(Chest. 2001;119:1210-1221.)
© 2001
American College of Chest Physicians
Current Status of the Implantable Cardioverter-Defibrillator*
Michael H. Gollob, MD and
John J. Seger, MD
*
From the Section of Cardiology (Dr. Gollob), Baylor College of Medicine, Houston, TX; Division of Cardiology (Dr. Seger), Texas Heart Institute, Houston, TX.
Correspondence to: Michael H. Gollob, MD, Section of Cardiology, RM 507D, Baylor College of Medicine, One Baylor Plaza, Houston, TX 77030; e-mail: mgollob{at}bcm.tmc.edu
 |
Abstract
|
|---|
Clinical trials have established the superiority of the implantable
cardioverter-defibrillator (ICD) over antiarrhythmic drug
therapy in survivors of sudden cardiac death and in high-risk patients
with coronary artery disease. The ICD has evolved to overcome the
limitation of earlier devices that required thoracotomy for
implantation and were fraught with inappropriate shock delivery.
Current ICDs are implanted in a similar manner to cardiac pacemakers
and incorporate sophisticated rhythm-discrimination algorithms to
prevent inappropriate therapy. Managing the patient with an ICD
requires an understanding of the multiprogrammable features of modern
devices. Drug interactions and potential sources of electromagnetic
interference may adversely affect ICD function. Driving restrictions
may be necessary under certain conditions. The cost-effectiveness of
ICD therapy appears favorable, given the marked survival benefit seen
in randomized trials relative to antiarrhythmic drug treatment. The
growing number of ICD recipients necessitates an understanding of the
specialized features of the modern ICD and the role of device therapy
in clinical practice.
Key Words: implantable cardioverter-defibrillator sudden cardiac death
 |
Introduction
|
|---|
In
the United States alone, sudden cardiac death (SCD) accounts for an
estimated 350,000 lives lost annually.1
The
majority of these events occur outside the hospital setting and
are associated with consequential delays in implementing external
defibrillation. The ability of the implantable
cardioverter-defibrillator (ICD) to provide therapy within 5 to 15
s of arrhythmia detection allows for defibrillation success rates
approaching 100%.
Numerous clinical trials have established the superiority of the ICD
over drug therapy in reducing mortality rates for survivors of cardiac
arrest or patients at high risk for SCD.2
3
4
5
6
7
On the basis
of these studies, the ICD is now recommended as the treatment of choice
in survivors of SCD and in patients with symptomatic, sustained
ventricular arrhythmias.8
Current ICD implantation rates
exceed 30,000 per year and will continue to grow as clinical
indications evolve.9
The majority of physicians will be
involved in the care of ICD recipients. In this review, we provide the
generalist with an overview of ICD function and the management of ICD
patients in clinical practice.
 |
The Modern ICD
|
|---|
Early Beginnings
The first human ICD implant occurred in 1980.10
Approval for general use was granted by the US Food and Drug
Administration in 1985. These early devices consisted of a large pulse
generator (PG) (Fig 1
) and patch electrodes for defibrillation placed directly on the heart
or pericardium. Epicardial screw-in leads also were placed for rate
sensing (Fig 2
). Implant procedures required general anesthesia, thoracotomy, were
associated with longer hospital stays, and had a perioperative
mortality rate in the range of 4%.11

View larger version (142K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 2. An early ICD system showing epicardial patch
electrodes used for defibrillation and epicardial screw-in leads for
rate sensing. The large PG was placed in the abdominal wall.
|
|
In addition to the limitations imposed by size, early devices offered
few options for patient-specific programming. The ICD was ordered by
the physician specifying a detection heart rate and delivered by the
manufacturer to perform defibrillation should the patients
ventricular rate exceed this value. No programming options existed for
changing the prespecified detection rate. Therefore, should an
antiarrhythmic agent be required and have the effect of slowing
ventricular tachycardia (VT) below the preset detection rate, ICD
therapy would not be delivered. A further limitation of early devices
was the inability to discriminate rapid ventricular rhythms of
supraventricular origin from ventricular-based rhythms, leading to a
high incidence of inappropriate shocks. Analysis of stored electrograms
(EGMs) in later device models confirmed a rate of inappropriate shocks
in the range of 25 to 40%.12
13
14
15
Advances in ICD lead systems and defibrillation waveforms allowed for
successful transvenous, pectoral ICD implants (Fig 3
). ICD systems are now implanted in a manner similar to cardiac
pacemakers and without the need for general anesthesia. Perioperative
mortality is < 1%.16
Complications such as infection,
pneumothorax, pericardial tamponade, and pocket hematoma occur at rates
similar to those seen with pacemaker implantation
(< 3%).17
18
As ICD system hardware has evolved, so too
has the internal circuitry responsible for the specialized features of
the device. The modern ICD may now be programmed to detect several
specified tachycardia zones, with the alternative of less
aggressive antitachycardia pacing (ATP) therapy for slower,
hemodynamically stable VT. Further, sophisticated
rhythm-discrimination algorithms now exist to discern VT from rapid
supraventricular arrhythmias. This has resulted in the decline of the
inappropriate shock rate to < 5%.19
20

View larger version (21K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 3. Evolving ICD systems. Left,
A: the earliest ICD system, illustrating the epicardial
patch electrodes used for defibrillation and epicardial leads used for
rate sensing. This system required thoracotomy for implantation.
Middle, B: the development of leads
integrating both sensing and shocking elements allowed for
nonthoracotomy implants. PG size remained too large for routine
pectoral implantation. Right, C: the
modern ICD utilizes the PG as an active can in the
defibrillation circuit and is of sufficiently small size for pectoral
implantation.
|
|
ICD Rhythm Discrimination: Rapid Supraventricular Tachycardia vs
Ventricular Arrhythmia
Current ICDs allow for the programming of at least three
tachyarrhythmia detection zones. The highest rate zone is referred to
as the ventricular fibrillation (VF) zone. In the VF zone, the risk of
sudden hemodynamic collapse mandates 100% sensitivity in detection.
