2, and
area 4 is the region between t2 and TI when
N is again greater than the patients inspiratory
flow.
|
|
 |
Materials and Methods
|
|---|
The subjects had a diagnosis of CF18
and were being
followed up in the CF clinic of the Montreal Childrens Hospital. They
were selected for their willingness and the ability to undergo
pulmonary function testing. The Ethics Review Board of the institution
had approved the study, and written informed consent was obtained from
either the parent or guardian, or the patient, depending upon the age
of the patient. Spirometry was performed according to American Thoracic
Society guidelines.19
Lung volumes were obtained by
plethysmography.20
The FEV1,
expressed as a percent predicted from height,21
was used
as the index of air flow limitation, and the residual volume to total
lung capacity ratio (RV/TLC) was used as an index of gas trapping.
The subject was then seated comfortably, familiarized with the
nebulizer, and asked to breathe quietly and regularly through the
device with nose clips in place.
N was set at 6
L/min using a flowmeter that was specifically calibrated for the
nebulizer to avoid problems of inadequate back-pressure compensation of
the flowmeter.22
23
24
For analysis, the operator selected
epochs that contained 10 consecutive breaths, indistinguishable from
each other (by inspection). Data from patients who could not breathe
regularly were not analyzed.
Data analysis was performed, first by comparing the calculated
concentration of Ar at the mouth to that measured by MS, to validate
the model using the Bland and Altman25
limits of
agreement. Once the model was validated, the fraction of inspired
ventilation that contained undiluted aerosol was calculated from the
combination of the data provided from the model, coupled with the
measured
E of the patient. The rate of output for an
80-mg dose of tobramycin with 2 mL normal saline solution, using the
Hudson 1730 and 6 L/min driving flow had been previously
reported.24
Similarly, total aerosol output and output
contained within the respirable fraction (RF)22
26
have
been reported for this nebulizer and are 51 mg total with 21 mg in the
RF, with a mass median diameter (± 1 geometric SD) of 5.7 ± 0.26
µm. The inhaled mass per breath can be calculated by combining the
rate of output,26
and data from the model. This, when
multiplied by the RF,22
26
gives an estimate of pulmonary
deposition per breath. The ratio of the estimated deposition per breath
to the total output per breath is essentially the in vivo
efficiency of the device. By multiplying this ratio by the total output
of the device,22
the amount (in milligrams) of tobramycin
estimated to deposit in the lungs of each patient per nebulization
session could be calculated. The definition of the RF was the fraction
of aerosol contained in particles
5 µg that, if inhaled, are
expected to deposit below the vocal cords.27
28
29
Inherent
in this definition is the assumption that the RF will be the same for
children with disease as it is for normal adults, although it is
recognized that there are theoretical reasons why this may not be the
case.30
31
Multiple regression analysis was used to explore the factors that would
correlate with the estimated pulmonary deposition, including the
pattern of breathing, the
E, height, and the degree
of lung disease (FEV1 and RV/TLC).
 |
Results
|
|---|
Forty-three patients from 6 to 18 years of age participated in the
study. They ranged in height from 108 to 173 cm, and their
FEV1 values ranged from 26 to 124% predicted.
Figure 3
shows the Bland and Altman25
plot of limits of agreement
for the measured signal from the MS, and the calculated concentration
for Ar. The bias ± 2 SDs indicates that the mathematical model
successfully defined the pattern of ventilation and the dilution of the
aerosol with ambient air. Based on the validity of the model, the
subsequent analysis calculated the estimated deposition of aerosol for
each individual patient.

View larger version (17K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 3. Bland and Altman25
limits of
agreement between the difference in the Ar concentration recorded at
the mouth and that predicted from the model, plotted against the
recorded concentration (which was considered the true value). The 95%
confidence limits are also shown.
