|
|
||||||||
Guest Access | Sign In via User Name/Password |
|||||||||
* From the Intensive Care Unit (Drs. Vignon and Rambaud), Dupuytren University Hospital, Limoges; Department of Biostatistic and Medical Informatic (Dr. Preux), University of Limoges, France; and Noninvasive Cardiac Imaging Laboratories (Drs. Spencer, Krauss, Lang, and Ms. Balasia), Section of Cardiology, Department of Medicine, The University of Chicago Medical Center, Chicago, IL.
Correspondence to: Roberto M. Lang, MD, The University of Chicago Medical Center, 5841 S Maryland Ave, Chicago, IL 60637; e-mail: rlang{at}medicine.bsd.uchicago.edu
| Abstract |
|---|
|
|
|---|
Methods and results: During an 8-year period, patients at high risk of AD (n = 261) or TDA (n = 90) who underwent a TEE study and had confirmed final diagnoses were studied. In an initial retrospective series, linear artifacts were observed within the ascending and descending aorta in 59 of 230 patients (26%) and 17 of 230 patients (7%), respectively. TEE findings associated with linear artifacts in the ascending aorta were as follows: displacement parallel to aortic walls; similar blood flow velocities on both sides; angle with the aortic wall > 85°; and thickness > 2.5 mm. Diagnostic criteria of reverberant images in the descending aorta were as follows: displacement parallel to aortic walls, overimposition of blood flow, and similar blood flow velocities on both sides of the image. In a subsequent prospective series (n = 121), systematic use of these diagnostic criteria resulted in improved TEE specificity for the identification of true intra-aortic flaps.
Conclusions: Misleading intra-aortic linear artifacts are frequently observed in patients undergoing a TEE study for suspected AD or TDA. Routine use of the herein-proposed diagnostic criteria promises to further improve TEE diagnostic accuracy in the setting of severely ill patients with potential need for prompt surgery.
Key Words: aorta diagnosis echocardiography imaging
| Introduction |
|---|
|
|
|---|
Multiple-path artifacts are the result of reverberations between strongly reflective surfaces. Multiple reflections occur when the ultrasound beam strikes an interface with large impedance mismatch between media (eg, soft tissue or fluidgas), particularly if the interface is oriented perpendicular to the direction of sound propagation.11 Reverberations between the interface and the esophageal transducer may occur, resulting in linear artifacts that do notcorrespond to anatomic structures. When located within the thoracic aorta, linear artifacts may be misinterpreted as intraluminal flaps and lead to false-positive diagnoses7 8 9 10 12 13 14 and needless thoracotomy.2 13 15
Linear artifacts within the ascending aorta are commonly encountered in patients with suspected AD.16 17 18 Little information is currently available on the TEE diagnostic criteria required to distinguish linear artifacts from true aortic flaps. In addition, to our knowledge, the incidence of these artifacts within the descending thoracic aorta and their potential impact on TEE diagnostic accuracy have not yet been studied. Accordingly, the aims of the present study were (1) to determine the incidence of intravascular linear artifacts within the ascending and descending thoracic aorta in a large cohort of patients undergoing a TEE study for a suspected acute aortic condition; (2) to establish the differential TEE diagnostic criteria that distinguish intra-aortic linear artifacts from true flaps associated with AD or TDA; and (3) to prospectively evaluate the impact of these criteria on the TEE diagnostic accuracy of aortic flaps.
| Materials and Methods |
|---|
|
|
|---|
In order to determine the incidence of intra-aortic linear artifacts and to establish the differential diagnostic criteria that would allow differentiation between multiple-path artifacts and true aortic flaps, we retrospectively studied 230 patients (172 men and 58 women; mean age, 58 ± 16 years; range, 17 to 85 years) who underwent TEE to rule out an AD or a TDA (protocol A). Subsequently, TEE diagnostic criteria that were independently predictive of the presence of underlying linear artifacts were prospectively tested in a second cohort of 121 patients (76 men and 45 women; mean age, 56 ± 17 years; range, 17 to 81 years) referred for suspicion of an acute aortic lesion (protocol B). In all patients, TEE interpretations were compared with the results of the reference methods.
TEE
All TEE studies were performed using either a 5-MHz monoplane
(n = 144) or multiplane probe (n = 207) connected to an ultrasound
system (Sonos 1500, 2500, or 5500; Hewlett-Packard; Andover, MA)
and recorded on videotape for off-line analysis. During the prospective
study (protocol B), all procedures were performed with a multiplane TEE
probe. Arterial BP, heart rate, and oximetry were monitored throughout
the procedure.
The TEE study was conducted as previously described,3 with particular attention directed toward the examination of the ascending, horizontal, and descending segments of the thoracic aorta, using two-dimensional echocardiography in conjunction with color Doppler echocardiographic flow mapping. Image depth and sector width were set to maximize frame rate, and the velocity scale was set between 60 cm/s and 80 cm/s in order to enable detection of low intra-aortic blood flow velocities, while limiting the aliasing effect. Gain settings were carefully adjusted to avoid the presence of color clutters outside the vascular lumen. When using the multiplane probe, the 110° to 140° echocardiographic plane was routinely used to visualize the ascending aorta to its fullest extent, while the descending thoracic aorta was examined in both the transverse (0°) and longitudinal views (80° to 110°). In addition, the TEE examination was also focused on excluding the presence of an associated cardiac abnormality (eg, pericardial effusion, aortic regurgitation).
Data Analysis
Protocol A:
Two experienced observers blinded to both the
clinical history and final diagnosis jointly reviewed the TEE studies.
In each patient, the ascending and descending segments of the thoracic
aorta were evaluated separately. The observers determined whether a
persistent linear intra-aortic image was present or absent, using
adequate gain settings. Accordingly, nonlinear artifacts such as mirror
images17
and reflection or comet tail
artifacts19
were not studied, because they usually do not
resemble intra-aortic flaps. Each linear intravascular image was then
described using the following qualitative and quantitative parameters.
Qualitative two-dimensional TEE parameters included the following: (1) mobility; (2) if mobile, the type of displacement (ie, free or parallel displacement to the aortic wall); (3) sharpness of the image borders; (4) image confined to the aortic lumen or extending outside the aortic wall; (5) presence of a pericardial effusion; and (6) presence of a left pleural effusion.20 Additional qualitative parameters were obtained from the color Doppler echocardiographic flow mapping of the thoracic aorta and aortic valve: (7) overimposition of blood flow on the linear image, or not; (8) different or similar blood flow velocities on both sides of the intra-aortic image; (9) presence of blood flow turbulence as reflected by a mosaic of colors surrounding the linear image, or normal laminar blood flow pattern; (10) presence of an entry or a reentry tear defined as the presence of blood flow through a typical communication of the linear image, or not21 ; and (11) presence of more-than-mild aortic regurgitation.21
Quantitative parameters were then measured independently by a third experienced observer who did not have access to the clinical charts and was unaware of qualitative data recordings. All measurements were performed at end-diastole (identified by the peak R wave of the ECG) in the transverse view, using electronic calipers. The following measurements were performed at the level of the ascending aorta: (1) diameter of the aorta obtained immediately above the sinuses of Valsalva; (2) diameter of the anatomic structure located posteriorly, ie, the left atrium or the right pulmonary artery according to the level of the tomographic plane (Fig 1 , left, A); and (3) in the presence of an intra-aortic linear image, the distance between the transducer (or the posterior wall of the left atrium or right pulmonary artery) and the posterior wall of the ascending aorta, and the distance separating the latter from the leading edge of this linear image in the same tomographic plane, along an axis perpendicular to the ascending aorta (Fig 1 , center, B). The ratio between the diameter of the ascending aorta and the adjacent posterior anatomic structure was calculated. The following measurements were performed at the level of the descending thoracic aorta: (1) diameter of the aorta at the level of the aortic isthmus, (2) and distance between the esophageal scope and the anteromedial descending aortic wall. Finally, when an intra-aortic linear image was observed, the maximal thickness of this image as well as the angle between the image and the vertical axis tangential to the aortic wall were measured (Fig 1 , right, C).