This is achieved by employing an X/Y algorithm that requires a
certain proportion (eg, 12/16) of R-R intervals to be within
the programmed zone for detection. Additional slower VT zones may be
programmed for detection of more hemodynamically stable ventricular
arrhythmias. In contrast to the more rapid rate of the VF zone,
rate criterion as the sole method for detection in VT zones
results in a significant number of inappropriate shocks.
Most unnecessary therapies are due to sinus tachycardia or
atrial fibrillation with ventricular rates overlapping with programmed
VT detection zones.11
21
22
23
24
Therefore,
rhythm-discrimination algorithms have been developed to prevent
inappropriate therapy and to increase specificity for VT within a
programmed VT zone. Unfortunately, increasing specificity is at the
expense of diminishing sensitivity. Thus, all rhythm-discrimination
algorithms are programmable only for the presumed more hemodynamically
stable, slower VT zones.
The two most commonly employed rhythm-discrimination algorithms utilize
onset and stability reference criteria. The onset
algorithm monitors for abrupt changes in the R-R interval from beat to
beat. In the scenario of sudden-onset tachycardia within a VT zone,
should the R-R interval of the first two beats of tachycardia decrease
(reflecting an increased heart rate) by more than the programmed onset
criterion (eg, 40 ms), then the tachycardia is considered to
be ventricular in origin. The onset algorithm is intended to prevent
inappropriate therapy for sinus tachycardia rates that progress into
the slow VT zone. Thus, a gradual change in R-R interval entering a VT
zone, as is the case in sinus tachycardia, would not satisfy onset
criteria and would be considered supraventricular in origin, averting
therapy. The limitation of this feature is the potential failure to
detect VT that arises during exercise and lacks abrupt onset. The onset
criterion has been reported to underdetect 5 to 13% of
VTs.25
26
27
Stability refers to the R-R interval variability of the
detected tachycardia. Ventricular rates due to atrial fibrillation
would be expected to show a wide range of R-R intervals, as opposed to
monomorphic VT. A programmable R-R stability algorithm would detect
as VT any rhythm with R-R intervals varying less than the programmed
value, for example, 40 ms. A potential difficulty is polymorphic VT,
although this rhythm tends to be faster and detected within the VF
zone.
The availability of dual-chamber ICD systems with the capability of A-V
sequential pacing and sensing has allowed the design of newer
enhancement-detection algorithms. The sensing of atrial EGMs permits
comparison of atrial and ventricular rates. Rapid tachycardias with an
atrial rate more than the ventricular rate would appropriately be
detected as supraventricular in origin. Similarly, dissociation of
atrial and ventricular rates during tachycardia may be recognized,
accurately identifying the rhythm as VT.
The assessment of ventricular EGMs has been incorporated as another
method to differentiate VTs from supraventricular tachycardias. The EGM
width criterion is based on the assumption that ventricular-based
rhythms will have a wider EGM compared to EGMs derived from normal AV
conduction. While the addition of this criterion to onset and stability
significantly lowers inappropriate shocks, it is known that EGM width
can change significantly at high sinus rates and over
time.28
Morphology discrimination (MD; Ventritex;
Sunnyvale, CA) is a novel algorithm whereby the device stores a
template of the baseline (sinus) ventricular EGM morphology (Fig 4
). During tachycardia detection, each complex is compared to the
template, and the algorithm determines the percent match. Should the
tachyarrhythmia EGM complex be less than a programmable percent
match, therapy is initiated. Limitations of EGM morphology exist
for patients with underlying bundle branch block or rate-related bundle
branch block.

View larger version (14K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 4. Morphology discrimination (Ventritex). A stored
ventricular EGM is compared to the detected ventricular EGM with
alignment of the three largest contiguous peaks. A morphology score or
percent match is calculated. The tachycardia is assumed to be
ventricular in origin if the percent match is less than a programmable
value or if the initial complex alignment is not possible.
|
|
Most electrophysiologists will utilize rhythm-discrimination algorithms
individually or in combination, depending on the clinical history of
the patient. Although these features increase detection specificity and
lower inappropriate shock rates to < 5%, underdetection and
inhibition of therapy for true VT may occur. A programmable safety
feature known as sustained rate duration can apply therapy
if the heart rate remains in a VT zone over a programmed duration of
time, thereby returning sensitivity for true VT to near
100%.19
20
26
ICD Therapy
All ICDs use electrical defibrillation as the only therapeutic
option for heart rates detected in the VF zone. Multiple or
tiered therapeutic options are available for VT detection
zones. A sequence of therapy is programmable and is followed as needed
until an episode is terminated. These programmable options include ATP,
low-energy synchronized cardioversion, and defibrillation.
The objective of ATP is to terminate monomorphic VT promptly with
little discomfort to the patient. Delivery of ATP is applied in
successive paced beats at a rate faster than the tachycardia cycle
length. ATP therapy successfully terminates approximately 90% of
episodes of spontaneous VT.29
In addition, the use of this
therapy is associated with a statistically significant 36 to 28%
reduction of first ICD shocks over a 2-year follow-up
period.30
The risk of this therapy is the potential for
accelerating VT to VF, which may occur in up to 10% of
attempts.31
Therefore, tiered therapy must always include
the programming of back-up defibrillation.
Diagnostic Storage
Early ICDs stored little information, noting only the number of
device discharges. Current devices store and display extensive data. An
updated therapy history is provided on each interrogation. A
large number of episodes may be stored and their details viewed
individually. Specifics regarding the date/time, therapy delivered,
detection zone and criteria satisfied, and total length of the episode
are displayed. Intracardiac EGMs of episodes requiring therapy are
retrievable, including the onset segment. Marker annotations indicating
R-R intervals and the devices interpretation of the ongoing rhythm
are displayed. Simultaneous EGMs from atrial and ventricular leads may
be viewed, assisting the physician in determining the appropriateness
of the applied therapy (Fig 5
).