|
|
From the model, 40 ± 10% of the total output of the nebulizer
entered the upper airway; based on the output within the
RF,22
the estimate of lower-airway deposition would be
8.6 ± 1.0 mg. This value was independent of the subjects size or
degree of lung disease (FEV1 percent
predicted21
and RV/TLC). However, it did correlate closely
with respiratory pattern as indicated by the ratio of the
TI to total respiratory time (respiratory duty cycle
[TI/TTOT]; r = 0.91,
p < 0.001; Fig 4
). TI/TTOT ranged from 0.36
to 0.64 with a mean of 0.44 ± 0.05. The
E also correlated with estimated
deposition, but more loosely (r = 0.60, p < 0.01). The
combination of TI/TTOT and
E regressed against estimated deposition
was highly predictive for individual patients (r = 0.93,
p < 0.0001), with TI/TTOT being
more than four times as important in explaining the variation than
E. The estimated deposition, expressed as
milligrams of tobramycin per kilogram of body weight, correlated
closely with height (r = - 0.91, p < 0.001; Fig 5
). On a milligram per kilogram basis, the estimated drug deposition in
smaller children was fourfold that of the taller children. With regard
to disease severity, neither FEV1 (percent
predicted) nor RV/TLC correlated with the estimated drug deposition by
linear regression. When using multiple linear regression along with
height, neither contributed significantly to the relationship seen with
height alone (Fig 5)
. While the distribution of disease severity was
skewed to the more mild degree of pulmonary impairment, there was no
significant relationship between age and FEV1
(percent predicted) with the majority of patients over all ages having
FEV1 = 75% predicted. Only one patient had an
FEV1 that was < 40% predicted. There was a
significant but loose correlation between estimated deposition
normalized for weight and tidal volume (VT;
r = - 0.42, p < 0.05), but not
E or
TI/TTOT. In other words,
the longer the duration of inspiration in relation to the total cycle
time, and the greater the
E, the greater
the estimate of total deposition. Total estimated deposition, however,
was not related to deposition per kilogram of body weight. Only the
VT and height correlated negatively with the
estimated deposition on a milligram per kilogram of weight basis.
 |
Discussion
|
|---|
In this study, we have developed a model that allows the
estimation of the inhaled mass in children with CF, over a wide range
of size and disease severity. The inhaled mass when coupled with RF
gives an estimate of pulmonary deposition of inhaled tobramycin. There
is a striking relationship between estimated deposition normalized for
body weight and the height of the child. This suggests that standard
doses of medications administered by unvented jet nebulization may
result in overdosing small children, underdosing larger children, or
both. While disease severity appeared to play a small role in
determining the amount of tobramycin that would be inhaled, it has been
shown to have a striking effect on the intrapulmonary
distribution.6
32
However, the lack of effect of disease
severity on the calculated inhaled mass may be partially attributed to
the lack of patients with very severe lung disease in this study.
The Bland and Altman25
plot (Fig 3)
shows that the model
accurately predicts the concentration of Ar that appears at the mouth.
This model is not perfect because there is a tendency to slightly
underestimate the inhaled Ar concentration in the smaller children who
have the highest concentration of Ar (Fig 3)
. Because these children
have small VTs, any error in the estimation of the
apparatus dead space will cause a proportionately larger error in the
calculated concentration of Ar. The single-value entry for apparatus
dead space assumes plug flow, in other words, no mixing of aerosol
output and ambient (or expired) air within the dead space.
Hence, it is assumed that the dead space will be filled with undiluted
aerosol during the early phase of inspiration (Fig 2
: t0 to
t1). Any discrepancy between these assumptions in reality
would give rise to the greatest errors when VT is small and
the dilution of the inhaled Ar is least. Another factor is the time
delay for the MS to sample and read the Ar concentration compared to
any signal delay from the pneumotachograph. Again, little children with
relatively rapid respiratory rates would magnify the effect on any
inaccuracy of the time delay, compared to the slower respiratory
pattern of the older children. Despite these constraints, the errors
because of such factors appear, from the limits of agreement (Fig 3)
,
to be small. In order to calculate the concentration of Ar at the
mouth, the model must be able to accurately partition the inspired gas
into the component of undiluted aerosol and the component of fresh
ambient air. Given that this partitioning is the basis for the
calculation of the estimated deposition (Appendix), similar accuracy
can be obtained for estimated deposition calculations.
