Chest ACCP Member Benefits
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Guest Access | Sign In via User Name/Password
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (26)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hinz, J.
Right arrow Articles by Hedenstierna, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hinz, J.
Right arrow Articles by Hedenstierna, G.
(Chest. 2003;124:314-322.)
© 2003 American College of Chest Physicians

Regional Ventilation by Electrical Impedance Tomography*

A Comparison With Ventilation Scintigraphy in Pigs

José Hinz, MD; Peter Neumann, MD, PhD; Taras Dudykevych, DIng; Lars Goran Andersson, MD; Hermann Wrigge, MD; Hilmar Burchardi, MD, PhD and Goran Hedenstierna, MD, PhD, FCCP

* From the Department of Anesthesiology, Emergency, and Intensive Care Medicine (Drs. Hinz, Neumann, Dudykevych, and Burchardi), University of Göttingen, Göttingen, Germany; the Department of Clinical Physiology (Drs. Andersson and Hedenstierna), University of Uppsala, Uppsala, Sweden; and the Department of Anesthesiology and Intensive Care Medicine (Dr. Wrigge), University of Bonn, Bonn, Germany. This study was supported by grants from the Swedish Medical and Technical Research Councils (5315 and 299-2000-688), the Swedish Heart and Lung Foundation, and departmental funds.

Correspondence to: José Hinz, MD, Department of Anesthesiology, Emergency, and Intensive Care Medicine, Georg-August-University, Robert-Koch-Str 40, D-37075 Göttingen, Germany; e-mail: mail{at}josehinz.de


    Abstract
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusion
 References
 
Study objective: The validation of electrical impedance tomography (EIT) for measuring regional ventilation distribution by comparing it with single photon emission CT (SPECT) scanning.

Design: Randomized, prospective animal study.

Settings: Animal laboratories and nuclear medicine laboratories at a university hospital.

Participants: Twelve anesthetized and mechanically ventilated pigs.

Interventions: Lung injury was induced by central venous injection of oleic acid. Then pigs were randomized to pressure-controlled mechanical ventilation, airway pressure-release ventilation, or spontaneous breathing.

Measurements and results: Ventilation distribution was assessed by EIT using cross-sectional electrotomographic measurements of the thorax, and simultaneously by single SPECT scanning with the inhalation of 99mTc-labeled carbon particles. For both methods, the evaluation of ventilation distribution was performed in the same transverse slice that was approximately 4 cm in thickness. The transverse slice then was divided into 20 coronal segments (going from the sternum to the spine). We compared the percentage of ventilation in each segment, normalized to the entire ventilation in the observed slice. Our data showed an excellent linear correlation between the ventilation distribution measured by SPECT scanning and EIT according to the following equation: y = 0.82x + 0.7 (R2 = 0.92; range, 0.86 to 0.97).

Conclusion: Based on these data, EIT seems to allow, at least in comparable states of lung injury, real-time monitoring of regional ventilation distribution at the bedside.

Key Words: regional ventilation • thoracic electrical impedance tomography


    Introduction
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusion
 References
 
Electrical impedance tomography (EIT), which was developed in the early 1980s by Barber and Brown,1 generates cross-sectional images of the impedance distribution within a measured object. The basic principle of EIT is based on an alternate-current injection and voltage measurement via surface electrodes. If surface electrodes are placed around the thorax, changes in the electrical impedance during ventilatory maneuvers parallel to changes of aeration within the lungs can be measured. This enables the measurement of regional ventilation. EIT has been used increasingly as an experimental, noninvasive, lung-imaging technique.2 3 4 5 6 7 8 9 10 11 12 13 14 EIT has been shown to detect physiologic events related to anatomic settings.2 5 8 15 Validation of the technique by an established clinical method, however, has not been obtained. The aim of this study was therefore to validate the measurement of regional ventilation, assessed by a newly developed high-performance EIT device, with ventilation scintigraphy. Since EIT may have the potential to be a bedside technique in the intensive care setting, we performed the tests in a porcine lung damage model, during different ventilatory support modes.


