* From the Departments of Surgery (Drs. Van De Water and Dalton, and Mr. Miller and Mr. Mount) and Family Practice (Dr. Vogel), Mercer University School of Medicine, Macon, GA.
Correspondence to: Joseph M. Van De Water, MD, FCCP, Department of Surgery, Mercer University School of Medicine, Macon, GA 31201; e-mail: Vandewater.joseph{at}mccg.org
| Abstract |
|---|
|
|
|---|
Design: Prospective study.
Setting: A cardiovascular-thoracic surgery ICU in a community university-affiliated hospital.
Patients: Post-coronary artery bypass graft patients (n = 53) in whom 210 pairs of CO measurements were made.
Interventions: None.
Measurements and main results: The CO-ICG was determined simultaneously while the nurse was performing the CO-TD. Variability within the monitoring method was better for CO-ICG compared to CO-TD (6.3% vs 24.7%, respectively). The correlation, bias, and precision of the CO-ICG was good compared to CO-TD (r2 = 0.658; r = 0.811; bias, -0.17 L/min; precision, 1.09 L/min; CO-ICG = 1.00 x CO-TD - 0.17; p < 0.001). A steady improvement in agreement of the previous ICG methodologies compared to CO-TD was observed as follows: (1) CO-K: r2 = 0.309; r = 0.556; bias, -1.71 L/min; precision, 1.81 L/min; CO-K = 0.78 x CO-TD - 0.45; p < 0.001; (2) CO-S: r2 = 0.361; r = 0.601; bias, -1.46 L/min; precision, 1.63 L/min; CO-S = 0.80 x CO-TD - 0.36; p < 0.001; and (3) CO-SB: r2 = 0.469; r = 0.685; bias, -0.77 L/min; precision, 1.69 L/min; CO-SB = 1.03 x CO-TD - 0.95; p < 0.001. The CO-ICG demonstrated the closest agreement to CO-TD.
Conclusion: The latest ICG technology for determining CO (CO-ICG) is less variable and more reproducible in an intrapatient sense than is CO-TD, it is equivalent to the average accepted CO-TD in post-coronary artery bypass graft patients, and showed marked improvement in agreement with CO-TD compared to measurements made using previous generation ICG CO equations.
Key Words: bioimpedance cardiac output coronary artery bypass graft impedance cardiography pulmonary artery catheter thermodilution
| Introduction |
|---|
|
|
|---|
The most commonly used method of measuring CO in the ICU is with the indicator-dilution technique of thermodilution (CO-TD), which utilizes a pulmonary artery catheter (PAC) to estimate CO.2 3 However, the risks of estimating CO-TDs via PAC are well-documented to include infection, sepsis, and arrhythmias, as well as increased morbidity and mortality.4 5 6 As such, hospitals have reserved the technique for only the most critically ill patients or have discontinued using the PAC altogether. Furthermore, CO-TD estimations can be inaccurate and variable due to the assumptions used by the technology, operator technique, and/or patient factors.7 8 9 10 11 Clearly, there is a need for a low-cost, accurate, noninvasive alternative.
A noninvasive alternative method for determining CO and SVR is that of impedance cardiography (ICG). ICG measures the beat-to-beat changes of thoracic bioimpedance via four dual sensors applied on the neck and thorax in order to calculate stroke volume (SV). ICG was first advanced in 1940 by Nyboer et al,12
who, using the resistivity (
) of blood and the length (L) of the chest, established the relationship of impedance change (
Z) and base impedance (Zo) to the volume change (
V) of the tissue under measurement as follows:
![]() |
In the 1960s, Kubicek et al13
were commissioned by the National Aeronautics and Space Administration to develop a noninvasive method of determining CO. The result of that work was the development of the Minnesota Impedance Cardiograph and a new equation for SV that incorporated the maximum value of the first derivative of the impedance waveform (dZ/dtmax) and the left ventricular ejection time (LVET), as follows:
![]() |
In the 1980s, Sramek et al14
developed a less cumbersome ICG device with a new SV equation that substituted the cylindrical model of the chest used by Kubicek et al13
with that of a truncated cone. At the same time, L was approximated as 17% of the patients height (H). This yielded the following equation:
![]() |
In 1986, Bernstein15
modified the equation of Sramek et al14
by introducing the term
(the actual weight divided by the ideal weight), which accounted for deviations from ideal body weight as established by Metropolitan Life insurance tables. The purpose was to determine more accurately the volume of the thorax, as follows:
![]() |
Advancements in hardware and software with the latest generation ICG monitor (BioZ; CardioDynamics; San Diego, CA), including digital signal processing (DISQ digital impedance signal quantifier; CardioDynamics)16 and the creation of a proprietary modification to the Sramek-Bernstein equation (ZMARC impedance-modulating aortic compliance) have yielded significantly better results.17 18 19 20 These advancements and the experience of this investigator led to a desire to reevaluate ICG CO measurements using the ZMARC equation (CO-ICGs) in post-CABG patients comparing them to CO-TDs, as well as to those derived from predecessor ICG CO equations in the same patient population.