Given the theoretical results of Collis et al,12
the
results presented here follow the expected pattern. However, Collis et
al12
used patterns of breathing derived from other studies
to illustrate their point, whereas the present study used actual
patients with CF breathing from the same type of nebulizer that they
use in the clinical setting. Furthermore, unlike Collis et
al,12
these results provide a quantitative analysis of the
factors that would influence estimated pulmonary deposition of
tobramycin, and hence can act as a specific guide to individualizing
doses of nebulized medication. The lack of a significant relationship
between size of the child and estimated deposition was because of the
predominant role played by TI/TTOT, which is
independent of size, and the less significant role played by
E, which is related to size. For unvented
nebulizers, where the output is constant and independent of the
respiratory cycle, drug is inhaled during inspiration and lost to the
environment during expiration, explaining the very significant
relationship between estimated deposition and
TI/TTOT. Where
E does play a
role is in the smaller children. With lower peak inspiratory flows, and
hence a longer period where the inspiratory flow, after the onset of
inspiration, is less than
N, more aerosol is lost to
the environment during inspiration. This happens at the start and end
of inspiration (Fig 2)
. For larger children with a high
E and the accompanying high inspiratory flow,
inspiratory flow is less than
N for very short
periods of time, so little aerosol is lost during inspiration. With
estimate of deposition independent of size, it is not surprising that
there is as much as a fourfold difference in estimated deposition
(expressed in terms of milligrams drug per kilogram of body weight)
between the smaller and the larger children. Because clinical
studies2
have shown both beneficial effects of 80 mg of
tobramycin administered three times daily2
and a lack of
toxic effects of doses of tobramycin several-fold higher,1
it is clear that the drug is potent and safe when given by inhalation
(presumably because little is absorbed through the pulmonary
epithelium, and that which is expectorated and swallowed has little GI
absorption). On the other hand, newer treatments such as gene therapy
using either liposome or viral vectors may not be as innocuous and may
have a significantly narrower therapeutic index. In such situations, it
will be important to be able to estimate the dose required to deliver a
prescribed amount of drug by aerosol, based on body weight. If the
output of an unvented jet nebulizer driven at 6 L/min is known, the
data in Figure 4
can be used to derive a rough estimate of the
pulmonary deposition. If further accuracy is required, it will be
necessary to make the appropriate measurement on an individual patient
using a model similar to that described.
There are two types of limitations and cautions in the interpretation
of these data, as there are with all in vitro data, which
are assumed to represent the real events in vivo. First,
other than the pattern of breathing, the model does not take any other
factor concerning disease severity or size into consideration. Of
particular concern is the extrapolation of the definition of the RF
that was established primarily in studies of normal adults rather than
in children with CF. The primary mechanisms of deposition of inhaled
particles in the airway are by inertial impaction and
sedimentation,33
with Brownian diffusion playing lesser
role. Inertial impaction can be roughly defined as the lack of ability
to follow the air carrying the particle around a curve, such as in the
posterior pharynx or bifurcation of a bronchus. Hence, it is quite
possible that differences in anatomy and size of the upper airway as
well as velocity of inspiratory flow may effect the RF. With regard to
sedimentation, retention time and airway size (eg, distance
for the particle to fall) may also play a role31
and would
be expected to differ according to size and disease
state.34
Furthermore, airway lining liquid viscosity, a
factor frequently abnormal in CF, has also been shown to play a role in
determining deposition.35
However, errors because of the
inappropriate definition of RF cannot be excessively large because the
current study used a very efficient unvented nebulizer and found a mean
estimated deposition of the nebulizer charge of 10.6%. This compares
to values between 5% and 10% found in deposition
studies6
32
36
in children with CF. Part of this
discrepancy may have occurred because the present model estimated total
deposition below the cords, while many studies6
32
36
exclude the trachea from the regions of interest. Compared to frequent
clinical practice, in this study, all subjects breathed through a
mouthpiece with a nose clip in place. It is clear that the nose is very
effective in removing inhaled particulate matter, and values
approximately half of that calculated here would be expected in older
children when breathing through a face mask.13