    Materials and Methods
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusion
 References
 
Study Protocol
After approval by the local animal ethics committee, 12 pigs of mixed breed (ie, Hampshire, Yorkshire, and Swedish country breed) [mean (± SD) weight, 30 ± 4 kg] were anesthetized and mechanically ventilated. Acute lung injury was induced in all pigs by IV injections of oleic acid (Apoteksbolaget; Göteborg, Sweden) that was suspended in 20 mL isotonic saline solution. The dosing of the oleic acid was guided by intermittently taken arterial blood samples in order to achieve a PaO2/fraction of inspired oxygen (FIO2) ratio of about 200 mm Hg. After the induction of lung injury, a stabilization period of at least 2 h was allowed before we proceeded with the experimental protocol, so that lung injury was stable, as assessed by oxygenation and respiratory mechanics. Another hour later, the pigs were transferred to the nuclear medicine laboratories, the animals were placed in the supine position, with the front legs stretched cranially, and simultaneous measurements of regional ventilation distribution using EIT16 and single photon emission CT (SPECT) scanning17 were performed. The ventilation distribution was assessed in an approximately 4 cm-thick transverse slice of the lung. The slice then was divided into 20 equally thick coronal segments going from the sternum to the spine. For both methods, the percentage of ventilation in each coronal segment was calculated and compared, as shown in Figure 1 .



View larger version (20K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 1. Analysis procedure. For each pig, simultaneous transversal measurements from SPECT scanning and EIT were performed in a 4-cm slice of the thorax. Each slice was divided into 20 anterior-posterior segments (going from the sternum to the spine). Ventilation for both methods was calculated by the percentages of either regional minute volume for SPECT scanning or by the percentages of regional impedance change for EIT. The correlation of ventilation in each segment was calculated. Ä = {Delta}.

 
The investigation was performed in the experimental laboratories of the Department of Clinical Physiology at the University Hospital of Uppsala, Sweden.

Anesthesia
Azaperone (Stresnil; Janssen Pharmaceutica; Beerse, Belgium), 40 mg IM, were administered as premedication. General anesthesia was induced with remifentanil (Ultiva; Glaxo Wellcome; Hamburg, Germany) [0.04 mg/kg/min] and ketamine (Ketanest; Parke-Davis; Berlin, Germany), followed by a constant IV infusion of both drugs to maintain either controlled ventilation or spontaneous breathing. The animals received tracheotomies and were either spontaneously breathing or mechanically ventilated through a cuffed tube. Prior to measurements, 1,000 mL Ringer-acetate (Pharmacia AB; Stockholm, Sweden) at body temperature was infused.

Ventilation
Mechanical ventilation was provided with a ventilator (EVITA 4; Dräger; Lübeck, Germany). Respirator settings were initially in the pressure-controlled mode with a peak airway pressure of 15 cm H2O, a positive end-expiratory pressure of 5 cm H2O, a respiratory rate of 20 breaths/min, an inspiratory/expiratory ratio of 1:2, and an FIO2 of 0.5.

After oleic acid administration and a stabilization period of approximately 2 h, the pigs were randomized to pressure-controlled ventilation (PCV) without spontaneous breathing, PCV with spontaneous breathing (ie, airway pressure release ventilation [APRV]), or spontaneous breathing (ie, continuous positive airway pressure [CPAP]). The pigs also were studied for other purposes, which are reported on elsewhere. Another hour later, the pigs were moved to the gamma camera for the study, keeping the ventilator settings constant in the PCV group, and adjusting the ventilator support in the APRV and CPAP groups so that spontaneous breathing was resumed. Arterial blood gas samples were analyzed (ABL 300 and OSM 3 Hemoximeter; Radiometer; Copenhagen, Denmark).

Hemodynamics
For arterial blood gas sampling, an 18-gauge catheter was inserted into the carotid artery, together with a thermistor-tipped fiberoptic catheter (Pulsiocath 4F FT PV 2024; Pulsion Medical System; Munich, Germany), which was advanced into the descending aorta for measurements of extravascular lung water index (EVLWI). EVLWI was calculated automatically (Pulsion COLD Z-021; Pulsion Medical System) after injecting 8 to 10 mL 1 mg/mL indocyanine green (ICG-Pulsion; Pulsion Medical System), which was mixed in sterile water (temperature range, 5°C to 7°C) randomly within the respiratory cycles.18 19 The mean of triplicate measurements was used.