View larger version (77K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 5. Intracardiac EGMs. Top: the panel
illustrates the successful termination of VT by a single burst of ATP.
Middle: the panel demonstrates the electrical
cardioversion of VT. Bottom: the panel indicates a rapid
supraventricular tachycardia with ventricular rates detected in a VT
zone. Rhythm-discrimination algorithms correctly identified this as
atrial fibrillation/flutter (AF) and prevented inappropriate therapy.
AR = atrial sensed beats during pacing refractory periods;
AS = atrial sense; CD = charge delivered; TD = tachycardia
detected; TP = antitachycardia paced beat; TS = tachycardia sense;
VR = ventricular sensed beats during pacing refractory periods;
VS = ventricular sense VP = ventricular paced beat.
|
|
 |
Clinical Trials of the ICD
|
|---|
Secondary Prevention of SCD
The impact of the ICD on survival in patients with a history of
life-threatening arrhythmias has now been assessed in three randomized
trials (Table 1
).
The Antiarrhythmics Versus Implantable Defibrillator Trial
(AVID)2
evaluated the efficacy of the ICD in reducing
total mortality in patients with an ejection fraction (EF) < 40% and
a history of SCD or sustained VT with syncope or hemodynamic
compromise. In the control arm of the study, 90% of patients received
empiric amiodarone, and the remainder received Holter-guided sotalol.
This trial was terminated early due to a clear benefit in the
ICD-treated group. At follow-up after 3 years, survival rates were 75%
for the ICD group vs 61% for the antiarrhythmic group. Although
ß-blocker use was more prevalent in the ICD group, adjusting for this
imbalance did not alter the mortality rate reduction attributable to
the ICD.
The Cardiac Arrest Study-Hamburg Trial4
randomly assigned
survivors of cardiac arrest to ICD or to treatment with propafenone,
metoprolol, or amiodarone. The propafenone arm was withdrawn early due
to an observed excess mortality rate. The all-cause mortality rate was
12.1% in the ICD arm vs 19.6% in the combined arms of amiodarone and
metoprolol at follow-up after 2 years, a statistically significant
reduction. Interestingly, no difference was observed between the
amiodarone and metoprolol arms.
The Canadian Implantable Defibrillator Study (CIDS) randomized a
patient population, which was similar to the one in the AVID trial, to
an ICD or to empiric amiodarone. At 5 years, a trend in favor of
reduced all-cause mortality in the ICD treatment group was present,
although this was not statistically significant.3
The
trial design of CIDS may have lessened the impact of ICD therapy in
this study. In contrast to the AVID trial, this study did not mandate a
poor EF for all patients enrolled. Included in enrollment were patients
with syncope without spontaneous VT who subsequently underwent
electrophysiologic (EP) study. Evidence from an EP study of inducible
VT satisfied enrollment criteria, without regard to EF. This is
reflected in the difference between the average EF for the CIDS patient
population (34%) and that for the AVID population (27%). The
difference in EF between the patient populations of the CIDS and the
AVID trial also extends to New York Heart Association (NYHA) heart
failure class, a known correlate to the risk of SCD.32
In
CIDS, > 50% of patients were of asymptomatic heart failure status,
whereas in the AVID trial > 70% of patients had NYHA class II-III
heart failure. Thus, the results of CIDS suggest that patients
with symptomatic VT and milder degrees of left ventricular (LV)
dysfunction (EF, > 35%) may be adequately treated with amiodarone,
whereas patients experiencing episodes of VT who have poorer EF values
and NYHA status are best treated with ICD implantation. Indeed,
a subgroup analysis of CIDS indicates a much stronger trend to improved
survival in ICD-treated patients with EFs < 35%.
Presently, treatment with an ICD is a class I indication for secondary
prevention in survivors of cardiac arrest not due to a reversible
cause, for patients with syncope of unknown etiology and inducible
VT/VF, and in patients with spontaneous, sustained VT (Table 2
).8
Primary Prevention of SCD
Since only 2 to 30% of persons who have cardiac arrest survive, a
strong impetus to identify high-risk patients for the primary
prevention of SCD exists.33
34
35
EP testing is a
well-established procedure that has been shown to predict the risk of
SCD in patients with coronary artery disease.36
37
This
test has played a key role in clinical trial designs for primary
prevention.
The Multicenter Automatic Defibrillator Trial
(MADIT)5
was the first prospective, randomized
trial assessing the value of the ICD. Patients with prior myocardial
infarctions (MIs), LV EFs < 35%, evidence of documented nonsustained
VT (NSVT) and inducible VT on EP study that was not suppressed by IV
procainamide randomly received an ICD or conventional therapy. A
statistically significant difference in the total mortality rate was
observed, 15% in the ICD group and 39% in the conventional therapy
group over an average follow-up period of 2.5 years. Although a
striking benefit for the ICD was present, the trial was criticized for
the lack of a unified approach to drug therapy in the conventional
treatment group. At last patient contact, only 5% of patients in the
conventional group were receiving ß-blocker therapy vs 27% in the
ICD group. The proportion of participants in the conventional group
receiving therapy with angiotensin-converting enzyme inhibitors was
51%, and the proportion receiving amiodarone was 45%.5
The recently completed Multicenter Unsustained Tachycardia
Trial6
randomized a patient population similar to that in
the MADIT. Patients with inducible VT received medical therapy alone
(excluding antiarrhymic drugs) or an EP-guided approach. Patients in
the EP-guided group whose conditions were suppressible by
antiarrhythmic drug therapy were maintained on regimens with their
respective drugs. Patients with VT not suppressible by antiarrhythmic
drug therapy received an ICD. At a median follow-up of 39 months, a
statistically significant benefit was evident with respect to the
primary end point of SCD or resuscitated cardiac arrest in the
EP-guided group (25%) compared to that in the medical therapy group
(32%). Further analyses indicated that the benefit seen with
the EP-guided approach was due solely to the ICD group (p < 0.001).