A second issue is equipment. The nebulizer used in this study is an
unvented disposable device with an output not influenced by any
respiratory effort of the subject.26
It is clear that
reusable vented or "breath-enhanced" nebulizers will increase their
output of tobramycin in relation to the inspiratory flow drawn through
them.26
Intuitively, this would increase output for
increasing
E, and this would be expected to reduce
the discrepancies seen between the smaller and the larger children (who
have larger values for
E). Hence, the
breath-enhanced nebulizers are becoming popular for home use, but the
much less expensive disposable unvented devices are commonly used in a
hospital or clinic environment. The values calculated for estimated
deposition are based on measurement of the actual drug
output37
(not just weight change) and particle size
distribution of this drug output determined by laser diffraction
techniques (which do not dry the aerosol cloud37
38
) for
this particular nebulizer charged with 80 mg of tobramycin in a 4-mL
volume. There are wide variations of nebulizer
performance,39
40
so to apply this information to any
other nebulizer driven at 6 L/min, it would be necessary to
characterize the output of the nebulizer while nebulizing the
particular drug of interest with the intended fill
volume.22
For nebulizing flows different from 6 L/min, the
analysis would have to be repeated as the dilution of aerosol with
ambient air depends on the patients inspiratory flow in relation to
the
N. The data for the parameters of the Hudson
1730 nebulizer are derived specifically for a
N of 6
L/min measured at the exit of the nebulizer, rather than what is
indicated by the flowmeter. While standard flowmeters are stated by the
manufacturer to be "back-pressure compensated," in reality, this is
not the case, and a specific calibration curve is required for each
brand and type of nebulizer to be evaluated.22
26
40
There
are other techniques to calculate inhaled mass, the most popular of
which is to place a filter between the patient and the nebulizer and
then quantify the drug on the filter.14
Chemical analysis
of the drug on filter is not simple for drugs like tobramycin that do
not contain a chromophore for easy ultraviolet spectrophotometric
analysis.41
Alternative techniques41
are
possible but labor intensive and would result in fewer subjects being
studied. The present model is sufficiently automated to allow many
patients to be tested. With the limitations of the assumption of the RF
as mentioned earlier, before any estimate of pulmonary deposition, the
inhaled mass must be known. The agreement with predicted and measured
concentration of Ar seen in this study would suggest that the model is
capable of predicting inhaled mass. If more data become available to
improve the definition of RF in this patient group, estimates of
deposition could easily be recalculated from the inhaled mass.
In conclusion, we have developed a model that can be used to calculate
the inhaled mass allowing the estimation of deposition of aerosol below
the vocal cords in children with CF. The results indicate that smaller
children breathing through a mouthpiece from an unvented jet nebulizer
are likely to receive four times the dose of tobramycin on a per
kilogram of weight basis, compared to older and bigger children. The
notion that the dose of drugs to be given by aerosol does not have to
be individualized for the size of the child is applicable only when
giving relatively large doses of nontoxic medication. Taking into
consideration the limitations discussed above, this approach can be
extended to other therapeutic agents and other devices to estimate
pulmonary deposition of an inhaled agent.
 |
Appendix 1
|
|---|
Volume areas 1 to 4 (Fig 2)
, when summed together, represent the
total nebulizer output (aerosol). The equations for each subvolume are
as follows:
 | (1) |
 | (2) |
 | (3) |
 | (4) |
Also, note that at times t1 and t2,
N =
p(t1) and
p(t2). Now,
% Aerosol
 | (1A) |
Defining
pnt(t) as positive for
inspiratory flow when
pnt(t) >
N, then:
or:
then, from (1a) we get:
 | (1B) |
 | (1C) |
 | (1D) |
In order to determine the Ar concentration at the mouth at
discrete times during the VT, consider the following three
equations.
At some time between t0 and t1:
 | (2A) |
At some time between t1 and
t2:
 | (2B) |
 | (2C) |
At some time between t1 and TI:
 | (2D) |
where FAra = concentration of Ar in ambient
air.
Putting (2a) and (2d) together over the whole excursion of
 | (2E) |
or using an MS (and thereby measuring at the sample port):
 | (5) |
Now if theory and practice agree, values for (2e) should be the
same as (3).
 |
Footnotes
|
|---|
Abbreviations:
CF = cystic fibrosis; MS = mass spectrometer; RF = respirable
fraction; RV/TLC = residual volume/total lung capacity ratio;
TI = inspiratory time;
TI/TTOT = respiratory duty cycle;
VT = tidal volume;
E = minute
ventilation;
N = flow driving the nebulizer;
p(t) flow to the patient;
pnt(t) = flow through the pneumotachograph
Supported by the Canadian Cystic Fibrosis Foundation.
Received for publication April 20, 2000.