|
Protocol B: All TEE studies were reviewed by the same investigators, who again were unaware of the final diagnoses. In each patient who exhibited an intra-aortic linear image, both investigators used the TEE diagnostic criteria established in protocol A to determine whether the linear image was an artifact or a true flap. The diagnostic accuracy of TEE for the identification of intra-aortic flaps was then determined and compared to that obtained in protocol A.
Statistical Methods
Protocol A:
In order to determine the independent TEE
parameters predictive of the presence of a linear artifact in patients
exhibiting an intra-aortic linear image, a logistic regression model
was developed for both the ascending and descending thoracic aorta.
Initially, continuous variables were compared between patients with
linear artifacts and patients with true flaps, using the nonparametric
Mann-Whitney test. Qualitative variables were compared using the
2 test or the Fishers Exact Test, when appropriate.
Parameters with a probability value < 0.25 in the univariate analysis
were then included in the regression model. Odd ratios and 95%
confidence intervals (CIs) were calculated for each parameter.
To confirm the origin of multiple-paths linear artifacts, the diameters of both the ascending aorta and the anatomic structure located posteriorly (ie, the left atrium or the right pulmonary artery), as well as the ratio between these two measurements were compared between patients with linear artifacts and patients with normal ascending aortas, using the Students t test. Results were expressed as mean ± SD, and p < 0.05 was considered statistically significant. All tests of significance were two tailed.
Finally, the sensitivity and specificity, as well as the positive and negative predictive values, were determined by comparing the initial interpretation of TEE studies with the corresponding results of the reference imaging techniques, or with anatomic findings. Ranges were obtained by considering initially inconclusive TEE studies as either positive or negative results.
Protocol B: The frequency of intra-aortic linear artifacts and true flaps in the ascending and descending segments of the thoracic aorta was compared between protocols A and B, using the Z test. The diagnostic accuracy of TEE for the identification of aortic flaps was determined using the diagnostic criteria established in protocol A. Diagnostic accuracy was obtained using either a single or a combination of TEE criteria. The combination providing the highest diagnostic accuracy was then established.
| Results |
|---|
|
|
|---|
|
|
Diagnostic Criteria Associated With Linear Artifacts: TEE parameters predictive of the presence of a linear artifact within the ascending aorta using univariate analysis are presented in Table 1 . Linear artifacts were thicker than aortic flaps (3.9 ± 1.6 mm vs 2.6 ± 1.2 mm; p < 0.001) and consistently appeared nearly horizontal in the aortic lumen in the transversal view (Fig 4 , top left, A, and top right, B), as reflected by a larger angle between the linear image and the ascending aortic wall (91 ± 10° vs 72 ± 39°; p < 0.001). None of the patients in whom a linear artifact was noticed within the ascending aorta had an associated pericardial effusion or evidence for an entry tear. In the presence of an intimal flap in the ascending aorta (Fig 4 , bottom left, C), the flap had different orientations within the aortic lumen and overimposition of blood flow was never observed using color Doppler echocardiographic mapping (Fig 4 , bottom right, d). Using the logistic regression model, the following criteria were selected as independent predictors of the presence of linear artifact in the ascending aorta: (1) displacement of the linear image parallel to aortic walls (odds ratio, 7.2; CI, 1.3 to 40.0; p = 0.02); (2) similar blood flow velocities on both sides of the linear image (odds ratio, 47.6; CI, 3.4 to 500.0; p = 0.004); (3) thickness of the linear image > 2.5 mm (odds ratio, 4.8; CI, 0.9 to 26.2; p = 0.06); and (4) an angle between the linear image and the aortic wall > 85° (odds ratio, 5.9; CI, 1.1 to 32.2; p = 0.04).
