* From the Department of Anesthesiology, Emergency, and Intensive Care Medicine (Drs. Hinz, Neumann, Dudykevych, and Burchardi), University of Göttingen, Göttingen, Germany; the Department of Clinical Physiology (Drs. Andersson and Hedenstierna), University of Uppsala, Uppsala, Sweden; and the Department of Anesthesiology and Intensive Care Medicine (Dr. Wrigge), University of Bonn, Bonn, Germany. This study was supported by grants from the Swedish Medical and Technical Research Councils (5315 and 299-2000-688), the Swedish Heart and Lung Foundation, and departmental funds.
Correspondence to: José Hinz, MD, Department of Anesthesiology, Emergency, and Intensive Care Medicine, Georg-August-University, Robert-Koch-Str 40, D-37075 Göttingen, Germany; e-mail: mail{at}josehinz.de
| Abstract |
|---|
|
|
|---|
Design: Randomized, prospective animal study.
Settings: Animal laboratories and nuclear medicine laboratories at a university hospital.
Participants: Twelve anesthetized and mechanically ventilated pigs.
Interventions: Lung injury was induced by central venous injection of oleic acid. Then pigs were randomized to pressure-controlled mechanical ventilation, airway pressure-release ventilation, or spontaneous breathing.
Measurements and results: Ventilation distribution was assessed by EIT using cross-sectional electrotomographic measurements of the thorax, and simultaneously by single SPECT scanning with the inhalation of 99mTc-labeled carbon particles. For both methods, the evaluation of ventilation distribution was performed in the same transverse slice that was approximately 4 cm in thickness. The transverse slice then was divided into 20 coronal segments (going from the sternum to the spine). We compared the percentage of ventilation in each segment, normalized to the entire ventilation in the observed slice. Our data showed an excellent linear correlation between the ventilation distribution measured by SPECT scanning and EIT according to the following equation: y = 0.82x + 0.7 (R2 = 0.92; range, 0.86 to 0.97).
Conclusion: Based on these data, EIT seems to allow, at least in comparable states of lung injury, real-time monitoring of regional ventilation distribution at the bedside.
Key Words: regional ventilation thoracic electrical impedance tomography
| Introduction |
|---|
|
|
|---|
| Materials and Methods |
|---|
|
|
|---|
|
Anesthesia
Azaperone (Stresnil; Janssen Pharmaceutica; Beerse, Belgium), 40 mg IM, were administered as premedication. General anesthesia was induced with remifentanil (Ultiva; Glaxo Wellcome; Hamburg, Germany) [0.04 mg/kg/min] and ketamine (Ketanest; Parke-Davis; Berlin, Germany), followed by a constant IV infusion of both drugs to maintain either controlled ventilation or spontaneous breathing. The animals received tracheotomies and were either spontaneously breathing or mechanically ventilated through a cuffed tube. Prior to measurements, 1,000 mL Ringer-acetate (Pharmacia AB; Stockholm, Sweden) at body temperature was infused.
Ventilation
Mechanical ventilation was provided with a ventilator (EVITA 4; Dräger; Lübeck, Germany). Respirator settings were initially in the pressure-controlled mode with a peak airway pressure of 15 cm H2O, a positive end-expiratory pressure of 5 cm H2O, a respiratory rate of 20 breaths/min, an inspiratory/expiratory ratio of 1:2, and an FIO2 of 0.5.
After oleic acid administration and a stabilization period of approximately 2 h, the pigs were randomized to pressure-controlled ventilation (PCV) without spontaneous breathing, PCV with spontaneous breathing (ie, airway pressure release ventilation [APRV]), or spontaneous breathing (ie, continuous positive airway pressure [CPAP]). The pigs also were studied for other purposes, which are reported on elsewhere. Another hour later, the pigs were moved to the gamma camera for the study, keeping the ventilator settings constant in the PCV group, and adjusting the ventilator support in the APRV and CPAP groups so that spontaneous breathing was resumed. Arterial blood gas samples were analyzed (ABL 300 and OSM 3 Hemoximeter; Radiometer; Copenhagen, Denmark).