Accepted for publication November 3, 2000.
 |
References
|
|---|
-
Ramsey, BW, Dorkin, HL, Eisenberg, JD, et al (1993) Efficacy of aerosolized tobramycin in patients with cystic fibrosis. N Engl J Med 328,1740-1746[Abstract/Free Full Text]
-
MacLusky, IB, Gold, R, Corey, ML, et al (1989) Long-term effects of inhaled tobramycin in patients with cystic fibrosis colonized with Pseudomonas aeruginosa. Pediatr Pulmonol 7,42-48[ISI][Medline]
-
Stead, RJ, Hodson, ME, Batten, JC (1987) Inhaled ceftazadime compared with gentamicin and carbenicillin in older patients with cystic fibrosis infected with Pseudomonas aeruginosa. Br J Dis Chest 81,272-279[CrossRef][ISI][Medline]
-
Littlewood, JM, Miller, MG, Ghoneim, AT, et al (1985) Nebulized colomycin for early Pseudomonas colonization in cystic fibrosis [letter]. Lancet 1,865[ISI][Medline]
-
Hodson, ME, Penketh, ARL, Batten, JC (1981) Aerosol carbenicillin and gentamicin treatment of Pseudomonas aeruginosa infection in patients with cystic fibrosis. Lancet 2,1137-1139[ISI][Medline]
-
Ilowite, JS, Gorvoy, JD, Smaldone, GC (1987) Quantitative deposition of aerosolized gentamycin in cystic fibrosis. Am Rev Respir Dis 136,1445-1449[ISI][Medline]
-
Alderson, PO, Secker-Walker, RH, Strominger, DB, et al (1974) Pulmonary deposition of aerosols in children with cystic fibrosis. J Pediatr 84,479-484[CrossRef][ISI][Medline]
-
Schuh, S, Johnson, DW, Callahan, S, et al (1995) Efficacy of frequent nebulized ipratropium bromide added to frequent high-dose albuterol therapy in severe asthma. J Pediatr 126,639-645[CrossRef][ISI][Medline]
-
Kerem, E, Levison, H, Schuh, S, et al (1993) Efficacy of albuteral administered by nebulizer versus spacer device in children with acute asthma. J Pediatr 123,313-317[CrossRef][ISI][Medline]
-
Steinkamp, G, Tummler, B, Gappa, M, et al (1989) Long-term tobramycin aerosol therapy in cystic fibrosis. Pediatr Pulmonol 6,91-98[ISI][Medline]
-
LeSouëf, PN, Geelhoel, GC, Turner, DJ, et al (1989) Response of the normal infants to inhaled histamine. Am Rev Respir Dis 139,62-66[ISI][Medline]
-
Collis, GG, Cole, CH, LeSouëf, PN (1990) Dilution of nebulised aerosols by air entrainment in children. Lancet 336,341-343[CrossRef][ISI][Medline]
-
Chua, HL, Collis, GG, Newbury, AM, et al (1994) The influence of age on aerosol deposition in children with cystic fibrosis. Eur Respir J 7,2185-2191[Abstract]
-
Smaldone, GC (1996) Aerosolized drug delivery in the 90s. Chest 110,316-317[Free Full Text]
-
Smaldone, GC (1994) Drug delivery by nebulization: "reality testing." J Aerosol Med 7,213-216[ISI][Medline]
-
Bain, JA, Spoerel, WE (1972) A streamlined anesthetic system. Can Anesth Soc J 25,30-36
-
Meakin, G, Coates, AL (1983) An evaluation of rebreathing with the Bain circuit system during anesthesia with spontaneous respiration. Br J Anaesth 55,487-496[Abstract/Free Full Text]
-
. for the Cystic Fibrosis Foundation Consensus PanelRosenstein, BJ, Cutting, GR (1998) The diagnosis of cystic fibrosis: a consensus statement. J Pediatr 132,589-595[CrossRef][ISI][Medline]
-
. American Thoracic Society. (1995) Standardization of spirometry: 1994 update. Am J Respir Crit Care Med 152,1107-1136[ISI][Medline]
-
Coates, AL, Peslin, R, Rodenstein, DO, et al (1997) Measurement of lung volumes by plethysmography. Eur Respir J 10,1415-1427[CrossRef][ISI][Medline]
-
Knudson, RJ, Lebowitz, MD, Holberg, CJ, et al (1983) Changes in the normal maximal expiratory flow-volume curve with growth and aging. Am Rev Respir Dis 127,725-734[ISI][Medline]
-
Coates, AL, MacNeish, CF, Meisner, D, et al (1997) The choice of jet nebulizer, nebulizing flow, and the addition of albuterol affects the output of tobramycin aerosols. Chest 111,1206-1212[Abstract/Free Full Text]