|
|
|
|
|
TEE Diagnostic Accuracy: The diagnostic accuracy of TEE for the identification of true intra-aortic flap is detailed in Table 3 . Three false-negative TEE diagnoses were attributed to the presence of limited dissections involving the descending thoracic aorta or the aortic arch, and one false-negative TEE diagnosis was attributed to the presence of a small traumatic medial tear confined to the aortic isthmus. Two patients underwent unnecessary thoracotomy based on the presence of a linear intraluminal image erroneously diagnosed as an intra-aortic flap. The first patient sustaining a type-B AD also had a dilated ascending aorta (65 mm), but without evidence of proximal dissection (Fig 6 , left, A). The second patient was suspected of sustaining a traumatic disruption of the aortic isthmus (Fig 6 , right, B), but visual inspection during surgery revealed only the presence of a hemomediastinum with a normal proximal descending thoracic aorta.
|
Protocol B
Among the 121 patients studied, 73 patients (60%) underwent the
TEE study to rule out a spontaneous AD, while the remaining 48 victims
of violent deceleration accidents (40%) were examined for suspected
TDA.
Of 33 patients with proven AD, 14 patients (42%) sustained a Stanford type-A AD while the remaining 19 patients (58%) had a type-B AD. Traumatic disruption of the aortic isthmus was confirmed in eight patients sustaining severe blunt chest trauma. Linear artifacts were observed in the ascending and descending segments of the thoracic aorta in 22 of 121 patients (18%) and 7 of 121 patients (6%), respectively. No acute associated aortic condition was found in 12 of 22 patients (54%) with a linear artifact in the ascending aorta, and in all patients with a reverberant image within the descending thoracic aorta. Of them, a single patient exhibited a linear artifact in both segments of the thoracic aorta. In 8 of 19 patients (42%) with proven Stanford type-B AD and in 2 of 8 patients (25%) with confirmed traumatic laceration of the aortic isthmus, a linear artifact in the ascending aorta was also observed. The frequency of linear artifacts and true flaps associated with either AD or TDA was similar between protocols (data not shown).
In two patients with true flap, all TEE diagnostic criteria individually predictive of linear artifacts were observed. One patient sustained a type-A acute AD, without evidence for an entry tear. In this case, the diagnosis was based on the presence of an associated significant aortic regurgitation and pericardial effusion, and on TEE findings consistent with a dissection of the descending thoracic aorta with a thrombosed false lumen. The other patient had a traumatic disruption of the aortic isthmus. TEE clearly disclosed the presence of a false aneurysm formation with a medial flap. The systematic use of the above-described TEE diagnostic criteria resulted in the absence of false-positive results. The combination of three of the four TEE criteria that were independent predictors of the presence of a linear artifact in the ascending aorta yielded the highest diagnostic specificity (Table 3) . Similarly, the presence of a single TEE criterion identified using the logistic regression model provided the highest diagnostic accuracy (Table 3) . When compared to protocol A, TEE specificity for the identification of true flaps in both segments of the thoracic aorta improved, and no patient underwent unnecessary surgery.
| Discussion |
|---|
|
|
|---|
Incidence of Intra-aortic Linear Artifacts
The incidence of linear artifacts within the ascending aorta has
been reported16
17
18
to be as high as 44 to 55% in
patients undergoing a TEE study to rule out an acute aortic
abnormality. In the present series, the incidence of linear artifacts
was lower (81 of 351 patients; 23%). This difference is conceivably
due to the larger size of our study population and to our use of strict
criteria to select artifacts that needed to be differentiated from true
intimal or medial flaps. As a result, easily identifiable pitfalls,
such as reverberations of aortic wall calcifications or comet tail
artifacts, were not recorded. As previously reported,7
linear artifacts were less commonly present in the descending thoracic
aorta, being identified in only 24 of 351 of our patients (7%) studied
for an AD or TDA. Interestingly, these linear artifacts were more
frequently observed in victims of severe blunt chest trauma at high
risk for TDA (13 of 90 patients vs 11 of 261 patients; p < 0.001).
This finding is of particular relevance because linear artifacts appear
to be more frequent in the vicinity of the aortic isthmus, where TDA
predominantly occur.3
4
5
TEE Identification of Linear Artifacts
The presence of a thick linear image (> 2.5 mm) and similar
blood flow velocities on both sides of the image were strong individual
predictors of an underlying ultrasound reverberation artifact within
the ascending aorta. Nevertheless, these TEE findings may be present in
certain patients suffering from type-A AD. Indeed, when involving a
substantial part of the media, aortic intimal flaps may at times appear
as thick linear images, similar to multiple paths artifacts. In
addition, blood flow velocities may be similar on both sides of the
intimal flap, particularly when a large entry tear located in the
proximal ascending aorta results in a circulating false channel. In
these cases, the use of the remaining TEE diagnostic criteria
associated with linear artifacts (ie, angle > 85°, and
parallel displacement to the aortic walls) remains useful, because in
the current series linear artifacts were consistently found to be
nearly horizontal within the ascending aorta. In contrast, aortic flaps
had various orientations across the vascular lumen. As previously
described,17
intra-aortic artifacts did not display the
oscillatory movement frequently seen in intraluminal flaps, but rather
exhibited a regular displacement that was parallel to aortic walls.
Using M-mode echocardiography, Evangelista et al16
demonstrated the clinical value of this sign to improve the diagnostic
accuracy of TEE for the identification of ascending AD. Finally, the
presence of oppositely directed blood flow in the two lumina delimited
by the intra-aortic linear image is suggestive of an underlying
AD.21
In the current study, displacement of the intra-aortic linear image parallel to the aortic walls in conjunction with overimposition of blood flow with similar velocities on both sides of the suspect image were individually predictive of the presence of a linear artifact within the descending thoracic aorta. Importantly, these TEE criteria may not be as valid in patients with suspected TDA, in whom multiple paths linear artifacts were more frequently encountered. In this condition, medial flaps are usually thicker than the intimal flaps associated with AD because they commonly involve the entire depth of the medial layer of the aorta.3 Although the medial flap fails to delimit two distinct channels with different blood velocities, high-velocity blood flow turbulence is usually observed in the surroundings of the disrupted aortic wall.3 In contrast, a near-normal laminar descending aortic flow pattern is consistently noted in the presence of linear artifacts.16 17 Finally, subadventitial TDA is frequently associated with false aneurysm formation,3 whereas the aortic contour usually remains normal in the presence of intraluminal linear artifacts.
Impact on TEE Accuracy
Although highly sensitive for the diagnosis of
AD1
7
8
16
22
and disrupted aortic
isthmus,5
23
the major widely acknowledged drawback of TEE
is its relatively poor specificity.7
8
9
10
24
25
Interestingly, this relative lack of specificity has been predominantly
reported for the diagnosis of Stanford type-A AD, whereas TEE
diagnostic specificity is usually higher when the dissecting process is
confined to the descending thoracic aorta.7
8
The presence
of linear artifacts within the ascending aorta has been hypothesized to
be responsible for most of the false-positive TEE
results.1
2
7
8
12
13
14
With the exception of one TEE study limited by suboptimal imaging quality, all inconclusive examinations encountered in protocol A of the current series (n = 24) were retrospectively attributed to the difficult interpretation of the presence of intra-aortic linear images. Of them, 21 images (87%) were linear artifacts, while only 3 images (13%) were related to the presence of an underlying intimal flap in the descending thoracic aorta. Interestingly, 13 of 21 linear artifacts (62%) were located in the ascending aorta. In this protocol, TEE diagnostic accuracy was similar to that reported in previous studies.1 14 16 22 The specificity of TEE for the identification of intimal flaps located within the ascending aorta was fairly higher than that reported by Nienaber et al.7 8 This discrepancy is presumably related to our larger cohort and the use of multiplane TEE probes in 37% of patients included in this protocol. In contrast with previous studies,7 8 we found a similar TEE diagnostic accuracy for the identification of intimal or medial flap in both the ascending and descending segments of the thoracic aorta (Table 3) . Although in the latter anatomic segment of the aorta the incidence of linear artifacts was lower (7%), TEE has frequently been inconclusive when such images were encountered in patients with suspected TDA.