Compliance
Respiratory compliance was measured during controlled mechanical ventilation before and after the induction of acute lung injury. Airway pressure (Huba Control; Würenlos, Switzerland) was measured at the y-piece of the respirator. Tidal volume (VT) was calculated from gas flow measured by pneumotachography (flow head: Fleisch No. 2; Fleisch; Lausanne, Switzerland; differential pressure transducer: Huba Control). Data were sampled at 200 Hz and were stored for offline evaluation on a personal computer. Compliance was calculated according to the equation VT/(Pei - Pee), where Pei is the end-inspiratory pressure, and Pee is the end-expiratory pressure. Compliance was averaged over a 3-min recording period.

EIT
An applied potential tomography system (Goe-MF; EIT-Group Göttingen; Göttingen, Germany) was used for electrical impedance measurements. Sixteen commercially available radiograph transparent surface electrodes (Blue sensor BR-50-K; Medicotest A/S; Olstyke, Denmark) were placed around the thorax of the pigs at the level of the xiphoid process. For data collection, an alternating current (5 mA p-p, 50 kHz) was injected between one pair of adjacent electrodes. The resulting surface potentials were measured between the remaining adjacent electrode pairs. All 16 adjacent electrode pairs were used as the injecting electrodes, one pair after the other, with the surface potential being measured with the remaining electrodes. One data collection cycle was completed when all pairs of adjacent electrodes had been used once as injecting electrodes. The sampling rate of the used EIT device was 13 cycles per second, and 1,000 cycles were measured during each observation period. Thus, the mean acquisition time was about 77 s, corresponding to approximately 35 to 60 respiratory cycles, depending on the applied respiratory mode. According to the method described by Barber and Brown,20 the resulting 208 surface potentials were normalized to the mean surface potential during the measurement period, and the normalized surface potentials were used subsequently for the reconstruction of regional impedance changes by a back-projection. Functional EIT16 (f-EIT) was performed to determine the local lung impedance change as an indicator for regional ventilation. The generation of f-EIT is based on the assessment of the amplitude of impedance changes in every region of an EIT image. An example of an f-EIT image is shown in Figure 1 .

For further evaluation, this f-EIT image then was divided into 20 equally thick coronal segments going from the sternum to the spine. The impedance change in each segment as well as in the whole EIT image was calculated. Ventilation in each segment then was calculated as a percentage of the summarized change in impedance in each segment divided by the summarized impedance change of the whole EIT image. The data were stored for off-line evaluation on a personal computer.

SPECT Scanning
Ventilation distribution was assessed by SPECT scanning using 99mTc-labeled carbon particles (Technegas; Tetley Medical Limited; Lucas Heights, NSW, Australia). The particle size is approximately 0.1 µm, and its distribution in the lung has been shown to be similar to that of radioactive gas.17 This "pseudogas" was injected continuously during mechanical ventilation over a period of 3 to 5 min from a 2-L syringe that was connected to the Y-piece. During the injection of the carbon particles, a pressure of approximately 31 cm H2O was maintained within the syringe, as controlled by a manometer connected to the syringe via a three-way stopcock, resulting into a continuous flow of carbon particles into the Y-piece during the whole respiratory cycles. Images were acquired on a dual-head gamma camera (Maxxus; General Electric Systems; Milwaukee, WI) that was equipped with all-purpose low-energy collimators. The SPECT acquisitions were performed in 64 projections (32 projections per head) and were stored in a 64 x 64 matrix. The acquisition time was 15 s for each projection.

The data were reconstructed on a computer workstation (HERMES; Nuclear Diagnostics; Stockholm, Sweden). The acquired data were prefiltered with a two-dimensional Butterworth filter (cutoff frequency, 0.14; filter order, 10). Filtered back-projection reconstruction was performed without applying attenuation correction. After corrections for the background, the number of counts was measured in each volume element of the lungs.

Minute ventilation was assessed by pneumotachography (as described above). Ventilation in the 4 cm-thick slice near the EIT electrodes was calculated as the number of counts within the slice, times minute ventilation, divided by the total number of counts within the whole lung. By adding coronal planes, the slices were divided into 20 equally thick segments in the ventral-dorsal direction going from the sternum to the spine. The ventilation of each segment finally was calculated as the fractional activity in each segment times minute ventilation.