-
MacNeish, CF, Meisner, D, Thibert, R, et al (1997) A comparison of pulmonary availability between Ventolin (albuterol) nebules and Ventolin (albuterol) Respirator Solution. Chest 111,204-208[Abstract/Free Full Text]
-
Coates, AL, MacNeish, CF, Lands, LC, et al (1998) Factors influencing the rate of drug output during the course of wet nebulization. J Aerosol Med 11,101-111
-
Bland, JM, Altman, DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 8,307-310
-
Coates, AL, MacNeish, CF, Lands, LC, et al (1998) A comparison of the availability of tobramycin for inhalation from vented vs unvented nebulizers. Chest 113,951-956[Abstract/Free Full Text]
-
Gerrity, TM, Lee, PS, Hass, FJ, et al (1979) Calculated deposition of inhaled particles in the airway generation of normal subjects. J Appl Physiol 47,867-873[Abstract/Free Full Text]
-
Newman, SP, Clarke, SW (1983) Therapeutic aerosols: 1. Physical and practical considerations. Thorax 38,881-886[ISI][Medline]
-
Newman, SP, Pellow, PGD, Clay, MM, et al (1985) Evaluation of jet nebulisers for use with gentamycin solution. Thorax 40,671-676[Abstract]
-
Taulbee, DB, Yu, CP (1975) A theory of aerosol deposition in the human respiratory tract. J Appl Physiol 38,77-85[Abstract/Free Full Text]
-
Goldberg, IS, Lourenço, RV (1973) Deposition of aerosols in pulmonary disease. Arch Intern Med 131,88-91[CrossRef][ISI][Medline]
-
Mukhopadhyay, S, Staddon, GE, Eastman, C, et al (1994) The quantitative distribution of nebulized antibiotic in the lung in cystic fibrosis. Respir Med 88,203-211[ISI][Medline]
-
Yeh, H, Phalen, RF, Raabe, OG (1976) Factors influencing the deposition of inhaled particles. Environ Health Perspect 15,147-152[ISI][Medline]
-
Kim, CS, Lewars, GA, Sackner, MA (1988) Measurement of total lung aerosol deposition as an index of lung abnormality. J Appl Physiol 64,1527-1536[Abstract/Free Full Text]
-
Kim, CS, Abraham, WM, Chapman, GA, et al (1985) Influence of two-phase gas-liquid interaction on aerosol deposition in airways. Am Rev Respir Dis 131,618-623[ISI][Medline]
-
Thomas, SHL, ODoherty, MJ, Graham, A, et al (1991) Pulmonary deposition of nebulised amiloride in cystic fibrosis: comparison of two nebulisers. Thorax 46,717-721[Abstract]
-
Dennis, JH (1995) Standardisation in drug nebulisers: BS7711; part 3. J Aerosol Med 8,313-315
-
Clark, AR (1995) The use of laser diffraction for the evaluation of the aerosol clouds generated by medical nebulizers. Int J Pharm 115,69-78[CrossRef]
-
Hollie, MC, Malone, RA, Skufea, RM, et al (1991) Extreme variability in aerosol output of the DeVilbiss 646 nebulizer. Chest 100,1339-1344[Abstract/Free Full Text]
-
Ho, SL, Coates, AL (1999) The effect of dead volume on the efficiency and the cost to deliver medications in cystic fibrosis with four disposable nebulizers. Can Respir J 6,253-260[Medline]
-
Kwong, E, MacNeish, CF, Meisner, D, et al (1998) The use of osmometry as a means of determining changes in drug concentration during jet nebulization. J Aerosol Med 11,89-100
This article has been cited by other articles:

|
 |

|
 |
 
M. R. Miller, J. Hankinson, V. Brusasco, F. Burgos, R. Casaburi, A. Coates, R. Crapo, P. Enright, C. P. M. van der Grinten, P. Gustafsson, et al.
Standardisation of spirometry
Eur. Respir. J.,
August 1, 2005;
26(2):
319 - 338.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. Reychler, A. Keyeux, C. Cremers, C. Veriter, D. O. Rodenstein, and G. Liistro
Comparison of Lung Deposition in Two Types of Nebulization: Intrapulmonary Percussive Ventilation vs Jet Nebulization
Chest,
February 1, 2004;
125(2):
502 - 508.
[Abstract]
[Full Text]
[PDF]
|
 |
|