| Materials and Methods |
|---|
|
|
|---|
CO-ICGs were performed according to the manufacturers guidelines at times coinciding with the routine CO-TD obtained by the nursing staff during each thermodilution fluid bolus injection. As the nurse injected each fluid bolus (ie, a room temperature saline solution), attempting to obtain a minimum of three CO-TD values with a < 10% variation, we recorded the CO-ICG values. Unlike the CO-TD values, which are obtained with individual injections of saline solution, usually at various times during the respiratory cycle, the CO-ICG values were continually displayed and updated on the ICG monitor every 10 beats. In this manner, at least three pairs of CO measurements were obtained for each patient. The nurse-determined acceptable CO-TDs were averaged to obtain the final CO-TD for each patient. The CO-ICGs corresponding to the accepted CO-TDs also were averaged for each patient to obtain the final CO-ICG for each patient. All CO-TDs, including discarded measurements, were recorded for further analysis of variability associated with the thermodilution technique. All CO-ICGs also were recorded for variability analysis without any being either objectively or subjectively rejected.
In addition to displaying the CO-ICG values continuously, this ICG monitor also electronically recorded and stored other values (ie, heart rate, LVET, dZ/dtmax, and Zo) that allowed us to retrospectively calculate the ICG CO measurements via the previous-generation SV equations (ie, those by Kubicek [CO-K], Sramek [CO-S], and Sramek-Bernstein [CO-SB]). The CO-S (L = 0.17 x H) was used to calculate the thoracic length for the CO-K. Fifty-two of the 53 patients (36 men and 16 women) had data stored for further analysis due to the inadvertent deletion of one patient file from the ICG monitor.
The following statistical analyses for comparing the average CO-ICGs to CO-TDs were performed: Pearson correlation; regression analysis; and Bland-Altman analysis for bias and precision.21 In addition to comparing the paired averages for the two CO methods, intramethod variability also was analyzed by restricted maximum likelihood estimation and intramethod linear regression analysis.
| Results |
|---|
|
|
|---|
|
Intermethod Comparison
The results of the average CO-ICGs corresponding to the average of the accepted CO-TDs are listed in Table 2
. The results for Pearson correlation coefficient, p values, bias, SD, and linear regression analysis with slope and intercept are listed in Table 2
and are graphically represented in a scattergram in Figure 1
. When the 53 paired average CO-ICGs vs CO-TDs were tested with the Bland-Altman analysis,21
the bias and precision were -0.18 and 1.09 L/min, respectively, with a 95% level of agreement between -2.36 and 1.99 L/min (Fig 2
). All differences fell within the 95% upper and lower levels of agreement.
|
|
|
| Discussion |
|---|
|
|
|---|
CO is a prime example of such a clinical measurement. For example, CO may be calculated via any of the following independent physical measurements: the Fick method measures O2 concentration and consumption; the thermodilution technique measures temperature; and ICG measures bioimpedance. In this study, we wanted to determine whether CO-ICG was accurate enough to be used clinically in postoperative CABG patients. In other words, how well does CO-ICG agree with the reference method? To answer this question, we had to define the term accurate enough. Therefore, we evaluated the reference technology (in this case, CO-TD via PAC) to see how well intrapatient CO-TDs compared.
Intramethod Comparison
The difficulties in obtaining a reproducible and accurate CO-TD via PAC have been well-documented.7
8
9
10
11
This study experienced similar results. Comparisons of the first three serially measured CO-TDs resulted in a mean absolute difference between CO-TDs of 12% (SD, 13%). While the variation of the same paired CO-ICGs resulted in a lower mean absolute difference of 5.9% (SD, 6.5%). Linear regression analysis of the intramethod CO comparisons clearly reveals a wider variability and less agreement in the CO-TDs, compared to the case with the CO-ICGs (Table 1)
. While restricted maximum likelihood estimation produced a substantially larger variability (24.7%) for intramethod CO-TD compared to that for the intramethod CO-ICG (6.3%), it cannot be concluded that CO-TD is less accurate than CO-ICG. It does, however, indicate that the intrapatient precision of CO-ICG is better than that of CO-TD. Nevertheless, with so much intrapatient variation between the CO-TDs within a given patient, it is fair to ask how well any new method for measuring CO can compare to the CO-TD method, given its inherent intramethod variability.
Intermethod Comparison
When comparing CO-ICG to the clinically accepted value for CO-TD, we found that the bias, precision, correlation, slope, and intercept were equivalent to the intramethod CO-TD values when compared to themselves (Tables 1
and 2)
.
Retrospective Intermethod Comparison
Retrospective analysis and comparison of the previous and current ICG CO equations (ie, CO-K, CO-S, CO-SB, and CO-ICG) all displayed a progression of improved agreement with CO-TD for the patients in this study (Table 2)
. Each subsequent generation of method for ICG CO measurement resulted in a higher correlation, a slope more closely approaching unity, an intercept more closely approaching zero, less bias, and lower SD values than its predecessor.
| Conclusion |
|---|
|
|
|---|
Our retrospective intermethod comparison also confirmed the findings of other investigators that CO-K, CO-S, and CO-SB were not accurate enough to be used in assessing seriously ill patients. They also confirmed that clear advancements have been made with the ICG monitor used in this study, providing a level of agreement that is equivalent to CO-TD.
Currently, noninvasive pulse oximetry is considered to be a de facto vital sign by clinicians. While the accuracy of pulse oximetry has been established, it is not without its limitations. In addition, to date no conclusive outcome studies on prospective assessments, diagnoses, and treatments based on pulse oximetry values have been completed. Contrasted to oxygen saturation via pulse oximetry, CO and SVR are affected by a wide range of pharmacologic agents that are routinely prescribed for managing both acute and chronic diseases. In most cases, these agents are prescribed and the patients are treated without any direct, objective evaluation of the complete hemodynamic condition by the treating clinician. Based on our findings of the equivalent and clinically acceptable accuracy and reproducibility of the ICG monitor use in this study, we believe that there is substantial merit in declaring noninvasive CO to be a useful vital sign in the assessment, diagnosis, and treatment of postoperative CABG patients.
| Footnotes |
|---|
Received for publication June 11, 2002. Accepted for publication October 31, 2002.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A. Valipour, F. Schneider, W. Kossler, S. Saliba, and O. C. Burghuber Heart rate variability and spontaneous baroreflex sequences in supine healthy volunteers subjected to nasal positive airway pressure J Appl Physiol, December 1, 2005; 99(6): 2137 - 2143. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. R. Tomczak, W. Wojcik, E. F.G. Busse, and R. G. Haennel Phase I and Phase II Oxygen Uptake Kinetics During Atrioventricular Dyssynchrony in Chronotropically Competent Pacemaker Patients Chest, September 1, 2005; 128(3): 1782 - 1789. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. A. H. Critchley, Z. Y. Peng, B. S. Fok, and A. E. James The Effect of Peripheral Resistance on Impedance Cardiography Measurements in the Anesthetized Dog Anesth. Analg., June 1, 2005; 100(6): 1708 - 1712. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. A. Critchley, Z. Y. Peng, B. S. Fok, A. Lee, and R. A. Phillips Testing the Reliability of a New Ultrasonic Cardiac Output Monitor, the USCOM, by Using Aortic Flowprobes in Anesthetized Dogs Anesth. Analg., March 1, 2005; 100(3): 748 - 753. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Engoren and D. Barbee Comparison of Cardiac Output Determined by Bioimpedance, Thermodilution, and the Fick Method Am. J. Crit. Care., January 1, 2005; 14(1): 40 - 45. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. M. Albert, M. D. Hail, J. Li, and J. B. Young Equivalence of the Bioimpedance and Thermodilution Methods in Measuring Cardiac Output in Hospitalized Patients With Advanced, Decompensated Chronic Heart Failure Am. J. Crit. Care., November 1, 2004; 13(6): 469 - 479. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. M. Albert Ask the Experts Crit. Care Nurse, December 1, 2003; 23(6): 58 - 60. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||