Since animals with varying degrees of pulmonary edema and damage were studied, the risk of carbon particles being trapped in the edematous airways was considered to be possible. The intrapulmonary distributions of the carbon particles therefore were compared with a reference technique, the inhalation of radioactive krypton gas, in four pigs. 81mKr was produced as a gas in a 81Rb/81mKr generator (KryptoScan; Mallinckrodt Medical BV; Petten, the Netherlands). The krypton was continuously administered directly into the ventilator during the whole acquisition time. The scintigraphic acquisition of carbon particles and 81mKr was performed simultaneously in two separate windows of 140 keV (± 10%) and 186 keV (± 10%), respectively, for technetium and krypton. Except for an extension of the acquisition time per projection to 35 s, the same camera and acquisition settings were used as described above. After correction for crossover into the 140-keV window, reconstruction was performed as described above.

Statistical Analysis
Calculations were performed with a statistical software package (Statistica, version 5.1; StatSoft Inc; Tulsa, OK) on a personal computer (Pentium II 233 MHz processor; Intel; Santa Clara, CA; and Windows 95; Microsoft; Redmond, WA). All data are presented as the minimum to maximum (median). A p value of < 0.05 was chosen as the level of significance.

The Pearson correlation analysis was used to calculate the correlation between the ventilation distribution assessed by the two methods. In addition, EIT and SPECT data were compared using the statistical analysis method of Bland and Altman.21 Bias was defined by the mean differences between EIT and SPECT, and it represented the systematic error. The SD of the mean differences is considered to represent the random error variability between the two techniques. Bias ± 2 SDs was the limit of agreement of both methods.


    Results
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusion
 References
 
Four pigs each were ventilated in the PCV, APRV, and CPAP modes, and the median values of the PaO2/FIO2 ratio were 176, 197, and 212 mm Hg, respectively. Respirator settings, blood gas data, and respiratory compliance are summarized in Table 1 .


View this table:
[in this window]
[in a new window]

 
Table 1. Results of Randomization to Various Respiratory Modes*

 
Correlation of Inhaled Krypton and Nebulized Carbon Particles
A good correlation between the distributions of inhaled krypton and nebulized carbon particles was obtained, according to the following equation: VregTc = 0.97 x VregKr + 4.3 (R2 = 0.98; mean Vreg, approximately 350 mL), where Vreg is registered volume, Tc is technetium, and Kr is krypton. This was considered proof that the carbon particles could be used for the subsequent comparisons between EIT and SPECT could also be use in the present pigs with oleic acid-induced pulmonary edema.

Pooled Data for Regional Ventilation Distribution
A highly significant linear correlation between regional ventilation measured by EIT and SPECT scanning was found, according to the equation y = 0.82x + 0.73 (R2 =0.92; range, 0.86 to 0.97) [Fig 2 ]. However, the regression coefficient was < 1.0 (p < 0.001), and we found a positive intercept (p < 0.001).



View larger version (18K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 2. Linear correlation. Comparison of regional lung ventilation measured with EIT and SPECT scanning in 240 regions of interests in 12 pigs. Values given as the percentage of ventilation assessed by EIT and SPECT scanning.

 
As can be seen from a Bland-Altman analysis, EIT tended to overestimate ventilation in regions that were poorly ventilated and to underestimate ventilation in well-ventilated regions (Fig 3 ). Thus, the magnitude of the regional ventilation affected the correlation between EIT and SPECT scanning values. However, the difference between EIT and SPECT scanning did not exceed 10% and was mostly much less than that.



View larger version (17K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 3. Bland-Altman analysis. Comparison of regional lung ventilation measured with EIT and SPECT scanning in 240 regions of interests in 12 pigs. Values given as the percentage of ventilation assessed by EIT and SPECT scanning.

 
Influence of Spontaneous and Mechanical Ventilation
The same correlation between EIT and SPECT data was seen whether the pigs were receiving continuous mechanical ventilation (ie, PCV) or were breathing spontaneously (ie, CPAP). A combination of mechanical and spontaneous ventilation (approximate ratio, 80:20) [ie, APRV] also resulted in a similar correlation (Fig 4 ). Thus, the respiratory mode had no effect on the relationship between EIT and SPECT scanning. Figure 5 shows the spatial ventilation distribution in the dorsoventral orientation (going from the sternum to the spine) of the different ventilatory modes as examined with EIT and SPECT scanning.