Hemodynamics
For arterial blood gas sampling, an 18-gauge catheter was inserted into the carotid artery, together with a thermistor-tipped fiberoptic catheter (Pulsiocath 4F FT PV 2024; Pulsion Medical System; Munich, Germany), which was advanced into the descending aorta for measurements of extravascular lung water index (EVLWI). EVLWI was calculated automatically (Pulsion COLD Z-021; Pulsion Medical System) after injecting 8 to 10 mL 1 mg/mL indocyanine green (ICG-Pulsion; Pulsion Medical System), which was mixed in sterile water (temperature range, 5°C to 7°C) randomly within the respiratory cycles.18
19
The mean of triplicate measurements was used.
Compliance
Respiratory compliance was measured during controlled mechanical ventilation before and after the induction of acute lung injury. Airway pressure (Huba Control; Würenlos, Switzerland) was measured at the y-piece of the respirator. Tidal volume (VT) was calculated from gas flow measured by pneumotachography (flow head: Fleisch No. 2; Fleisch; Lausanne, Switzerland; differential pressure transducer: Huba Control). Data were sampled at 200 Hz and were stored for offline evaluation on a personal computer. Compliance was calculated according to the equation VT/(Pei - Pee), where Pei is the end-inspiratory pressure, and Pee is the end-expiratory pressure. Compliance was averaged over a 3-min recording period.
EIT
An applied potential tomography system (Goe-MF; EIT-Group Göttingen; Göttingen, Germany) was used for electrical impedance measurements. Sixteen commercially available radiograph transparent surface electrodes (Blue sensor BR-50-K; Medicotest A/S; Olstyke, Denmark) were placed around the thorax of the pigs at the level of the xiphoid process. For data collection, an alternating current (5 mA p-p, 50 kHz) was injected between one pair of adjacent electrodes. The resulting surface potentials were measured between the remaining adjacent electrode pairs. All 16 adjacent electrode pairs were used as the injecting electrodes, one pair after the other, with the surface potential being measured with the remaining electrodes. One data collection cycle was completed when all pairs of adjacent electrodes had been used once as injecting electrodes. The sampling rate of the used EIT device was 13 cycles per second, and 1,000 cycles were measured during each observation period. Thus, the mean acquisition time was about 77 s, corresponding to approximately 35 to 60 respiratory cycles, depending on the applied respiratory mode. According to the method described by Barber and Brown,20
the resulting 208 surface potentials were normalized to the mean surface potential during the measurement period, and the normalized surface potentials were used subsequently for the reconstruction of regional impedance changes by a back-projection. Functional EIT16
(f-EIT) was performed to determine the local lung impedance change as an indicator for regional ventilation. The generation of f-EIT is based on the assessment of the amplitude of impedance changes in every region of an EIT image. An example of an f-EIT image is shown in Figure 1
.
For further evaluation, this f-EIT image then was divided into 20 equally thick coronal segments going from the sternum to the spine. The impedance change in each segment as well as in the whole EIT image was calculated. Ventilation in each segment then was calculated as a percentage of the summarized change in impedance in each segment divided by the summarized impedance change of the whole EIT image. The data were stored for off-line evaluation on a personal computer.
SPECT Scanning
Ventilation distribution was assessed by SPECT scanning using 99mTc-labeled carbon particles (Technegas; Tetley Medical Limited; Lucas Heights, NSW, Australia). The particle size is approximately 0.1 µm, and its distribution in the lung has been shown to be similar to that of radioactive gas.17
This "pseudogas" was injected continuously during mechanical ventilation over a period of 3 to 5 min from a 2-L syringe that was connected to the Y-piece. During the injection of the carbon particles, a pressure of approximately 31 cm H2O was maintained within the syringe, as controlled by a manometer connected to the syringe via a three-way stopcock, resulting into a continuous flow of carbon particles into the Y-piece during the whole respiratory cycles. Images were acquired on a dual-head gamma camera (Maxxus; General Electric Systems; Milwaukee, WI) that was equipped with all-purpose low-energy collimators. The SPECT acquisitions were performed in 64 projections (32 projections per head) and were stored in a 64 x 64 matrix. The acquisition time was 15 s for each projection.