A false-positive diagnosis of intra-aortic flaps in the setting of patients who undergo a TEE for a suspected acute aortic condition may lead to unnecessary surgery,2 13 15 as shown in two of our patients from the retrospective series. Accurate identification of multiple-path artifacts is crucial in the presence of a linear image located in the ascending aorta, because surgical treatment of a dissection, or rarely a traumatic disruption,26 of this anatomic segment of the thoracic aorta is unequivocally advocated. In the present study, 21 of 64 patients (33%) sustaining a type-B AD also exhibited linear artifacts in the ascending aorta.7 18 As previously described,2 13 15 this image has been misdiagnosed as a true flap, resulting in unnecessary surgery in one patient who had a type-B AD rather than a type-A AD (Fig 6 , left, A). In the setting of patients with severe blunt chest trauma, accurate diagnosis of linear artifacts within the descending thoracic aorta, as opposed to intraluminal medial flap, is also crucial because rapid surgery of subadventitial TDA is widely advocated to avoid lethal adventitial rupture. As in one of our patients (Fig 6 , right, B), needless thoracotomy because of false-positive TEE results has been previously reported.27
Importantly, the systematic use of strict TEE diagnostic criteria applied to intra-aortic linear images in protocol B substantially improved the specificity of this imaging modality for the identification of true flaps, within both the ascending and descending segments of the thoracic aorta (Table 3) . Hence, no false-positive TEE results were encountered and no patient underwent unnecessary surgery during the prospective study. However, the learning curve during protocol A may also have contributed to the higher diagnostic accuracy observed in protocol B (Table 3) .
Origins of Linear Artifacts
Using in vitro experiments, Appelbe et
al17
confirmed the origin of multiple-path linear artifact
by imaging two water-filled latex balloons of varying size placed in
series in a water tank. A linear artifact was only observed within the
balloon representing the ascending aorta when its size exceeded that of
the second balloon representing the left atrium. In addition, the
distance between the transducer and balloon interface was equal to the
distance separating this interface from the linear
artifact.16
Using M-mode echocardiography, Evangelista et
al16
recently described three distinct types of linear
artifacts within the ascending aorta: type-A artifacts (53%), which
were located twice as far from the transducer as the posterior aortic
wall; type-B artifacts (40%), which were located at the same distance
from the posterior aortic wall as the latter was from the right
pulmonary posterior wall; and type-C artifacts (7%), which were
located at the same distance from the posterior aortic wall as a
reverberation from the right pulmonary artery posterior wall. All these
artifacts exhibited a movement parallel to the posterior aortic wall,
as in the current series. The reverberation images observed in the
present study were type A (Fig 6
, left, A) and
type B (Fig 1 , center, B; Fig 3
, bottom
right, D; Fig 4
, top left, A)
artifacts. Type-A artifacts are thought to be generated when the echo
of the posterior aortic wall is partly reflected back by the
transducer,16
17
while type-B artifacts correspond to
ultrasound reverberations from a moving target (ie, the
right pulmonary posterior wall) and a moving mirror (ie, the
posterior aortic wall).16
The origin of type-A and type-B artifacts was confirmed in
vivo in the present study. In patients with intraluminal linear
artifacts, the diameter of the ascending aorta consistently exceeded
that of the adjacent posterior anatomic structure (diameters ratio
> 1), and was significantly greater than that seen in patients
without intra-aortic linear images. As a result, reverberation images
were more commonly observed when relative aortic dilatation was present
(Fig 6 , left, A), a finding commonly associated
with aortic dissection.3
13
18
These results are in
keeping with those recently reported by Losi et al,18
who
showed that a diameter of the ascending aorta (> 5 cm) that exceeds
the diameter of the left atrium (with a ratio
0.6), was predictive
of the presence of an underlying linear artifact. In addition, in our
patients with intra-aortic linear artifacts, the mean distance between
the transducer (type-A artifacts), or the posterior wall of the right
pulmonary artery (type-B artifacts), and the posterior wall of the
ascending aorta was similar to that separating the latter wall from the
leading edge of the intraluminal image (28.8 ± 6.1 mm vs
28.6 ± 5.7 mm; p = 0.9; Fig 1
, center, B;
Fig 3
, bottom right, D; Fig 4
, top
left, A; Fig 6
, left, A).
Similarly, Losi et al18
reported that linear artifacts
were consistently located in the aortic lumen twice as far from the
transducer as the posterior aortic wall, such as type-A
artifacts.16
In patients with linear artifacts within the descending thoracic aorta, the mean diameter of this vessel was similar to that of patients without reverberation images. However, in the presence of misleading linear artifacts, the descending thoracic aorta was consistently shifted anteriorly. This explains probably the higher incidence of linear artifacts within the descending thoracic aorta in patients with severe blunt chest trauma in whom a traumatic hemomediastinum, which usually results in an increased distance between the esophageal probe and the anteromedial aortic wall,28 was frequently observed (Fig 5 , top left, A; Fig 6 , right, B). Because hemomediastinum is frequently associated with TDA,3 accurate diagnosis of linear artifacts within the aortic isthmus in this clinical setting is critical to avoid false-positive TEE results.
Limitations
Because multiplane TEE probe was not available in all patients in
protocol A, the additional diagnostic value of various tomographic
planes for accurate identification of reverberation images was not
specifically studied. However, most of the proposed diagnostic
criteria may also be used in other TEE planes, as shown in Figure 7
. In addition, because of the inability of monoplane TEE probes to
adequately image the distal ascending aorta, the aortic arch was not
studied. Because linear artifacts may be confounding images only when
interpreted as aortic flaps, patients sustaining acute aortic condition
without associated intraluminal flaps were not studied. Therefore, the
reported diagnostic capability of TEE for the diagnosis of spontaneous
AD and TDA did not take into account certain clinical presentations
(eg, thrombosed false lumen, intramural hematoma, traumatic
intimal tear) for which a diagnosis may be challenging.
|
Other types of multiple-path artifacts, such as mirror artifacts, are frequently encountered during TEE examination of the descending thoracic aorta.17 We did not study these reverberation images because they are rarely misleading and visualized only if the field of view is expanded to accommodate a second signal.29
| Conclusion |
|---|
|
|
|---|
| Acknowledgements |
|---|
| Footnotes |
|---|
Received for publication February 7, 2000. Accepted for publication October 12, 2000.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
T. M. Beaver, F. N. Herrbold, P. J. Hess Jr, C. T. Klodell, and T. D. Martin Transferring Diagnosis Versus Actual Diagnosis at a Center for Thoracic Aortic Disease Ann. Thorac. Surg., June 1, 2005; 79(6): 1957 - 1960. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Sonker, D. F. P. M. Peek, J. J. C. van Der Meij, R. H. Heijmen, C. A. Seldenrijk, and P. J. Knaepen Primary lung tumor mimicking acute type A aortic dissection Ann. Thorac. Surg., May 1, 2004; 77(5): 1841 - 1843. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK |