| Abstract |
|---|
|
|
|---|
Setting and patients: Over a 15-month period at two tertiary-care teaching hospitals, 42 patients with PAH underwent repeated, paired CPETs using cycle ergometry.
Interventions and measurements: Each patient underwent one to six pairs of cycle ergometry tests to maximal tolerance. Each pair of tests was separated by 3 months, with each test in the pair separated by 1 to 3 days. Specific guidelines were given to the independent evaluators for the key measurements assessed from each CPET study: peak O2 uptake (
O2), peak heart rate, peak O2 pulse, anaerobic threshold (AT), and end-tidal PO2, end-tidal PCO2, and the ventilatory equivalent for CO2 at the AT (
E/
CO2@AT).
Results: There were no fatalities or complications occurring among the 242 tests performed on 42 patients. The mean peak
O2 was 722 mL/min or 41% of predicted; 34 patients were Weber class C or D. Using the specific guidelines to measure the variability of measurements made by the six independent evaluators, the coefficients of variation were < 2.2% for peak
O2, peak heart rate, peak O2 pulse, end-tidal values at the AT, and
E/
CO2@AT, while for the AT, it was 8.5%. There were no significant differences in these measurements between the first and second tests of any pair or between the earlier and later sets of pairs.
Conclusions: Using specific guidelines, key CPET measurements can be safely, reliably, and reproducibly assessed even in patients with severe exercise intolerance.
Key Words: anaerobic threshold exercise testing O2 pulse peak O2 uptake reproducibility ventilatory efficiency ventilatory equivalent for CO2
| Introduction |
|---|
|
|
|---|
O2), which defines the maximal transport of extractable O2 to the tissues; heart rate at peak
O2 (peak HR); O2 pulse at peak
O2 (peak O2P), which reflects peak stroke volume and cardiac output; anaerobic threshold (AT), which assesses maximal sustainable aerobic work capacity; ventilatory equivalent for CO2 at the AT (
E/
CO2@AT), which is a good estimate of gas exchange abnormalities and is often described as ventilatory efficiency;25 and end-tidal CO2 value near the AT (PETCO@AT) and end-tidal O2 value near the AT (PETO2@AT), which also assess gas exchange and alveolar ventilation.
Severe disability could increase the variability and decrease the confidence of interpretation of these key CPET measurements, as suggested in one series13 that showed higher measurement variability in comparing severe to less severe heart failure. The paired CPET data from a number of patients with pulmonary artery hypertension (PAH) evaluated as part of a therapeutic study at two sites were available for evaluation. Because of the overall severity of these patients illness, whether assessed by Weber classification26 or percentage of predicted peak
O2,27 we thought it important to evaluate the reproducibility of their key CPET measurements to ascertain if their single test values were as reliable as CPET values from less disabled patients in other reported series. Additionally, in this population, we recognized that right-to-left shunting of CO2-rich blood through a patent foramen ovale, which frequently occurs at the onset of exercise in patients with PAH, might interfere with the assessment of the AT.28
Because commercial breath-by-breath systems have differing time frames for selecting these key variables, we established predetermined rules for the use of the six study-blinded independent evaluators to record their measurements. Using paired CPETs conducted 1 to 3 days apart, we assessed not only the variability between these measurements in 228 tests (114 paired) from 42 patients (observer plus patient performance variability), but also the variability of evaluators in measuring key measurements in a single test (interobserver variability).
| Materials and Methods |
|---|
|
|
|---|
Protocol
At both sites, after familiarization with the CPET equipment, patients underwent an incremental CPET on a cycle ergometer under physician supervision, breathing through a mouthpiece with nose clipped. The protocols were identical at both sites: 3 min of sitting rest, 3 min of unloaded cycling, increasing work rate in a ramp fashion at a rate of 5 to 15 W/min (mostly 10 w/min) until symptom-limited, followed by 2 to 3 min of recovery.