Patients receiving EP-guided antiarrhythmic drug therapy did not show
improved outcomes compared to medically treated patients not receiving
antiarrhythmic drug therapy. Patients enrolled in this
trial who were not inducible at EP study, and therefore
were considered to be at low risk for SCD, were followed-up in a
registry. Interestingly, this patient population had an unexpectedly
high mortality rate of 24% at follow-up.6
The Coronary Artery Bypass Graft-Implantable Cardioverter-Defibrillator
Study (CABG-PATCH) study randomized patients prior to coronary
surgery with EFs < 35% and positive signal-averaged ECGs to either
the ICD group or the control group. No difference in overall
mortality was observed.7
The study did not require NSVT or
inducible VT for trial enrollment. It is suggested that patients with a
similar degree of LV dysfunction but without inducible VT may be at a
lower risk for SCD.38
39
In addition, revascularization
may have decreased the risk for ischemia-induced arrhythmias in this
patient population. These factors may help to explain the much lower
mortality rate in the CABG-PATCH control arm (18%) than in the MADIT
control group (39%).
The ongoing Sudden Cardiac Death in Heart Failure Trial will
present the most stringent control intervention thus far vs the ICD.
Patients with symptomatic (ie, NYHA class II-III) ischemic
or nonischemic cardiomyopathy and EFs < 35% are being randomized to
treatment with ICD, amiodarone, or placebo. A key component to this
trial is the strong encouragement for ß-blocker use, targeting 70%
of the patient population. The relevance of this is highlighted by
recent trials40
41
indicating a significant reduction in
all-cause and sudden death mortality in congestive heart
failure patients randomized to treatment with ß-blockers.
Furthermore, the combination of ß-blocker and amiodarone has been
suggested in post hoc analyses to have a significant impact
on reducing the number of deaths from arrhythmia.42
In summary, the current evidence suggests that patients with a history
of cardiac arrest or sustained VT and syncope are best treated with an
ICD. Patients with EFs < 35%, coronary artery disease, and NSVT
should be referred for EP study. If inducible, they should receive an
ICD.
 |
Managing Patients With ICDs
|
|---|
Troubleshooting Inappropriate ICD Function
The first issue in assessing a patient with a recent
defibrillation discharge is to determine whether therapy was
appropriate. Despite the sophisticated rhythm-discrimination algorithms
that are in current use, rapid atrial fibrillation remains the most
common cause of inappropriate shocks.21
22
The main reason
for this is atrial fibrillation ventricular rates that meet VF
zone criteria in which detection is based on the ventricular rate alone
and the programming of rhythm-discrimination algorithms is not
permitted. Preventive options include increasing the VF detection rate,
adding AV nodal blocking agents, or considering AV nodal ablation.
Inappropriate therapy also may result from the sensing of
electrical chatter. Electrical chatter develops in the
presence of lead fractures or insulation breaks, or may result from a
loose connection to the PG.43
44
45
These problems may be
evident on analysis of stored EGMs in which nonphysiologic R-R
intervals (ie, those of < 150 ms) may be observed. Lead
problems should be suspected on device interrogation when marked
variation of R-wave sensing, pacing thresholds, and impedance
measurements exist. Occasionally, having the patient perform physical
maneuvers such as straining or arm movements may reproduce electrical
chatter while observing ICD telemetry. An overpenetrated chest
radiograph may localize a lead fracture. Issues related to lead failure
or to loose connections require surgical intervention. Acute management
of a patient with inappropriate, incessant shocks not caused by a
ventricular arrhythmia should include placing a magnet over the device.
This maneuver will disable therapy delivery but will have no effect on
required pacing.
The inability of therapy to convert an arrhythmia may have dire
consequences. Lead failure or malposition may be a culprit, resulting
in insufficient energy delivery. Altered tissue substrate may render
ATP unsuccessful due to failure to adequately capture. Defibrillation
thresholds (DFTs) may be increased due to amiodarone initiation or to
severe electrolyte/acid-base abnormalities.46
Increasing
the programmed shock energy level may suffice, but a thorough device
evaluation is required. Managing an acute ventricular arrhythmia not
successfully terminated by ICD therapy should not differ from the
approach used in patients without ICDs. An ICD discharge during patient
contact by a resuscitating team member will not harm or pose a risk to
the individual.
Antiarrhythmic Drug-ICD Interactions
ICD implantation may obviate the need for long-term antiarrhythmic
drug therapy in a large number of patients. However, frequent shocks
due to atrial tachyarrhythmias and/or ventricular arrhythmias will
require the initiation of drug therapy in the management of certain
patients. In addition to the possibility of decreasing VT rates out of
programmed detection zones, therapy with antiarrhythmic drugs also may
cause proarrhythmia. An increase in ICD therapies correlating with
recent initiation of an antiarrhythmic agent should raise this
suspicion.
The effect of antiarrhythmic drugs on the minimum energy requirement
for successful defibrillation, or DFT, requires special attention. At
ICD implantation, the DFT is determined and a 10-J safety margin
typically is added to ensure defibrillation efficacy. Some
antiarrhythmic agents may alter the DFT as a result of their EP
properties (Table 3
). In general, class IA antiarrhythmic drugs, such as quinidine and
procainamide, appear to have little effect on the DFT at therapeutic
doses.46
The short-term administration of class IB agents,
including lidocaine, phenytoin, and mexiletine, has been shown to
increase the DFT.47
The class III agents amiodarone and
sotalol are commonly used for the maintenance of sinus rhythm in
patients with paroxysmal atrial fibrillation. Their effects on the DFT
are disparate. Amiodarone has been shown in several studies to elevate
the DFT, while sotalol consistently has no effect or lowers the
DFT.46
48
The addition of an antiarrhythmic agent to the therapy of ICD patients
may have a significant effect on the efficacy of ICD therapy. It is
necessary, therefore, to review programmed tachycardia zones and to
consider repeat DFT testing after drug loading.
