* From the Division of Public Health Biology and Epidemiology (Drs. Mortimer, Fallot, and Tager), University of California, Berkeley; and Division of Occupational and Environmental Medicine (Dr. Balmes), University of California at San Francisco, San Francisco, CA.
Correspondence to: Kathleen Mortimer, ScD, MPH, 1918 University Ave, #3C, Berkeley, CA 94704; kmort{at}uclink.berkeley.edu
| Abstract |
|---|
|
|
|---|
Design: Based on data from a cohort of 92 children with asthma, we evaluated the agreement between a portable spirometer and a office-based spirometer, using an in-line technique to evaluate measures from the same effort. We compared a range of pulmonary function parameters collected during office-based tests, and also evaluated whether adequate adherence and data quality could be achieved in a home-based study of children with asthma.
Results: The agreement between the devices for the actual values of peak expiratory flow, FEV1, and forced expiratory flow at 25% of FVC was excellent. The portable device was programmed with customized software to grade each curve using revised American Thoracic Society acceptability and reproducibility criteria. For 74% of the curves, quality grade agreed with a grade assigned by physician review of the curve from the office-based spirometer. During 2 weeks of twice-daily monitoring at home, children completed an average of 23 of 28 possible sessions (83%). Of these, 84% had at least two acceptable and two reproducible curves. Although children ≥ 8 years old were not more adherent, they were significantly more likely to achieve acceptable and reproducible curves.
Conclusions: Portable spirometers can provide measurements that are highly comparable to those obtained from "gold standard" laboratory spirometers, and high-quality tracings can be achieved both at home and in the office setting. Visual inspection of the curves by experienced reviewers identified unacceptable curves that were not rejected by the quality control software. Portable spirometers are an important contribution to epidemiologic and clinical studies that require frequent measures of a more broad range of pulmonary function parameters than can be provided by peak flowmeters.
Key Words: pediatric asthma pulmonary function testing spirometry panel studies
| Introduction |
|---|
|
|
|---|
Two studies6 7 have evaluated the practicality and validity of home spirometry in children with asthma. Unfortunately, neither study validated the device used by comparison with a "gold standard," laboratory-based spirometer. Although both studies examined adherence and reproducibility of efforts in the home setting, neither study mentions the use of quality control measures to ensure acceptability of efforts prior to evaluation of their reproducibility.
We performed a direct comparison of a portable spirometer (EasyOne; ndd Medical Technologies; Andover, MA) with an office-based volume spirometer (model RS232; Morgan Scientific; Winchester, MA), based on an in-line technique to simultaneously measure forced expiratory maneuvers from both instruments. The acceptability of each curve was determined independently for each device. Customized quality control software was used to determine the acceptability of curves obtained by the portable spirometer. The acceptability of curves generated on the office-based spirometer was initially determined by the technician and underwent further independent review by three physicians experienced in the interpretation of spirometry (A.F., J.R.B., I.B.T.). Volume-time and flow-volume curves were reviewed. The curves from each effort in a given session were considered during the review. Any differences in interpretation were resolved by consensus. The agreement between the two methods of acceptability grading was compared. Additionally, a 2-week home study using only the portable spirometer was performed with the same children to assess adherence and the acceptability and reproducibility of efforts.
| Materials and Methods |
|---|
|
|
|---|
The FACES design consists of longitudinal and panel components with > 4 years of follow-up. Every 6 months, each child will undergo lung function testing. For the panel component, children will be observed for three 2-week panels per year for up to 4 years. During panel periods, daily data will be obtained on the following: twice-daily forced expiratory volumes, symptoms, and medication use. For both the panel and longitudinal components, lung function is the primary health outcome of interest. Given that it is not feasible to do office-based testing on a daily basis for several weeks, it was desirable to obtain an affordable and portable device that would give comparable measurements to the office spirometer. Key requirements included the following: (1) the ability to record and store full flow-volume curves and answers to a series of medication and symptom questions; (2) inclusion of quality control software and prompts to obtain acceptable and repeatable efforts; (3) time and date stamping of all records; (4) easy transfer of specific flows and volumes to a personal computer database; (5) ability, with minimal cleaning, to be reused to test multiple children; (6) easy calibration; (7) inclusion of an incentive program to improve adherence; (8) reminder alarm and limitations on the times of day that entries can be collected; and (9) compliance with American Thoracic Society (ATS) criteria for spirometer performance. Because FACES is not an intervention study, we wanted to mask the results to avoid influencing the childrens behavior with respect to their asthma, based on pulmonary function readings. The EasyOne device met all of these criteria.