The data were reconstructed on a computer workstation (HERMES; Nuclear Diagnostics; Stockholm, Sweden). The acquired data were prefiltered with a two-dimensional Butterworth filter (cutoff frequency, 0.14; filter order, 10). Filtered back-projection reconstruction was performed without applying attenuation correction. After corrections for the background, the number of counts was measured in each volume element of the lungs.
Minute ventilation was assessed by pneumotachography (as described above). Ventilation in the 4 cm-thick slice near the EIT electrodes was calculated as the number of counts within the slice, times minute ventilation, divided by the total number of counts within the whole lung. By adding coronal planes, the slices were divided into 20 equally thick segments in the ventral-dorsal direction going from the sternum to the spine. The ventilation of each segment finally was calculated as the fractional activity in each segment times minute ventilation.
Since animals with varying degrees of pulmonary edema and damage were studied, the risk of carbon particles being trapped in the edematous airways was considered to be possible. The intrapulmonary distributions of the carbon particles therefore were compared with a reference technique, the inhalation of radioactive krypton gas, in four pigs. 81mKr was produced as a gas in a 81Rb/81mKr generator (KryptoScan; Mallinckrodt Medical BV; Petten, the Netherlands). The krypton was continuously administered directly into the ventilator during the whole acquisition time. The scintigraphic acquisition of carbon particles and 81mKr was performed simultaneously in two separate windows of 140 keV (± 10%) and 186 keV (± 10%), respectively, for technetium and krypton. Except for an extension of the acquisition time per projection to 35 s, the same camera and acquisition settings were used as described above. After correction for crossover into the 140-keV window, reconstruction was performed as described above.
Statistical Analysis
Calculations were performed with a statistical software package (Statistica, version 5.1; StatSoft Inc; Tulsa, OK) on a personal computer (Pentium II 233 MHz processor; Intel; Santa Clara, CA; and Windows 95; Microsoft; Redmond, WA). All data are presented as the minimum to maximum (median). A p value of < 0.05 was chosen as the level of significance.
The Pearson correlation analysis was used to calculate the correlation between the ventilation distribution assessed by the two methods. In addition, EIT and SPECT data were compared using the statistical analysis method of Bland and Altman.21 Bias was defined by the mean differences between EIT and SPECT, and it represented the systematic error. The SD of the mean differences is considered to represent the random error variability between the two techniques. Bias ± 2 SDs was the limit of agreement of both methods.
| Results |
|---|
|
|
|---|
|
Pooled Data for Regional Ventilation Distribution
A highly significant linear correlation between regional ventilation measured by EIT and SPECT scanning was found, according to the equation y = 0.82x + 0.73 (R2 =0.92; range, 0.86 to 0.97) [Fig 2 ]. However, the regression coefficient was < 1.0 (p < 0.001), and we found a positive intercept (p < 0.001).
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
EIT vs SPECT
The validation of the EIT technique has so far been based on correlation analysis with global parameters or on the reproduction of known physiologic phenomena or anatomic settings.2
3
4
5
11
15
16
A partitioning of ventilation between the right and left lung by EIT and a comparison with radionuclide scanning was performed by Kunst and coworkers.12
However, they needed to average 10 data collection cycles to obtain one EIT image. Although the time resolution was rather poor, they found a good correlation between the EIT and radionuclide methods for separate lungs. With the advanced hardware and software9
25
used in the present study, the technique also should be suitable for the regional analysis of ventilation within the lungs. We obtained an excellent linear correlation between f-EIT and ventilation scintigraphy (ie, SPECT scanning). Whether breathing was spontaneous or mechanically delivered did not affect the correlation. Moreover, we did not see any influence of the severity of lung damage in this correlation on a global level (ie, no correlation to respiratory compliance and EVLWI) and on a regional level (ie, no difference in the relationship between EIT and SPECT in dependent, more severely affected lung regions or upper, nondependent regions).
However, a regression coefficient of 0.82 and the positive intercept between SPECT scanning (independent variable) and EIT (dependent variable) showed that EIT slightly underestimated ventilation compared to SPECT scanning in rather well- ventilated regions and overestimated it in poorly ventilated areas. The difference in regional ventilation as measured by both methods was < 10%, and this is comparable to the coefficient of variation of most physiologic methods. The EIT method thus seems to be acceptable as a monitoring device in, for example, the intensive care setting. Potential causes of the measured differences between EIT and SPECT are discussed in the following paragraph.
Linearity of EIT
Holder and Khan26
determined the linear operating range of the impedance change caused by various polyacrylamide gels with different impedance characteristics. As a physiologic model, they used a saline solution-filled tank and an EIT system (APT System Mark I; IBEES; Sheffield, UK). Impedance changes increased in a linear fashion up to 20%. Impedance changes of > 20% were underestimated by the EIT system. This effect may depend on the electrical properties of the EIT system27
or on the image reconstruction algorithm,28
which is the same as the one that we have used. The algorithm for image reconstruction is based on the assumptions that the object is two-dimensional and circular. Furthermore, the distribution of resistivity initially has to be uniform, the changes in resistivity should be small, and the electrodes should be spaced equally around the thorax. Nevertheless, EIT images can be acquired even if these assumptions are violated. The respiratory patterns in the present study included VT values
15 mL/kg, which confined impedance changes to a maximum of 25%. This may be the reason for the underestimation of larger regional volume changes, as seen in the present study.
Possibilities and Limitations of EIT
Value of Information:
Although we calculated only 20 regions of interest from one f-EIT image, the spatial resolution allows the calculation of a maximum of 912 regions of interest per image. According to Hahn et al,29
the minimal detectable lung volume by EIT is in the range of 9 to 29 mL, whereas the spatial resolution using SPECT scanning is approximately 1 mL. Thus, regional ventilation of well-ventilated lung areas, which are located close to poorly or nonventilated lung regions, may be underestimated by EIT and vice versa.
By measuring EIT over time, this spatial resolution combined with the time resolution of
44 Hz offers the possibility of following regional ventilation. A lower sampling rate of 1 Hz is sufficient to monitor slow events like stepwise changes in lung volume during breath-holding procedures.30
Higher sampling rates may be suitable for rapid physiologic events, like a forced expiration in spontaneously breathing subjects.
Since EIT images are generated by relatively small local impedance changes, whereas absolute impedance values may vary considerably among different subjects, EIT is most suitable for intraindividual monitoring, with each subject serving as his own reference.
Clinical Application:
As it is noninvasive and easily transportable, EIT might be suitable for monitoring regional lung function in mechanically ventilated patients at the bedside. It may be used to acquire information, such as regional lung ventilation, regional distribution of VT, regional lung volume, and functional residual capacity.31
Furthermore, fluid accumulation, redistribution of ventilation inside the lungs,32
and regional compliance curves11
may be reconstructed from EIT data. Thus, EIT could be used, for example, to monitor atelectasis and hyperinflation during different ventilatory strategies or in opening procedures. In patients with COPD, EIT measurement of regional lung function during therapeutic interventions such as bronchodilator administration or noninvasive ventilation also might be possible. In neonates, for whom radiation exposure is undesirable, EIT potentially may replace radiographic investigations in certain situations.7
However, EIT should not be applied to obtain morphologic information similar to CT scanning or MRI, since the latter diagnostic methods provide anatomic information with a much higher spatial and morphologic resolution compared to EIT.
| Conclusion |
|---|
|
|
|---|
| Footnotes |
|---|
Received for publication July 11, 2002. Accepted for publication January 29, 2003.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Nirmalan, M. Niranjan, T. Willard, J. D. Edwards, R. A. Little, and P. M. Dark Estimation of errors in determining intrathoracic blood volume using thermal dilution in pigs with acute lung injury and haemorrhage Br. J. Anaesth., October 1, 2004; 93(4): 546 - 551. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Hedenstierna Using Electric Impedance Tomography to Assess Regional Ventilation at the Bedside Am. J. Respir. Crit. Care Med., April 1, 2004; 169(7): 777 - 778. [Full Text] [PDF] |
||||
![]() |
J. A. Victorino, J. B. Borges, V. N. Okamoto, G. F. J. Matos, M. R. Tucci, M. P. R. Caramez, H. Tanaka, F. S. Sipmann, D. C. B. Santos, C. S. V. Barbas, et al. Imbalances in Regional Lung Ventilation: A Validation Study on Electrical Impedance Tomography Am. J. Respir. Crit. Care Med., April 1, 2004; 169(7): 791 - 800. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||