View larger version (17K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 4. Bland-Altman analysis. Separate analysis of regional lung ventilation from pigs that were breathing spontaneously (ie, CPAP), that were mechanically ventilated (ie, PCV), and that breathed APRV. Eighty regions of interests were measured in four pigs. The linear correlation is given in the plot. Values given as the percentage of ventilation assessed by EIT and SPECT scanning.

 


View larger version (54K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 5. Comparison of ventilation distributions in different ventilatory modes in dorsal-to-ventral orientation of a transversal plane at the processus xiphoideus measured by SPECT scanning and EIT. Twelve pigs with oleic acid-induced lung injury were studied. Four pigs were mechanically ventilated (ie, PCV), four pigs were mechanically ventilated with spontaneous breathing (ie, APRV), and four pigs breathed spontaneously (ie, CPAP).

 
Influence of Lung Edema, Minute Ventilation, and PaCO2
Since the distribution of the oleic acid-induced edema is gravity-dependent,22 dorsal lung regions are more severely affected than anterior regions. Therefore, an analysis was made about whether this (ie, the upper, healthier lung regions vs the lower, sicker lung regions) had an effect on the relationship between EIT and SPECT scanning, but no such effect was found (Fig 6 ). Moreover, the compliance of the respiratory system as well as of the EVLWI were compared with the correspondence between EIT and SPECT scanning. No correlation was found between compliance or ELVWI, either in absolute numbers or in terms of the relative change from baseline, and the difference between EIT and SPECT scanning.



View larger version (23K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 6. Bland-Altman analysis. Separate analyses of dorsal and ventral regions from 12 pigs were made. Lung ventilation in 120 regions of interests in 12 pigs was measured. The linear correlation is given in the plot. Values given as the percentage of ventilation assessed by EIT and SPECT scanning.

 
Minute ventilation and PaCO2 varied considerably between pigs, with several pigs displaying hypoventilation as a consequence of the oleic acid-induced lung damage and the use of spontaneous breathing modes (Table 1) . This may have been due to increased dead space ventilation during spontaneous breathing, and it necessitated an analysis of the influence of minute ventilation and PaCO2 on the performance of EIT. However, no correlation between minute ventilation, or PaCO2, and the difference between EIT and SPECT scanning was seen. Thus, the accuracy of measuring regional ventilation by EIT was not affected by the degree of ventilation.


    Discussion
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusion
 References
 
EIT is a noninvasive technique with the potential to monitor regional ventilation distribution. Since its introduction, the hardware and software have been improved continuously.23 24 Hahn and coworkers25 developed more advanced algorithms for the analysis of dynamic physiologic phenomena with low amplitudes than was initially supplied (Mark I or DAS-01P; Royal Hallamshire Hospital; Sheffield, UK). Thus, they introduced the f-EIT and the averaging technique.16 These improvements, together with other refinements of the technique, have resulted in a higher cycle rate and a better signal-to-noise ratio, which enable the system to monitor impedance changes during for example respiration. The major advantages of EIT are thereby that it is noninvasive, it is easy to use at the bedside, and data collection can be performed with a high time resolution.

EIT vs SPECT
The validation of the EIT technique has so far been based on correlation analysis with global parameters or on the reproduction of known physiologic phenomena or anatomic settings.2 3 4 5 11 15 16 A partitioning of ventilation between the right and left lung by EIT and a comparison with radionuclide scanning was performed by Kunst and coworkers.12 However, they needed to average 10 data collection cycles to obtain one EIT image. Although the time resolution was rather poor, they found a good correlation between the EIT and radionuclide methods for separate lungs. With the advanced hardware and software9 25 used in the present study, the technique also should be suitable for the regional analysis of ventilation within the lungs. We obtained an excellent linear correlation between f-EIT and ventilation scintigraphy (ie, SPECT scanning). Whether breathing was spontaneous or mechanically delivered did not affect the correlation. Moreover, we did not see any influence of the severity of lung damage in this correlation on a global level (ie, no correlation to respiratory compliance and EVLWI) and on a regional level (ie, no difference in the relationship between EIT and SPECT in dependent, more severely affected lung regions or upper, nondependent regions).