Electromagnetic Interference
Electromagnetic interference may cause an ICD to falsely detect
tachycardia and to deliver inappropriate therapy. In the home
environment, there have been no reports of normally functioning
domestic appliances causing any inappropriate shocks in modern ICDs.
Microwave ovens, portable telephones, or personal computers have not
been linked to interference in current ICD systems. Inappropriate
shocks from electric razors have been reported rarely.49
Hand-held radiofrequency remote controls may produce inappropriate
sensing when held in close vicinity to the ICD device but have not been
shown to adversely affect device function when held > 10 cm from the
chest wall.50
Patients should be advised to avoid the strong magnetic fields of
electronic theft surveillance systems or to walk through them without
pausing. Similarly, patients should present their device
identification card to airport security personnel and walk briskly
through the security gate. The ICD device may trigger the alarm. In
these instances, brief passage of a hand-held metal detector over the
device is innocuous; however, prolonged exposure (ie, > 30
s) should be avoided as this may inactivate programmed therapies in
some devices. Caution is needed in the use of cellular telephones. When
used, the phone should be held on the side opposite to the device. The
avoidance of close contact with the PG and lead system is recommended,
and, therefore, phones should not be placed in nearby
pockets.51
52
53
Patients with occupational exposures to
large magnetic fields may have their work environment assessed by an
ICD company representative for the possibility of inappropriate device
sensing.
In the hospital setting, MRI imaging is generally
contraindicated. The use of electrocautery in surgery may cause
electrical chatter and oversensing. The device is best turned off in
this setting with appropriate monitoring and an external defibrillator
nearby. In addition, electrocautery should be avoided in close
proximity to the device to prevent the risk of damage to internal
circuitry.
ICDs and Driving
The privilege of driving is a quality-of-life issue for ICD
recipients. Although the presence of an ICD will terminate sudden,
malignant ventricular arrhythmias, loss of consciousness may not be
prevented in up to 15% of episodes.54
Current data
suggest that the risk of fatal motor vehicle accidents involving ICD
recipients is low. European data indicate that only 1.5 to 3.4% of
road accidents are attributed to sudden driver
incapacity.55
A study observing the driving history of 291
ICD patients over an average of 3 years noted only 11 traffic
accidents. Of these accidents, no fatalities occurred and there were no
accidents associated with driver syncope or defibrillation therapy. In
the remaining patient cohort, 5% of patients received ICD therapy
while driving, which was not associated with syncope or
accident.56
At present, most US states have no specific laws regarding
driving for ICD patients. A consensus statement providing
recommendations has been published by the North American Society of
Pacing and Electrophysiology/American Heart Association.57
These guidelines recommend against noncommercial driving for a period
of 6 months following ICD implantation for survivors of
life-threatening arrhythmias. Similarly, following successful ICD
therapy for a ventricular arrhythmia, the patient should refrain from
driving for 6 months. The basis for these time frames stems from data
indicating that the risk of recurrent arrhythmia is greatest soon after
an index event and decreases to < 0.7% per month after the
seventh month.58
It is recommended that aircraft piloting
and commercial driving be prohibited in patients receiving an ICD.
Similar precautions should be followed for patients involved in the
handling of heavy machinery.
 |
Cost-Effectiveness
|
|---|
Advancements in ICD technology leading to nonthoracotomy, pectoral
implantation have resulted in the vast majority of procedures being
undertaken in the EP laboratory. This transition from the operating
room has significantly reduced the cost associated with ICD
implantation. A recent cost-analysis for ICD procedures compared the
expense of procedures in an operating-room setting vs an EP
laboratory.59
Total costs were significantly less in the
EP laboratory ($4,541) than in the operating room ($9,431). This lower
cost was attributable to lower physician fees, hospital charges, and
the shorter length of postprocedural convalescence.
Present data analyzing the cost-effectiveness of the ICD relative to
conventional medical therapy are obscured given the rapid technological
advances of the ICD. Owens et al60
provided an economic
model estimating that a 30% reduction in mortality by the ICD relative
to amiodarone would satisfy the current standards of cost-effectiveness
(<$50,000 per life-year gained). This estimate compares favorably to
the observed risk reduction of 31% at 3 years in the AVID trial and
the 59% risk reduction at 2 years in MADIT.
Prior clinical information regarding the cost-effectiveness of ICDs has
been limited by small patient numbers and trial
design.61
62
Completion ofMADIT in 1996 has allowed for
an adequate duration of patient follow-up to assess the costs
accumulated by each treatment group.63
The expenses of
recurrent hospitalizations, physician visits, medications, laboratory
tests, and procedures were analyzed. In view of the significant
mortality reduction in MADIT, the resulting cost-effectiveness ratio
was $27,000 per life-year gained. In patients who received
nonthoracotomy procedures, this ratio was reduced to $23,000 per
life-year gained.
Advances in ICD technology will reduce the cost of this therapeutic
strategy further. Improved batteries are increasing device longevity,
better diagnostic features avoid inpatient or ambulatory Holter
monitoring, and dual-chamber pacing capabilities prevent the need for
separate pacemaker implantation when required.
 |
Future Developments
|
|---|
Future advances in ICD technology are motivated by the
demographics of an aging population and by the success of various
medical interventions in improving the survival of cardiac patients
prone to SCD. A goal of newer-generation ICD devices will be to provide
intervention prior to arrhythmia onset, avoiding discomforting
cardioversion or defibrillation therapy. ECG or physiologic parameters
known to increase ventricular arrhythmia risk, such as long-short R-R
intervals, T-wave alternans, heart rate variability, or hemodynamic
instability, may be intervened on by novel ICD therapies. Such
programmable therapies may include pacing to avoid long-short coupling,
intermittent antiarrhythmic drug infusion, or multisite pacing to
improve hemodynamics.64
65
66
67
68
Further advances in device
technology will undoubtedly expand the role of the ICD in the primary
and secondary prevention of SCD.