Equipment and Data Display
At both sites, CPETs were performed using computercontrolled cycle ergometers with either a MedGraphics Cardiorespiratory Diagnosis System (Medical Graphics; St. Paul, MN) or a SensorMedics Vmax System (SensorMedics; Yorba Linda, CA). Calibration was done according to the instructions of the manufacturers. Gas exchange and cardiovascular data were retrieved electronically after exercise. Each exercise study was retrieved from its electronic file. The data were computer averaged for either 10-s or 15-s periods and formatted by one investigator into graphical and tabular displays at one of the sites so that the graphical displays were similar from the two sites. The data given to each grader consisted of nine-panel graphs on a single page,25 an enlarged CO2 output (
CO2) vs
O2 graph with equal axes on a single page for optimal discrimination of the AT by the V-slope method,29 and tabular data on one to two pages.
Data Evaluation
Each study set of three to four pages was coded, randomized, given a new identity number, and copied so that evaluators receiving the printed data were blinded as to the patients identities, dates of the study, their cohort, or any investigators prior interpretations of their studies. For each test, six evaluators, four from one site and two from the other site, agreed to independently record their numerical interpretation of the data on preprinted blank forms, following specific instructions. Each evaluator received the guidelines shown in Table 1
to determine the seven exercise measurements listed.2829 To obtain maximal stability, PETCO2@AT and
E/
CO2@AT values were averaged at and immediately after the AT was reached, while PETO2@AT values were averaged immediately before and at the AT.30 Evaluator experience in interpreting the AT ranged from 1 year to > 25 years.
|
Statistical Analysis
Single Tests:
For the 228 exercise tests evaluated, the mean, SD, and coefficient of variation (CoV) for the group for each variable were calculated. To compare the role of investigator/evaluator experience on the variability of AT measurements, paired t tests with Bonferroni corrections were used to compare the two most experienced, two intermediate experienced, and two least experienced evaluators.
Paired Tests:
Using the above mean values and SDs for each variable for the 114 pairs of CPET studies, Pearson correlation coefficients (r) and CoV were calculated. Absolute and percentage differences between the first and second tests of each pair were summarized so that their mean and median values of the variable differences could be calculated. (For example, if the peak
O2 for tests 1 and 2 in one patient were 950 mL/min and 1,050 mL/min, respectively, the mean, SD, CoV, and absolute and percentage differences would be 1,000 mL/min, 70.7 mL/min, 7.1%, 100 mL/min, and 10%, respectively.) The group mean values for each variable for the first and second of each paired test were graphed using equal X-Y coordinates according to the Bland-Altman method.31 To assess the importance of the size of the measured value, we also calculated the r, slope, and intercept values as each variable increased. Analysis of variance was performed to ascertain whether or not differences in patient age or the number of repeated paired tests caused significant trends or differences in any measurement variability; p < 0.05 was considered significant.
To assess what magnitude of differences between serial tests might be clinically significant in a single patient with severe heart or lung disease, we considered that physiologic improvement might show consistent increases in peak
O2, peak O2P, AT, PETCO2@AT, and decreases in
E/
CO2@AT and PETO2@AT greater than those found in reproducibility of the paired tests reported here. Physiologic deterioration would show the opposite changes. Therefore, we averaged the consistent change of these six separate independent physiologic variables, all weighed equally, for each pair of tests (either improving or deteriorating), and sorted the 114 paired tests from minimal to maximal change.
| Results |
|---|
|
|
|---|
O2 from 24 to 69% of predicted.25 Thirty-five of the 42 patients in the study were female (83%). All but eight were Weber class C or D (Table 2
).
|
O2, peak HR, peak O2P, and PETO2@AT were < 1%, for PETCO2@AT and
E/
CO2@AT were approximately 2%, and for AT was 8.5% (Table 3
). There were no significant differences or trends in variability with repeated testing.
|
For 114 paired CPET studies, the absolute and percentage median differences between test 1 and test 2 measurements and CoV are shown in Table 4 . The median differences and CoV both ranged from 1 to 7%.
|
O2, peak O2P, and
E/
CO2@AT (Table 5
). For peak
O2, the absolute difference between paired tests increased by 64 mL/min for each 1,000 mL/min increase in
O2, or 6.4%; for peak O2P, the absolute difference increased by < 0.04 mL/beat per 1 mL/beat increase, or 4%. For
E/
CO2@AT, the significant r value (r = 0.446) became insignificant (r = 0.022) when its reciprocal times a constant, which equals the PECO2@AT, was calculated (Table 5, Fig 2). Other variable differences were insignificantly affected by a change in quantity.
|
|
|
|
| Discussion |
|---|
|
|
|---|
O2, peak HR, peak O2P, and
E/
CO2@AT were especially consistent.
Measurements of the AT were more variable, even for more experienced evaluators, but the differences were relatively small even for the less experienced. In our prior experiences, AT differences between evaluators are reduced when a single evaluator has the opportunity to evaluate tests in the same patient serially or if a group of evaluators take the opportunity to discuss specific tests together. Neither of these tactics was used in this study. Subjectively, we previously had noted that AT measurement tended to be more difficult in patients who exhibit right-to-left shunting through an open foramen ovale during exercise, as the sudden increase in
CO2, PETO2, and ventilatory equivalents, accompanied by a decrease in PETCO2, and usually decrease in oximetry saturation can be interpreted incorrectly as the onset of the AT, because the AT is also manifested by an increase in VCO2, R, and minute ventilation/
O2.28 Thus, we were surprised to find that there was no significant increase in the variability of AT measurement in those patients that exhibited such shunting.
Paired Tests
The evaluation of paired tests in our study indicates that a single, well-performed test in a cooperative patient, even with severe or very severe exercise limitation, is valid. Despite slight differences in site equipment and technicians, time of day, patient mood, intercurrent patient illnesses, or fluctuations in disease intensity, the measured values of a second test performed 1 to 3 days later usually differed only minimally from the first test. In fact, the median differences for paired tests for all variables ranged from only 1 to 7% for peak
O2, peak HR, peak O2P, AT, and
E/
CO2@AT (Table 4). The AT reproducibility may be especially important since some clinicians and investigators are reticent to have their patients exercise to their maximal tolerance for measurement of peak
O2, peak HR, and peak O2P, and thus may choose to use the AT as a surrogate for maximal aerobic capacity.
Measurement variability was not increased at either age extreme. Also, the variability between paired tests did not significantly change from the first pair to the second, third, fourth, fifth, or sixth pair of tests. Quantitatively, the modest increase in absolute differences between paired tests for peak
O2 and peak O2 pulse at higher values (Fig 2, Table 5) is not surprising. In fact, the lower absolute differences in patients with more severe disorders are salutary.
The variability of
E/
CO2@AT-paired differences significantly increased as their values increased from 30 to 90 U. In contrast, the variability of PECO2@AT-paired differences remained stable from 30 to 10 mm Hg (Fig 2, Table 5). The PECO2 and
E/
CO2 values are reciprocal. Thus, as the PECO2@AT values declined and ventilatory efficiency decreased, the variability between paired tests for
E/
CO2@AT significantly increased.
Factors Not Assessed
Our study did not assess the reproducibility of peak
CO2, peak respiratory exchange ratio, or peak ventilation because these values are somewhat dependent on recent diet and substrate availability.11 We did not assess peak work rate, as it is more variable and also of lesser clinical utility than peak
o2.26 Nevertheless, other investigators examining highly trained cyclists35 and a large number of normal adults3 found test-retest correlation coefficients > 0.86 for peak
o2, peak
CO2, peak ventilation, peak HR, and peak work rate. We did not assess
E/
O2 or
E/
CO2 at the peak work rate, because these values are dependent on the severity of acidemia during exercise and thus do not reflect maximal gas exchange efficiency.27 It would have been useful for us to have also assessed the nadir
E/
CO2 after the AT or slope of minute ventilation vs
CO2, as these have variability similar to that of
E/
CO2@AT in a normal population.30
Clinical Relevance
The high reproducibility of the key physiologic measurements in our study compares favorably with other studies involving patients with heart, lung, musculoskeletal, or renal disease with less severe exercise impairment. For example, in 11 patients with end-stage renal disease and moderate exercise impairment (mean peak
O2 of 1,500 mL/min), Koufaki et al14 reported CoVs for peak
O2, peak HR, and peak O2P of 4.7%, 5.9%, and 6.0%, respectively. In an exacting study of 17 patients with mild heart disease, who were hospitalized, fasted, and tested at the same time each day, Lehmann and Kolling15 reported correlation coefficients of 0.996 for peak
O2, 0.928 for peak HR, 0.991 for peak O2P, and 0.995 for
E/
CO2@AT. Their very high correlation coefficients can be partially attributed to the broad range of peak
O2 in their patients (803 to 3,927 mL/min). In 16 cardiac patients with moderate impairment (Weber class A for 6 patients, Weber class B for 5 patients, and Weber class C for 5 patients, mean peak
O2 of 26.5 mL/min/kg, 17.4 mL/min/kg, and 13.8 mL/min/kg, respectively) tested several times over a period of 3 to 22 months, Janicki et al13 found CoVs of 5.7%, 4.4%, and 9.2% for peak
O2, peak HR, and AT, respectively. In duplicate tests in 11 cardiac patients (4 patients were New York Heart Association class II, and 7 patients were class III), Meyer et al,19 in patients with mean peak
O2 of 13.9 mL/min/kg, found CoVs for peak
O2, peak HR, and peak O2P of 4.1%, 1.4%, and 4.4%, respectively.
Several studies of CPET reproducibility in patients with obstructive or restrictive lung disease have also been reported. In triplicate tests in nine adult patients with cystic fibrosis (mean predicted peak
O2 of 62%), McKone et al17 found a CoVs for peak
O2 and peak HR were 6.9% and 3.0%, respectively. In triplicate tests in six patients with restrictive lung disease and predicted peak
O2 of 58 ± 10%, Marciniuk et al16 found a CoVs of 5.3% for peak
O2 and 4.0% for peak HR. In 11 patients with obstructive lung disease, familiar with CPET and mean peak
O2 of 1.3 L/min, Brown et al6 found a CoV of 7%. In 11 patients with mild obstructive lung disease (peak
O2 of 1.95 L/min), Cox et al9 reported CoVs for peak
O2 and peak HR of 3% and 4%, respectively. In duplicate tests in 56 patients with COPD with a mean predicted peak
O2 of 65%, Covey et al8 found correlation coefficients and CoVs of 0.97 and 4.5%, and 0.92 and 3.0%, respectively, for peak
O2 and peak HR.
The patients with PAH in the current study appear to be more impaired than those in any previously reported CPET reproducibility study of patients with either cardiovascular or pulmonary diseases, as their mean ± SD percentage of predicted peak
O2 was only 43 ± 10% and their Weber classes were usually severe or very severe (81% C or D). Yet, the reproducibility of their key CPET measurements remained high. Most of the studies cited above, as did ours, used cycle ergometry, which has the advantage of being under patient control and also of allowing the calculation of the
O2/work rate relationship. However, for the measurements that were addressed in this study, there is no reason to expect that they would not be similarly reproducible using treadmill ergometry.
Thus we conclude that single, well-conducted gas exchange exercise tests can safely and reliably measure key and discriminating exercise variables, including peak
O2, in moderately impaired to very severely impaired children and adults. The median difference in peak
O2 was 7%, with lesser differences in the other measurements, all relatively independent of the absolute value. Thus, taken alone, a change in peak
O2 of 8 to 10% might be used as a cut off in deciding that there is a significant change in exercise tolerance. However, using the combination of peak
O2, peak O2P, AT, PETCO2@AT,
E/
CO2@AT, and PETO2@AT from one test to the next the average consistent change would be < 3% in half of all patients, < 5% in 7 in 10 of all patients, and < 10% in 9 in 10 of patients. To increase uniformity of reading and interpreting aerobic function and ventilatory efficiency using CPET, we recommend that other practitioners consider using the guidelines given in Table 1.
| Acknowledgements |
|---|
| Footnotes |
|---|
CO2 = CO2 output;
E/
CO2@AT = ventilatory equivalent for CO2 at and immediately after the anaerobic threshold;
O2 = O2 uptake Received for publication October 7, 2003. Accepted for publication May 6, 2004.
| References |
|---|
|
|
|---|
O2 max in prepubertal girls. Med Sci Sports Exerc 32,865-869[ISI][Medline]
O2 max. Int J Sports Med 1990;11,26-32[ISI][Medline]
O2 max in patients with chronic air-flow obstruction. Am Rev Respir Dis 1985;131,435-438[ISI][Medline]
O2 max during progressive and constant bicycle exercise in patients with chronic obstructive lung disease. Respiration 1984;45,197-206[ISI][Medline]