However, a regression coefficient of 0.82 and the positive intercept between SPECT scanning (independent variable) and EIT (dependent variable) showed that EIT slightly underestimated ventilation compared to SPECT scanning in rather well- ventilated regions and overestimated it in poorly ventilated areas. The difference in regional ventilation as measured by both methods was < 10%, and this is comparable to the coefficient of variation of most physiologic methods. The EIT method thus seems to be acceptable as a monitoring device in, for example, the intensive care setting. Potential causes of the measured differences between EIT and SPECT are discussed in the following paragraph.

Linearity of EIT
Holder and Khan26 determined the linear operating range of the impedance change caused by various polyacrylamide gels with different impedance characteristics. As a physiologic model, they used a saline solution-filled tank and an EIT system (APT System Mark I; IBEES; Sheffield, UK). Impedance changes increased in a linear fashion up to 20%. Impedance changes of > 20% were underestimated by the EIT system. This effect may depend on the electrical properties of the EIT system27 or on the image reconstruction algorithm,28 which is the same as the one that we have used. The algorithm for image reconstruction is based on the assumptions that the object is two-dimensional and circular. Furthermore, the distribution of resistivity initially has to be uniform, the changes in resistivity should be small, and the electrodes should be spaced equally around the thorax. Nevertheless, EIT images can be acquired even if these assumptions are violated. The respiratory patterns in the present study included VT values <= 15 mL/kg, which confined impedance changes to a maximum of 25%. This may be the reason for the underestimation of larger regional volume changes, as seen in the present study.

Possibilities and Limitations of EIT
Value of Information:
Although we calculated only 20 regions of interest from one f-EIT image, the spatial resolution allows the calculation of a maximum of 912 regions of interest per image. According to Hahn et al,29 the minimal detectable lung volume by EIT is in the range of 9 to 29 mL, whereas the spatial resolution using SPECT scanning is approximately 1 mL. Thus, regional ventilation of well-ventilated lung areas, which are located close to poorly or nonventilated lung regions, may be underestimated by EIT and vice versa.

By measuring EIT over time, this spatial resolution combined with the time resolution of <= 44 Hz offers the possibility of following regional ventilation. A lower sampling rate of 1 Hz is sufficient to monitor slow events like stepwise changes in lung volume during breath-holding procedures.30 Higher sampling rates may be suitable for rapid physiologic events, like a forced expiration in spontaneously breathing subjects.

Since EIT images are generated by relatively small local impedance changes, whereas absolute impedance values may vary considerably among different subjects, EIT is most suitable for intraindividual monitoring, with each subject serving as his own reference.

Clinical Application:
As it is noninvasive and easily transportable, EIT might be suitable for monitoring regional lung function in mechanically ventilated patients at the bedside. It may be used to acquire information, such as regional lung ventilation, regional distribution of VT, regional lung volume, and functional residual capacity.31 Furthermore, fluid accumulation, redistribution of ventilation inside the lungs,32 and regional compliance curves11 may be reconstructed from EIT data. Thus, EIT could be used, for example, to monitor atelectasis and hyperinflation during different ventilatory strategies or in opening procedures. In patients with COPD, EIT measurement of regional lung function during therapeutic interventions such as bronchodilator administration or noninvasive ventilation also might be possible. In neonates, for whom radiation exposure is undesirable, EIT potentially may replace radiographic investigations in certain situations.7 However, EIT should not be applied to obtain morphologic information similar to CT scanning or MRI, since the latter diagnostic methods provide anatomic information with a much higher spatial and morphologic resolution compared to EIT.


    Conclusion
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusion
 References
 
EIT data of regional ventilation correlated closely to ventilation distribution by SPECT scanning in a porcine, oleic acid-lung damage model. Whether breathing was spontaneous or mechanical had no effect on the relationship between EIT and SPECT scanning, and the severity of lung damage on a global or a regional level did not influence the relationship either. A slight overestimation of ventilation in well-ventilated areas and a slight underestimation in poorly ventilated areas was seen, but the difference was always < 10% of ventilation measured by SPECT scanning. Thus, EIT appears to be an interesting monitoring technique for use in the intensive care setting.


    Footnotes
 
Abbreviations: APRV = airway pressure release ventilation; CPAP = continuous positive airway pressure; EIT = electrical impedance tomography; f-EIT = functional electrical impedance tomography; EVLWI = extravascular lung water index; FIO2 = fraction of inspired oxygen; PCV = pressure-controlled ventilation; SPECT = single photon emission CT; Vreg = registered volume; VT = tidal volume

Received for publication July 11, 2002. Accepted for publication January 29, 2003.