 |
Footnotes
|
|---|
Abbreviations: ATP = antitachycardia pacing;
AVID = Antiarrhythmics Versus Implantable Defibrillator Trial;
CABG-PATCH = Coronary Artery Bypass Graft-Implantable
Cardioverter-Defibrillator Study; CIDS = Canadian Implantable
Defibrillator Study; DFT = defibrillation threshold;
EGM = electrogram; EF = ejection fraction;
EP = electrophysiologic; ICD = implantable
cardioverter-defibrillator; LV = left ventricular;
MADIT = Multicenter Automatic Defibrillator Trial; MI = myocardial
infarction; NSVT = nonsustained ventricular tachycardia; NYHA = New
York Heart Association; PG = pulse generator; SCD = sudden cardiac
death; VF = ventricular fibrillation; VT = ventricular tachycardia
Received for publication June 20, 2000.
Accepted for publication August 3, 2000.
 |
References
|
|---|
-
Gillum, RF (1989) Sudden coronary death in the United States: 19801985. Circulation 79,756-765[Abstract/Free Full Text]
-
. The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators (1997) A comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias N Engl J Med 337,1576-1583[Abstract/Free Full Text]
-
Connolly, SJ, Gent, M, Roberts, RS, et al (2000) Canadian Implantable Defibrillator Study (CIDS): a randomized study of the implantable cardioverter defibrillator against amiodarone. Circulation 101,1297-1302[Abstract/Free Full Text]
-
Kuck, KH, Cappato, R, Siebels, J, et al (2000) Randomized comparison of antiarrhythmic drug therapy with implantable defibrillator in patients resuscitated from cardiac arrest: the Cardiac Arrest Study Hamburg (CASH). Circulation 102,748-754[Abstract/Free Full Text]
-
Moss, AJ, Hall, WJ, Cannom, DS, et al (1996) Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. N Engl J Med 335,1933-1940[Abstract/Free Full Text]
-
Buxton, AE, Lee, KL, Fisher, JD, et al (1999) A randomized study of the prevention of sudden cardiac death in patients with coronary artery disease. N Engl J Med 341,1882-1890[Abstract/Free Full Text]
-
Bigger, JT, Jr (1997) Prophylactic use of implanted cardiac defibrillators in patients at high risk for ventricular arrhythmias after coronary artery bypass graft surgery: Coronary Bypass Graft (CABG) Patch Trial Investigators. N Engl J Med 337,1569-1575[Abstract/Free Full Text]
-
Gregoratos, G, Cheitlin, MD, Conill, A, et al (1998) ACC/AHA Guidelines for Implantation of Cardiac Pacemakers, and Antiarrhythmic devices: executive summary; a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee on Pacemaker Implantation). Circulation 97,1325-1335[Free Full Text]
-
Wilbur, SL, Marchlinski, FE (1999) Implantable cardioverter-defibrillator follow-up. Cardiol Rev 7,176-190[Medline]
-
Mirowski, M, Mower, MM, Reid, PR (1980) The automatic implantable defibrillator. Am Heart J 100,1089-1092[CrossRef][ISI][Medline]
-
. PCD Investigator Group. (1994) Clinical outcome of patients with malignant ventricular tachyarrhythmias and a multiprogrammable implantable cardioverter-defibrillator implanted with or without thoracotomy: an international multicenter study. J Am Coll Cardiol 23,1521-1530[Abstract]
-
Grimm, RF, Flores, BF, Marchlinski, FE (1992) Electrocardiographically documented unnecessary, spontaneous shocks in 241 patients with implantable cardioverter defibrillators. Pacing Clin Electrophysiol 15,1667-1673[CrossRef][Medline]
-
Neuzner, J, Pitschner, HF, Schleper, M (1995) Programmable VT detection enhancements in implantable cardioverter defibrillator therapy. Pacing Clin Electrophysiol 18,539-547[CrossRef][Medline]
-
Marchlinski, FE, Callans, DJ, Gottlieb, CD, et al (1996) Benefits and lessons learned from stored electrogram information in implantable defibrillators. J Cardiovasc Electrophysiol 6,832-851
-
Reiter, MJ, Mann, DE (1996) Sensing and tachyarrhythmia detection problems in implantable cardioverter defibrillators. J Cardiovasc Electrophysiol 7,542-558[ISI][Medline]
-
Bardy, GH, Yee, R, Jung, W (1996) Multicenter experience with a pectoral unipolar implantable cardioverter-defibrillator. J Am Coll Cardiol 28,400-410[Abstract]
-
Rosenqvist, M, Beyer, T, Block, M, et al (1998) Adverse events with transvenous implantable cardioverter-defibrillators: a prospective multicenter study. Circulation 98,663-670[Abstract/Free Full Text]
-
Smith, PN, Vidaillet, HJ, Hayes, JJ, et al (1998) Infections in nonthoracotomy implantable cardioverter-defibrillators: can these be prevented? Pacing Clin Electrophysiol 21,42-45[CrossRef][Medline]
-
Schaumann, A, von zur Muhlen, F, Gonska, BD, et al (1996) Enhanced detection criteria in implantable cardioverter-defibrillators to avoid inappropriate therapy. Am J Cardiol 78,42-50[ISI][Medline]
-
Trappe, HJ, Achtelik, M, Pfitzner, P, et al (1999) Single-chamber versus dual-chamber implantable cardioverter defibrillators: indications and clinical results. Am J Cardiol 83(suppl),8D-16D[CrossRef][ISI][Medline]
-
Schaumann, A (1999) Managing atrial tachyarrhythmias in patients with implantable cardioverter defibrillators. Am J Cardiol 83(suppl),214D-217D[CrossRef][ISI][Medline]
-
Anderson, MH, Murgatroyd, FD, Hnatkova, K, et al (1997) Performance of basic ventricular tachycardia detection algorithms in implantable cardioverter defibrillators: implication for device programming. Pacing Clin Electrophysiol 20,2975-2983[CrossRef][Medline]
-
Wood, MA, Stambler, BS, Damiano, RJ, et al (1994) Lessons learned from data logging in a multicenter clinical trial using a late-generation implantable cardioverter-defibrillator: the Guardian ATP 4210 Multicenter Investigators Group. J Am Coll Cardiol 24,1692-1699[Abstract]
-
Nunain, SO, Roelke, M, Trouton, T, et al (1995) Limitations and late complications of third-generation automatic cardioverter-defibrillators. Circulation 91,2204-2213[Abstract/Free Full Text]
-
Swerdlow, CD, Ahern, T, Chen, PS, et al (1994) Underdetection of ventricular tachycardia by algorithms to enhance specificity in a tiered-therapy cardioverter-defibrillator. J Am Coll Cardiol 24,416-424[Abstract]
-
Weber, M, Bocker, D, Bansch, D, et al (1999) Efficacy and safety of the initial use of stability and onset criteria in implantable cardioverter defibrillators. J Cardiovasc Electrophysiol 10,145-153[ISI][Medline]
-
Brugada, J (1999) Is appropriate therapy a resolved issue with current implantable cardioverter defibrillators? Am J Cardiol 83(suppl),40D-44D[ISI][Medline]
-
Barold, HS, Newby, KH, Tomassoni, G, et al (1998) Prospective evaluation of new and old criteria to discriminate between supraventricular and ventricular tachycardia in implantable defibrillators. Pacing Clin Electrophysiol 21,1347-1355[CrossRef][Medline]
-
Porterfield, JG, Porterfield, LM, Smith, BA, et al (1993) Conversion rates of induced versus spontaneous ventricular tachycardia by a third generation cardioverter defibrillator. Pacing Clin Electrophysiol 16,170-173[CrossRef][Medline]
-
Gross, JN, Sackstein, RD, Song, SL, et al (1993) The antitachycardia pacing ICD: impact on patient selction and outcome. Pacing Clin Electrophysiol 16,165-169[CrossRef][Medline]
-
Rosenqvist, M (1996) Antitachycardia pacing: which patients and which methods? Am J Cardiol 78,92-97[CrossRef][ISI][Medline]
-
. The SOLVD Investigators. (1991) Effect of enalapril on survival in patients with reduced left ventricular ejection fractions and congestive heart failure. N Engl J Med 325,293-302[Abstract]
-
De Vreede-Swagemakers, JJM, Gorgels, APM, Dubois-Arbouw, WI, et al (1997) Out-of-hospital cardiac arrest in the 1990s; a population-based study in the Maastricht area on incidence: characteristics and survival. J Am Coll Cardiol 30,1500-1505[Abstract]
-
Liberthson, RR, Nagel, EL, Hirschman, JC, et al (1974) Prehospital ventricular defibrillation: prognosis and follow-up course. N Engl J Med 291,317-321
-
Weaver, WD, Hill, D, Fahrenbruch, CE, et al (1988) Use of the automatic external defibrillator in the management of out-of-hospital cardiac arrest. N Engl J Med 319,661-666[Abstract]
-
Roy, D, Marcand, E, Theroux, P, et al (1985) Programmed ventricular stimulation in survivors of an acute myocardial infarction. Circulation 72,487-494[Abstract/Free Full Text]
-
Iesaka, Y, Nogmi, A, Aonuma, K, et al (1990) Prognostic significance of sustained monomorphic ventricular tachycardia induced by programmed ventricular stimulation using up to triple extrastimuli in survivors of acute myocardial infarction. Am J Cardiol 65,1057-1063[CrossRef][ISI][Medline]
-
Wilber, DJ, Olshansky, B, Moran, JF, et al (1990) Electrophysiologic testing and nonsustained ventricular tachycardia: use and limitations in patients with coronary artery disease and impaired ventricular function. Circulation 82,350-358[Abstract/Free Full Text]
-
Bhandari, AK, Widerhorn, J, Sager, PT, et al (1992) Prognostic significance of programmed ventricular stimulation in patients surviving complicated acute myocardial infarction: a prospective study. Am Heart J 124,87-96[CrossRef][ISI][Medline]
-
The Cardiac Insufficiency Bisoprolol Study II (CIBIS-II): a randomized trial. Lancet 1999; 353:913
-
Effect of metoprolol CR/XL in chronic heart failure: Metoprolol CR/XL Randomized Intervention Trial in Congestive Heart Failure (MERIT-HF). Lancet 1999; 353:20012007
-
Boutitie, F, Boissel, JP, Connolly, SJ, et al (1999) Amiodarone interaction with beta-blockers: analysis of the merged EMIAT and CAMIAT databases; The EMIAT (European Myocardial Infarct Amiodarone Trial) and CAMIAT (Canadian Myocardial Infarct Amiodarone Trial) Investigators. Circulation 99,2268-2275[Abstract/Free Full Text]
-
Lawton, JS, Ellenbogen, KA, Wood, MA, et al (1996) Sensing lead-related complications in patients with transvenous implantable cardioverter-defibrillators. Am J Cardiol 78,647-651[CrossRef][ISI][Medline]
-
Daoud, EG, Kirsh, MM, Bolling, SF, et al (1994) Incidence, presentation, diagnosis, and management of malfunctioning implantable cardioverter-defibrillator rate-sensing leads. Am Heart J 128,892-895[CrossRef][ISI][Medline]
-
Roelke, M, ONunsin, SS, Osswald, S, et al (1995) Subclavian crush syndrome complicating transvenous cardioverter defibrillator systems. Pacing Clin Electrophysiol 18,973-979[CrossRef][Medline]
-
Brode, SE, Schwartzman, D, Callans, DJ, et al (1997) ICD-antiarrhythmic drug and ICD-pacemaker interactions. J Cardiovasc Electrophysiol 8,830-842[ISI][Medline]
-
Naccarelli, GV, Dougherty, AH, Wolbrette, D (1995) Antiarrhythmic drug-implantable cardioverter-defibrillator interactions. Zipes, DP Jalife, J eds. Cardiovascular electrophysiology: from cell to bedside 2nd ed. ,1426-1433 WB Saunders Philadelphia, PA.
-
Dorian, P, Newman, D (1993) Effect of sotalol on ventricular fibrillation and defibrillation in humans. Am J Cardiol 72,72A-79A[CrossRef][Medline]
-
Seifert, T, Block, M, Borggrefe, M, et al (1995) Erroneous discharge of an implantable cardioverter defibrillator caused by an electric razor. Pacing Clin Electrophysiol 18,1592-1594[CrossRef][Medline]
-
Man, KC, Davidson, T, Langberg, JJ, et al (1993) Interference from a hand held radiofrequency remote control causing discharge of an implantable defibrillator. Pacing Clin Electrophysiol 16,1756-1758[CrossRef][Medline]
-
Barbaro, V, Bartolini, P, Belloci, F, et al (1999) Electromagnetic interference of digital and analog cellular telephones with implantable cardioverter defibrillators: in vitro and in vivo studies. Pacing Clin Electrophysiol 22,626-634[CrossRef][Medline]
-
Fetter, JG, Ivans, V, Benditt, DG, et al (1998) Digital cellular telephone interaction with implantable cardioverter-defibrillators. J Am Coll Cardiol 31,623-628[Abstract/Free Full Text]
-
Occhetta, E, Piebani, L, Bartnik, M, et al (1999) Implantable cardioverter defibrillators and cellular telephones: is there any interference? Pacing Clin Electrophysiol 22,983-989[CrossRef][Medline]
-
Kou, WH, Calkins, H, Lewis, RR, et al (1991) Incidence of loss of consciousness during automatic implantable cardioverter-defibrillator shocks. Ann Intern Med 115,942-945
-
Halinen, MO, Jaussi, A (1994) Fatal road accidents caused by sudden death of the driver in Finland and Vaud, Switzerland. Eur Heart J 15,888-894[Abstract/Free Full Text]
-
Trappe, HJ, Wenzlaff, P, Grellman, G (1998) Should patients with implantable cardioverter-defibrillators be allowed to drive?: observations in 291 patients from a single center over an 11-year period J Interv Card Electrophysiol 2,193-201[CrossRef][ISI][Medline]
-
Epstein, AE, Miles, WM, Benditt, DG, et al (1996) Personal and public safety issues related to arrhythmias that may affect consciousness: implications for regulation and physician recommendations. Circulation 94,1147-1166[Free Full Text]
-
Larsen, GC, Stupey, MR, Walance, CG, et al (1994) Recurrent cardiac events in survivors of ventricular fibrillation or tachycardia: implications for driving restrictions. JAMA 271,1335-1339[Abstract]
-
Vorperian, VR, Lawrence, S, Chlebowski, K (1999) Replacing abdominally implanted defibrillators: effect of procedure setting on cost. Pacing Clin Electrophysiol 22,698-705[CrossRef][Medline]
-
Owens, DK, Sanders, GD, Harris, RA, et al (1997) Cost-effectiveness of implantable cardioverter defibrillators relative to amiodarone for prevention of sudden cardiac death. Ann Intern Med 126,1-12[Abstract/Free Full Text]
-
Larsen, GC, Manolis, AS, Sonnenberg, FA, et al (1992) Cost-effectiveness of the implantable cardioverter-defibrillator: effect of improved battery life and comparison with amiodarone therapy. J Am Coll Cardiol 19,1323-1334[Abstract]
-
Wever, EF, Hauer, RN, Schrijvers, G, et al (1996) Cost-effectiveness of implantable defibrillator as first choice therapy versus electrophysiologically guided, tiered strategy in post-infarct sudden death survivors: a randomized study. Circulation 93,489-496[Abstract/Free Full Text]
-
Mushlin, AI, Hall, WJ, Zwanziger, J, et al (1998) The cost-effectiveness of automatic implantable cardiac defibrillators: results from MADIT (Multicenter Automatic Defibrillator Implantation Trial). Circulation 97,2129-2135[Abstract/Free Full Text]
-
Nearing, BD, Huang, AH, Verrier, RL (1991) Dynamic tracking of cardiac vulnerability by complex demodulation of the T wave. Science 252,437-440[Abstract/Free Full Text]
-
Zehender, M, Faber, T, Grom, A, et al (1994) Continuous monitoring of acute myocardial ischemia by the implantable cardioverter defibrillator. Am Heart J 127,1057-1063[CrossRef][ISI][Medline]
-
Steinhaus, DM, Lemery, R, Bresnahan, DR, Jr, et al (1996) Initial experience with an implantable hemodynamic monitor. Circulation 93,745-752[Abstract/Free Full Text]
-
Daubert, J (1997) The compensatory pause prevention algorithm: preliminary results at the ventricular level. ,168-179 Futura Publishing New York, NY.
-
Pinski, SL, Fahy, GJ (1999) Implantable cardioverter-defibrillators. Am J Med 106,446-458[CrossRef][ISI][Medline]
This article has been cited by other articles:

|
 |

|
 |
 
Z. Goldberger and R. Lampert
Implantable Cardioverter-Defibrillators: Expanding Indications and Technologies
JAMA,
February 15, 2006;
295(7):
809 - 818.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. A. Driscoll and K. C. Sumino
BRONCHOMEDIASTINAL FISTULA CAUSED BY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR EPICARDIAL PATCH ELECTRODE
Chest Meeting Abstracts,
October 1, 2005;
128(4):
409S - 410S.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
I. R. Hanna and J. J. Langberg
The Shocking Story of Azimilide
Circulation,
December 14, 2004;
110(24):
3624 - 3626.
[Full Text]
[PDF]
|
 |
|