Equipment and Setup
The EasyOne portable spirometer is a small, portable device that uses ultrasound flow sensors to measure airflow through a hollow disposable mouthpiece (thin-walled spirette). The ultrasonic flow measurement is independent of gas composition, pressure, temperature, and humidity, and eliminates errors due to these variables. Since the disposable spirette has no sensor elements, it does not perform a measurement function and does not require calibration. Hundreds of sessions of spirometric data can be stored in memory, and later can easily be uploaded to a computer database. All sessions are time and date stamped. The device has a screen that can display up to 40 characters of text as well as flow-volume curves. A numeric keypad allows the user to answer yes/no questions that are displayed on the screen. The device has no moving parts requiring calibration. The office-based volume spirometer used in this study is a 12-L dry rolling seal spirometer. The spirometer is attached to a monitor that displays flow-volume curves to assist the technician in the evaluation of acceptability and reproducibility.
Due to the inherent variability of pulmonary function parameters in children with asthma due to bronchial reactivity, we used reproducibility criteria that are less stringent than those proposed by ATS. The ATS criteria were established for more general testing, and do not account for the particular difficulties faced in testing children and asthmatics.8 Based on our criteria that were developed specifically for asthmatic children (Table 1 ) the technician can decide whether the data from each curve should be saved.
|
|
Children were coached through a prebronchodilator session of forced expiratory maneuvers and were seated, wearing nose-clips. The children were allowed up to eight attempts to achieve three acceptable efforts with at least two that met reproducibility criteria (Table 1) . Each effort was measured simultaneously by both machines with the in-line setup. Two puffs of albuterol were then administered via a metered-dose inhaler and spacer with facemask, and children were allowed to rest for 20 min. A postbronchodilator session was then performed. Children again were permitted up to eight efforts to achieve three acceptable maneuvers with at least two of them meeting reproducibility criteria. For participation in the baseline visit, children were given gift packages that included approximately $50 worth of coupons from local vendors and food establishments.
Home Panel Measurements
Within 1 month of completion of the baseline spirometry measurements, children were given an EasyOne portable spirometer with the customized FACES quality control software to perform twice-daily spirometry at home for 2 weeks. Prior to home delivery, new batteries were installed and each unit underwent verification of calibration at varying flow rates with the same 3-L syringe that is used in the laboratory.9
Children were given a set of photographic instructions accompanied with text instructions in both English and Spanish. The serial number of each unit was recorded in each childs chart. The portable spirometers were programmed with an alarm that activated during the time windows of 7 to 9 AM and from 7 to 10 PM to remind children to perform the tests. The spirometers operate only within these time windows, and the units only allow one session of up to six attempts during each time window. The spirometers used the same acceptability and reproducibility criteria noted previously (Table 1)
to evaluate each effort. If a criterion was not met, the message prompted the child on how to improve the subsequent effort, such as "Blow out longer." Children were allowed six attempts to achieve three acceptable efforts with two of them meeting all of the reproducibility criteria. Data were saved for each of the three best efforts even if the acceptability and reproducibility criteria were not satisfied during the session. The curve with the largest sum of FEV1 plus FVC was chosen as the best. The only exception was when an unacceptable curve had the greatest sum, but the session included an acceptable curve with a smaller sum of FEV1 plus FVC. All sessions were time and date stamped. At the end of each session, children were prompted to answer a series of questions about asthma symptoms and medication use by entering "yes" or "no" on the spirometer keypad. After the completion of the 2-week home panels, a $25 gift certificate to a toy store was mailed to each child.
After answering all the questions at the end of each session, the cumulative point total defined by the FACES incentive scoring system was displayed. The incentive system was designed to balance the need to reward children for performing daily readings, yet not discourage children who were unable to achieve acceptable or reproducible tracings. For each session, 500 points were awarded if three acceptable and at least two reproducible efforts were obtained and all questions were answered. If six attempts were made and all questions were answered but the efforts failed to meet either the acceptability or reproducibility criteria, the session was awarded 200 points. Over a 14-day period, a total of 14,000 points could be obtained. The number of points was stored in the software and children were allowed to pick out various prizes during their next visit, based on the total number of points.
Agreement Between Physician and Quality Control Software Grading
To assess the ability of the customized FACES quality control software to determine whether a forced expiratory maneuver was acceptable, the three physicians independently reviewed the office-based maneuvers stored by the volume spirometer, based on the same acceptability criteria applied to the portable spirometer. The reviewers included one additional criterion that was not programmed into quality control software. Efforts were declared unacceptable if the middle portion of the time-volume curves displayed any deviation as outlined by the ATS 1994 update of spirometry standardization.8
The reviewers assessment for each effort obtained from baseline sessions on the volume spirometer was then compared to the determination of acceptability for that effort of the portable spirometer. Because the quality control software was not programmed to evaluate the entire curve, the physicians visual review takes priority throughout this report.
For the 2-week home panel testing, the childs identifying information, time and date, pulmonary function testing measures, acceptability status, and answers to the questions for each session were uploaded to a single database for evaluation (Access 97 for Windows; Microsoft; Redmond, WA). Adherence during the 2-week panels was calculated as a percentage of sessions out of a total possible 28 sessions completed by a child. We also calculated the percentage of sessions with at least three acceptable efforts and with at least two curves that met all three of the reproducibility criteria (FEV1, FVC, and PEF). Additional reproducibility measures were calculated separately for FEV1, FVC, and PEF. The panel data included in this report have not yet undergone expert visual review to determine if they meet our more stringent acceptability and reproducibility criteria. This will be done in future health-effects analyses, in which case we can "override" the determination of the software of the quality of the data if necessary.
Statistical Methods
For each effort and several lung function parameters, the measured difference between the two devices (office-based minus portable device value) was plotted against the mean of the values measured by both instruments.10
11
Only efforts that were determined acceptable by both the reviewers and the portable spirometer were used in the plots. Due to the differing abilities of the children, the number of maneuvers contributed by each child to the plots was variable. A repeated-measures analysis using a random-effects model (lme routine of Splus; Insightful; Seattle, WA) was used to calculate a 95% confidence interval for each of the plots.10
The Shapiro-Wilk test and normal probability plots were used to test the normality of the differences observed between the devices for each of the spirometric indexes. Confidence intervals were also obtained nonparametrically using a bootstrap technique. Ten thousand bootstrap samples were generated to estimate the 2.5% and 97.5% quantiles of the distribution of the measurement differences between the two devices.
"Completeness" and "reproducibility" of the 2-week panel data were continuous variables that were compared across various demographic characteristics. Testing across dichotomous variables was done using PROC TTEST (SAS Institute; Cary, NC) and tests for categorical variables with more than two levels were assessed using PROC ANOVA (SAS Institute), with a significance level of p < 0.05; SAS version 6.12 was used for these analyses.12
| Results |
|---|
|
|
|---|
|
|
|
|
|
To obtain an additional quantitative measure of agreement, we also evaluated the devices using reproducibility criteria across devices, rather than within devices. The "best curve" from the session on the office-based spirometer was identified as the one with the maximum sum of FEV1 and FVC. We then calculated whether the portable spirometer data from the same effort met the reproducibility criteria using the best curve from the office-based device. For 72% of the sessions, all three reproducibility criteria were met. The reproducibility based on individual parameters was 97%, 75% and 98% for FEV1, FVC, and PEF, respectively.
Adherence During the Home Panels
The adherence information from the 2-week home panels is presented in Table 6
. All 92 children in this analysis were doing 2-week panels for the first time. On average, the children attempted 23 of 28 possible sessions (83%). Adherence was only slightly lower for morning vs evening sessions. Adherence was similar for the first and second weeks of monitoring, 82% and 85%, respectively. According to the grading of the quality control software, at least one acceptable effort was obtained during 94% of the sessions. Three acceptable and two reproducible curves were obtained in 1,271 of 2,151 sessions (59%) [Table 5
]. The median number of sessions per child that met these criteria was 16 (range, 0 to 28). When the reproducibility criteria were based only on one parameter, the median number of sessions per child that met these criteria was 74%, 75%, and 82% for FEV1, FVC, and PEF, respectively.
|
Neither adherence nor the number of points received were significantly different according to age, race, or gender. The percentage of sessions with three acceptable and two reproducible tracings was significantly higher in the children who were > 8 years old, but did not significantly differ by race or gender. More frequent medication use was associated with increased adherence but decreased ability to achieve acceptable and reproducible tracings. Similarly, children who were more symptomatic (ie, reported more symptoms in the 12 h prior to testing) had similar adherence, but were less likely to achieve acceptable and reproducible tracings. There was only one battery failure during these panels, and all of the devices were in working order when retrieved from the families.
| Discussion |
|---|
|
|
|---|
The second goal was to demonstrate whether a substantial portion of the asthmatic children in this age range (6 to 11 years) were able to provide good quality data, both in the office and home settings. When supervised, the percentage of sessions with two acceptable and two reproducible efforts was 66% in the office setting. In the unsupervised panel data, 62% of the sessions had at least two acceptable and reproducible tracings. We propose that requiring three rather than two acceptable tracings may be too stringent a requirement for young children with asthma.
The third goal was to provide a comparison of the physician review vs quality control software. Our experience indicates that visual inspection of curves by a qualified reviewer will identify unacceptable curves that would be classified differently by the criteria of the software. Although the agreement between the two methods of assessment was high, the software was not programmed to identify problems in the midportion of the curve, and 25% of the curves that were coded as acceptable by the software were rejected by physician review. In the subset of 56 children who had at least one discordant scoring (ie, the physicians rejected curves that the software accepted), the mean FEV1 of curves accepted by the physicians was 1.59 L vs 1.56 L for the curves accepted by the software. The mean FEV1 for the curves that were accepted only by the software was 1.50 L. This indicates the curves that the physicians rejected and the software accepted have lower FEV1s, leading to a lower mean. The extent to which the inclusion of those curves in a longitudinal analysis would impact the findings will be evaluated as we collect more data.
The final goal was to evaluate adherence during the 2-week panels. Adherence was as good and often better than that reported in other panel studies, which range from 60 to 85% over a similar time period.6 13 14 As more children complete the 2-week panels, we will examine additional characteristics that are predictive of good adherence. It is clear that some children were not able to produce acceptable and reproducible tracings, either on their own or with coaching; however, they were able to provide valuable symptom and medication information. While the older children were not more adherent, they were more likely to achieve reproducible curves. Future analyses will evaluate the improvement in childrens performance over the course of the study.
Clinical and epidemiologic studies often have been limited to measurements derived from peak flowmeters. This study suggests that it will be feasible to expand the range of pulmonary function measures that can be captured in these settings. These devices are substantially more expensive than peak flowmeters (approximately $600 vs $25); however, they are approved for multiple users (with disposable mouthpieces). The need for more sensitive and physiologically relevant measures may warrant that expense. For example, in a clinical setting, the accurate monitoring of pulmonary function may help avoid the use of expensive health-care services such as diagnostic testing or hospitalization. The ability of epidemiologic studies, such as those focused on air pollution, to observe evidence of functional alterations could be greatly improved if a wider range of parameters of pulmonary function were measured. Most studies of air pollution and asthma have found statistically significant, but clinically small effects on PEF rate and FEV1. One of the goals of FACES is to determine whether changes in FEF2575 and FEF75 are more sensitive indicators of air pollution health effect than FEV1 and PEF rate. Larger percentage effects of air pollution on lung growth already have been observed for FEF2575 and FEF75 compared to PEF and FEV1 in population samples of children.15 For that reason, we feel that it is essential that epidemiologic and clinical evaluations of lung function evaluate a wider range of pulmonary function measures than has been the case up to now. Evaluating the cost-effectiveness of such a device is not within the scope of this report.
In general, electronic devices provide many benefits over paper diaries or manual recording of pulmonary function measures. The data values and timing of collection cannot be fabricated and are easily transferred to a personal computer, eliminating the cost and errors associated with manual data entry of paper diaries. This device also provides immediate suggestions to improve technique as well as displays an incentive scoring system to improve adherence.
This adherence component of this study allows only for the evaluation of a portable spirometer as a data collection tool. Children were blinded to the results, and monitoring was not performed in conjunction with a clinical or case-management program; therefore, the results may not be generalizable to those settings. The findings of this study are based on 2-week panels with twice-daily readings and therefore cannot assess adherence with longer-term monitoring.
| Conclusion |
|---|
|
|
|---|
| Acknowledgements |
|---|
| Footnotes |
|---|
The study was funded by the California Air Resources Board, Contract 99-322.
Received for publication April 30, 2002. Accepted for publication December 17, 2002.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. Zanconato, G. Meneghelli, R. Braga, F. Zacchello, E. Baraldi, and on behalf of the Working Group Office Spirometry in Primary Care Pediatrics: A Pilot Study Pediatrics, December 1, 2005; 116(6): e792 - e797. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Fonseca, A. Costa-Pereira, L. Delgado, L. N. Silva, M. Magalhaes, M. G. Castel-Branco, and M. Vaz Pulmonary Function Electronic Monitoring Devices: A Randomized Agreement Study Chest, September 1, 2005; 128(3): 1258 - 1265. [Abstract] [Full Text] [PDF] |
||||
![]() |
P Enright Spirometry should be measured in the offices of general practitioners Chronic Respiratory Disease, April 1, 2005; 2(2): 109 - 111. [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||