    References
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusion
 References
 

  1. Barber, DC, Brown, BH (1984) Applied potential tomography. J Phys [E] 17,723-733[CrossRef]
  2. Adler, A, Amyot, R, Guardo, R, et al Monitoring changes in lung air and liquid volumes with electrical impedance tomography. J Appl Physiol 1997;83,1762-1767[Abstract/Free Full Text]
  3. Frerichs, I, Hahn, G, Golisch, W, et al Monitoring perioperative changes in distribution of pulmonary ventilation by functional electrical impedance tomography. Acta Anaesthesiol Scand 1998;42,721-726[ISI][Medline]
  4. Frerichs, I, Hahn, G, Hellige, G Gravity-dependent phenomena in lung ventilation determined by functional EIT. Physiol Meas 1996;17(suppl),A149-A157
  5. Frerichs, I, Hahn, G, Hellige, G Thoracic electrical impedance tomographic measurements during volume controlled ventilation-effects of tidal volume and positive end-expiratory pressure. IEEE Trans Med Imaging 1999;18,764-773[CrossRef][ISI][Medline]
  6. Frerichs, I, Hahn, G, Hellige, G Bestimmung der regionalen Ventilation bei einseitiger Lungenschädigung mittels funktioneller elektrischer Impedanztomographie. Intensiv und Notfallbehandlung 1999;24,87-92
  7. Frerichs, I, Hahn, G, Schiffmann, H, et al Monitoring regional lung ventilation by functional electrical impedance tomography during assisted ventilation. Ann N Y Acad Sci 1999;873,493-505[Abstract/Free Full Text]
  8. Frerichs, I, Hahn, G, Schroder, T, et al Electrical impedance tomography in monitoring experimental lung injury. Intensive Care Med 1998;24,829-836[CrossRef][ISI][Medline]
  9. Hahn, G, Frerichs, I, Golisch, W, et al Determining local lung ventilation by functional electrical impedance tomography under clinical circumstances. Biomed Tech (Berl) 1997;42 (Suppl),213-214
  10. Hahn, G, Frerichs, I, Kleyer, M, et al Local mechanics of the lung tissue determined by functional EIT. Physiol Meas 1996;17(suppl),A159-A166
  11. Kunst, PW, Bohm, SH, de Vazquez, A, et al Regional pressure volume curves by electrical impedance tomography in a model of acute lung injury. Crit Care Med 2000;28,178-183[CrossRef][ISI][Medline]
  12. Kunst, PW, Vonk, NA, Hoekstra, OS, et al Ventilation and perfusion imaging by electrical impedance tomography: a comparison with radionuclide scanning. Physiol Meas 1998;19,481-490[CrossRef][ISI][Medline]
  13. Kunst, PW, Vonk, NA, Straver, B, et al Influences of lung parenchyma density and thoracic fluid on ventilatory EIT measurements. Physiol Meas 1998;19,27-34[CrossRef][ISI][Medline]
  14. Kunst, PW, de Vries, PM, Postmus, PE, et al Evaluation of electrical impedance tomography in the measurement of PEEP-induced changes in lung volume. Chest 1999;115,1102-1106[Abstract/Free Full Text]
  15. Adler, A, Shinozuka, N, Berthiaume, Y, et al Electrical impedance tomography can monitor dynamic hyperinflation in dogs. J Appl Physiol 1998;84,726-732[Abstract/Free Full Text]
  16. Hahn, G, Sipinkova, I, Baisch, F, et al Changes in the thoracic impedance distribution under different ventilatory conditions. Physiol Meas 1995;16(suppl),A161-A173[CrossRef][ISI][Medline]
  17. Burch, WM, Sullivan, PJ, Lomas, FE, et al Lung ventilation studies with technetium-99m Pseudogas. J Nucl Med 1986;27,842-846[Abstract/Free Full Text]
  18. Lichtwarck-Aschoff, M, Beale, R, Pfeiffer, UJ Central venous pressure, pulmonary artery occlusion pressure, intrathoracic blood volume, and right ventricular end-diastolic volume as indicators of cardiac preload. J Crit Care 1996;11,180-188[CrossRef][ISI][Medline]
  19. Lichtwarck-Aschoff, M, Zeravik, J, Pfeiffer, UJ Intrathoracic blood volume accurately reflects circulatory volume status in critically ill patients with mechanical ventilation. Intensive Care Med 1992;18,142-147[CrossRef][ISI][Medline]
  20. Barber, DC, Brown, BH Recent developments in applied potential tomography-APT. Bacharach, SL eds. Proceedings of the 9th Conference of Information Processing in Medical Imaging 1986,106-121 Martinus Nijhoff. Dordrecht, the Netherlands:
  21. Bland, JM, Altman, DG Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1,307-310[CrossRef][ISI][Medline]
  22. Neumann, P, Berglund, JE, Mondejar, EF, et al Effect of different pressure levels on the dynamics of lung collapse and recruitment in oleic-acid-induced lung injury. Am J Respir Crit Care Med 1998;158,1636-1643[Abstract/Free Full Text]
  23. Koukourlis, CS, Kyriacou, GA, Sahalos, JN A 32-electrode data collection system for electrical impedance tomography. IEEE Trans Biomed Eng 1995;42,632-636[CrossRef][ISI][Medline]
  24. Li, JH, Joppek, C, Faust, U Fast EIT data acquisition system with active electrodes and its application to cardiac imaging. Physiol Meas 1996;17(suppl),A25-A32
  25. Hahn, G, Beer, M, Frerichs, I, et al A simple method to check the dynamic performance of electrical impedance tomography systems. Physiol Meas 2000;21,53-60[CrossRef][ISI][Medline]
  26. Holder, DS, Khan, A Use of polyacrylamide gels in a saline-filled tank to determine the linearity of the Sheffield Mark 1 electrical impedance tomography (EIT) system in measuring impedance disturbances. Physiol Meas 1994;15(suppl),A45-A50
  27. Brown, BH, Seagar, AD The Sheffield data collection system. Clin Phys Physiol Meas 1987;8,91-97
  28. Barber, DC Quantification in impedance imaging. Clin Phys Physiol Meas 1990;11(suppl),45-56
  29. Hahn, G, Hartung, C, Hellige, G Bestimmung der grösse minimal erfassbarer areale mit ventilationsstörungen. Thews, G eds. Elektrische impedanztomographie (EIT) als methode zur regionalen beurteilung der lungenventilation 1998,77 Gustav Fischer Verlag. Mainz, Germany:
  30. Harris, ND, Suggett, AJ, Barber, DC, et al Applications of applied potential tomography (APT) in respiratory medicine. Clin Phys Physiol Meas 1987;8(suppl),155-165
  31. Hinz, J, Hahn, G, Neumann, P, et al End-expiratory lung impedance change enables bedside monitoring of end- expiratory lung volume change. Intensive Care Med 2003;29,37-43[ISI][Medline]
  32. Hinz, J, Hahn, G, Sydow, M, et al Bed-side measurement of PEEP effects on ventilation distribution by f-EIT: preliminary results [abstract]. Intensive Care Med 1998;24(suppl),66



This article has been cited by other articles:


Home page
Br J AnaesthHome page
M. Nirmalan, M. Niranjan, T. Willard, J. D. Edwards, R. A. Little, and P. M. Dark
Estimation of errors in determining intrathoracic blood volume using thermal dilution in pigs with acute lung injury and haemorrhage
Br. J. Anaesth., October 1, 2004; 93(4): 546 - 551.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Respir. Crit. Care Med.Home page
G. Hedenstierna
Using Electric Impedance Tomography to Assess Regional Ventilation at the Bedside
Am. J. Respir. Crit. Care Med., April 1, 2004; 169(7): 777 - 778.
[Full Text] [PDF]


Home page
Am. J. Respir. Crit. Care Med.Home page
J. A. Victorino, J. B. Borges, V. N. Okamoto, G. F. J. Matos, M. R. Tucci, M. P. R. Caramez, H. Tanaka, F. S. Sipmann, D. C. B. Santos, C. S. V. Barbas, et al.
Imbalances in Regional Lung Ventilation: A Validation Study on Electrical Impedance Tomography
Am. J. Respir. Crit. Care Med., April 1, 2004; 169(7): 791 - 800.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (26)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hinz, J.
Right arrow Articles by Hedenstierna, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hinz, J.
Right arrow Articles by Hedenstierna, G.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS