| Abstract |
|---|
|
|
|---|
Methods and results: Reversible acute mitral regurgitation (AMR) was induced in eight calves by stenting the mitral valve using a vena cava filter. Full and partial AMR assist were compared with maximum IABP support in each animal. In full-support mode, both assist systems increased cardiac output (miniature intracardiac pump, 13% [p < 0.05]; IABP, 3% [p < 0.05]), mean aortic pressure (miniature intracardiac pump, 13% [p < 0.05]; IABP, 8% [p < 0.05]), carotid artery flow (miniature intracardiac pump, 29% [p < 0.05]; IABP, 5% [difference not significant]), and coronary blood flow (miniature intracardiac pump, 25% [difference not significant]; IABP, 34% [p < 0.05]). Again in full-support mode, both systems reduced left atrial pressure (miniature intracardiac pump, 2.4 mm Hg [p < 0.05]; IABP, 0.7 mm Hg [p < 0.05]), peak left ventricular (LV) pressure (miniature intracardiac pump, 13% [p < 0.05]; IABP, 5% [p < 0.05]), and external LV work (miniature intracardiac pump, 29% [p < 0.05]; IABP, 3% [p < 0.05]). Only full miniature intracardiac pump support reduced both end-diastolic LV volume (7%; p < 0.05) and end-systolic LV volume (10%; p < 0.05). IABP mainly improved coronary perfusion, while the miniature intracardiac pump proved more capable of genuinely unloading the LV.
Conclusions: We conclude that during severe acute LV failure, the miniature intracardiac pump is capable of more effective cardiac unloading and circulatory support than IABP.
Key Words: assist device hemodynamics intracorporeal circulation shock ventricular unloading
| Introduction |
|---|
|
|
|---|
The Impella intracardiac microaxial pump (IMP) [Impella Cardiotechnik AG; Aachen, Germany] is a new, catheter-based cardiac assist device that is designed to provide ventricular support in bypass mode. Although the IMP has been introduced clinically recently, the hemodynamic efficacy has not yet been extensively evaluated.1011 In the present study, left ventricular (LV) support using the IMP and an IABP (System 98 XT; Datascope; Fairfield, NJ) were evaluated in an animal model of acute mitral regurgitation (AMR).5
| Materials and Methods |
|---|
|
|
|---|
Cardiac Support Systems
The IMP features an integrated cannula-pump assembly that is 6.4 mm in diameter and 100 mm in length, which is suitable for placement across the aortic valve. The axial impeller of the pump (ie, an Archimede screw) continuously recruits blood from the LV via the cannula and drains the blood into the aorta just above the aortic valve. The pump housing contains an integrated solid-state differential pressure sensor that registers the pressure gradient across the pump. The device is powered and controlled by a pump console via a connecting catheter. The placement of the catheter pump occurs via percutaneous access or intraoperatively via aortic cannulation. The pump is designed for short-term to mid-term support (maximum duration, 7 days).1213
The IMP was introduced via the left carotid artery and was placed across the aortic valve with the inflow cannula in the LV, and with the pump housing and outlet just above the aortic valve. Positioning was facilitated by the pressure gradient reading on the pump console.1113 Routine checks with fluoroscopy confirmed the proper placement of the pump. Immediately after placement of the IMP, the pump rate was set to a minimum level that was sufficient to prevent aortic retrograde flow. The IMP console flow measurement was calibrated by the automated console calibration sequence after pump positioning.
A 40-mL intraaortic balloon catheter (Datascope) was introduced via the right femoral artery and was placed in the high descending aorta. The IABP was triggered on the ECG.
Instrumentation
A Swan-Ganz thermodilution catheter (Arrow International; Reading, PA) was advanced into the pulmonary artery via a jugular vein. A conductance catheter, incorporating a pressure sensor (CD Leycom; Zoetermeer, the Netherlands), was positioned in the LV via the left femoral artery. The conductance catheter was connected to a conductance console (Sigma 5/DF; CD Leycom), which was used in the dual-frequency mode.14 Solid-state pressure catheters (CD Leycom) were placed in the abdominal aorta via the left femoral artery and in the left atrium through a purse-string suture. BP in the aortic arch was measured via the lumen of the intraaortic balloon catheter with a standard pressure transducer (Baxter International Inc; Deerfield, IL). Ultrasonic flow probes (Transonic Systems; Ithaca, NY) were placed on the pulmonary artery, the right carotid artery, and on a side branch of the left anterior descending coronary artery. All hemodynamic data were acquired and stored using a 16-channel acquisition system (CD Leycom).
To induce AMR, an incision was made in the left atrium through which a retrievable steel wire cage (Günther Tulip Vena Cava Filter; Cook Europe; Bjaeverskov, Denmark) was placed in the mitral valve.59 The degree of AMR was assessed and controlled by the monitoring of real-time LV pressure and volume, and left atrial pressure (LAP). The induction of mitral regurgitation was consistently aimed at maximizing the regurgitant orifice.
Conductance Calibration
Parallel conductance was determined by injecting 7.5 mL 6.5% hypertonic saline solution into the pulmonary artery.15 A 5-mL blood sample was collected in a sampling cuvette for blood resistivity measurement (CD Leycom).16 The gain calibration factor was determined by comparing LV conductance-derived output (CDO) with pulmonary artery flow (QPA) in the healthy animal. The conductance data were analyzed with an appropriate software package (Circlab 2000; GTX Medical Software; Leiden, the Netherlands).
Measurements and Analysis
Both the IABP and the IMP were tested in each animal, in an alternated sequence in consecutive experiments. The IABP was driven in 1:1 mode and was tuned to maximize LV afterload reduction and diastolic aortic pressure augmentation.123 The IMP was run at medium speed (24,000 revolutions per minute; performance level, P5) and full speed (33,000 revolutions per minute; performance level, P9). Approximately 10 s after the creation of mitral regurgitation, assist was initiated and maintained for 30 s.
The hemodynamic data were analyzed and processed to obtain objective measures for left heart pump function, myocardial working conditions, and peripheral perfusion. At baseline (ie, during calibration) the CDO (ie, stroke volume x heart rate) was set equal to the QPA, and the flow through the aortic valve (QAV) was considered to be equal to the QPA. During mitral regurgitation, the CDO expresses both the forward output (ie, the QAV) and the regurgitant flow across the (stented) mitral valve (QR), and thus CDO = QAV + QR (in liters per minute). During IMP support, part of the QAV no longer occurs through the aortic valve but through the IMP, and QAV is calculated as QPA the flow of the IMP (QIMP). QAV is solely due to the contraction of the LV. QIMP is the flow measured through the IMP. The QPA is then the result of the combined effects of LV contraction and IMP aspiration.
Pulse pressure (PP) was calculated as the differential between the maximum and minimum values of ascending aortic pressure (PASC). The diastolic coronary flow fraction (DCF) was calculated as the ratio of antegrade diastolic coronary flow (QCOR) to the mean QCOR, multiplied by 100%. External LV work (LVW) was calculated as follows: (peak systolic LV pressure [PLVpeak] LAP) x (QAV + QR) [in Joules per second]. The QPA/LAP ratio was calculated as an index of left heart pump function.17 A measure for the balance in coronary demand and supply was defined by the LVW/QCOR ratio (in Joules per milliliter). For each setting, the on condition (ie, AMR plus assist) and the off condition (ie, AMR plus assist idle) were compared by the application of the nonparametric Wilcoxon signed rank test (eight subjects). Significance was assumed if the p values were < 0.05.
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Cardiac Unloading
The reduction of LVW and oxygen consumption is an essential element of cardiac unloading and cardioprotective therapy. Afterload is a major determinant of myocardial work.118 PLVpeak as a measure of LV afterload was reduced with use of the assist mode (Table 1). 19 In general, a suitable definition for LV pressure load would be PASC LAP,20 but this factor is prone to error in the case of LV and aortic pressure dissociation (Fig 5). Therefore, external LV pressure work was estimated as PLVpeak LAP.
The IMP, in P9 mode, reduced both regurgitant valve flow (15%) and aortic valve flow (52%), as well as external LVW (29%) and LV volume (EDV, 7%; ESV, 10%). These data suggest that during full IMP support total LVW and myocardial oxygen consumption are reduced.182122 The IABP showed less effective unloading capabilities.
Myocardial Support
In general, myocardial function is supported by reducing myocardial work and augmenting coronary perfusion. The IMP improved coronary perfusion (QCOR increase, 25%) and strongly reduced cardiac workload (LVW decrease, 29%; PLVpeak decrease, 13%).18 The IABP primarily improved coronary perfusion (QCOR increase, 34%) as work reduction was only marginal (LVW decrease, 3%; PLVpeak decrease, 5%). The coronary demand-supply balance was significantly improved (ie, LVW/QCOR ratio decreased) by both devices (Table 1), but the overall effectiveness of the IMP was greater than that of the IABP (Fig 3). The LVW/QCOR ratio reflects both the ventricular unloading and coronary perfusion aspects. In clinical practice, the priority may lie with either one of these. If the priority lies with myocardial oxygen supply (ie, control of ischemia), the IABP remains an effective, well-proven tool for therapy.23 When additional LV unloading and immediate circulatory support are mandatory, the IMP may present a solid alternative.
Impact of Preassist Condition
An analysis of the effect of support among individual subjects shows that functional improvements with assist are greater in weaker subjects (Fig 3, 4). In the P9 setting, the average IMP flow was 4.2 L/min (Fig 1). With preassist cardiac outputs of > 8 L/min, a bypass flow of 4 L/min naturally has only limited impact. In individuals with insufficient cardiac output (ie, < 4 L/min), pump function improvement was more pronounced (IMP, > 100%; IABP,
25%) [Fig 3]. These trends and the data in Table 1 suggest that the IMP has greater effectiveness in improving overall pump function than the IABP. Figure 4 shows that the IABP partially compensates for its poor volume unloading performance by improving the coronary perfusion aspect.159
LV Output and IMP Output
Table 1 gives the values for QPA, QR, and QAV. Figure 2 provides a detailed overview of the LV and device-related contributions to forward cardiac output. Apparently, CDO is the sum of QR, QAV, and QIMP. The slight discrepancy found when calculating CDO from Table 1 is due to rounding errors (ie, EDV ESV x HR
13.7 L/min, while QR + QAV + QIMP
14 L/min). Output analysis shows that part of the stroke volume ejected by the LV is via the IMP (ie, CDO contains QIMP) [Fig 2]. The QIMP is due to the simultaneous effect of LV contraction and IMP aspiration. Essentially, the IMP acts as an afterload reduction device that augments ventricular output toward the aorta, which effectively redistributes forward output and regurgitant flow. (The cumulative bar is lifted with respect to the zero output level in Fig 2.)
LVW was calculated from (PLVpeak LAP) x (QAV + QR), which implies that the fraction of LV stroke output through the IMP does not contribute to external LVW. Realistically, this assumption may introduce some error. While both the LV and IMP work simultaneously, the exact share of pump and cardiac work cannot be identified.
Pulsatility
Whereas the IMP reduces the pulsatility in the aorta, the IABP increases PP (Fig 5 and Table 1). The evaluation of the quality of pulseless perfusion based solely on hemodynamic measurements is controversial and therefore was not addressed further.24 Moreover, the present study is focused on acute hemodynamic collapse, which renders the long-term pulseless perfusion effects less relevant. Interestingly, however, the IMP did not flatten the QCOR curve but exhibited a behavior comparable to that of the IABP during systole (Fig 5). The decreased systolic flow seen with use of the IABP can be attributed to the intentional aortic pressure reduction at end-diastole. The decrease in systolic flow with the IMP may be caused by a pressure decline due to an accelerated jet (ie, the Venturi effect) from the outflow of the IMP.25
Model
Other investigators20 who tested a device similar to the IMP reported less promising results. In their bovine ischemic LV dysfunction model (coronary ligation), mean aortic pressures of 88 mm Hg existed in the presence of a cardiac index of 2.1 L/min/m2 (values for this study: aortic pressure, < 50 mm Hg; cardiac index, 3.2 L/min/m2). As the presence of high afterload limits the efficacy of axial flow pumps, insufficient hemodynamic collapse will not challenge these devices. As the IMP was designed to provide support up to 4.5 L/min, a substantial level of hemodynamic compromise is considered necessary to assess the potential of the device.9 The induction of AMR did not lead to substantially elevated LAPs.5 The pre-AMR values of QPA and PASC in Table 1 could be indicative of a distended peripheral arteriolar vascular bed at baseline. Vasodilation is known to reduce the effects of mitral regurgitation.26 Ultimately, the level of hemodynamic compromise obtained in the present study proved to be effective in characterizing the IMP and the IABP.
Clinical Implications
The present results suggest that the IMP can effectively reduce the cardiac workload while increasing cardiac output and that it outperforms the IABP in low-cardiac-output conditions. This implies that the IMP may be a better alternative in circulatory collapse patients.427 Moreover, in the critical situation of sudden left heart deterioration, the IMP may provide a safety net, buying time for the medical team to act.
Although the present study has illustrated the potential of the IMP to take over the left heart pump function (Fig 5), it must be realized that the pumping capacity heavily depends on the availability of preload. Right-sided circulatory failure seriously limits the effectiveness of LV assist devices and is a contraindication for their use.2829
In the treatment of unstable patients with angina, both the IMP and the IABP would be beneficial in terms of sustained afterload reduction and coronary perfusion enhancement.23 As timing and synchronization are critical when using the IABP, the continuous-flow IMP might perform more consistently than the IABP in rhythmically unstable patients (eg, atrial fibrillation).
The IMP is a device with a diameter of 6.4 mm (19F) that requires surgical implantation under direct vision. A smaller version of the IMP would allow easier percutaneous introduction and might stretch the indications for intracorporeal support beyond those for the IABP. Moreover, both systems could prove synergetic when applied simultaneously.
| Conclusion |
|---|
|
|
|---|
| Footnotes |
|---|
Received for publication July 29, 2003. Accepted for publication April 29, 2004.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
K. D. Reesink, L. D.C. Sauren, A. L.A.J. Dekker, E. Severdija, T. van der Nagel, G. G. Geskes, F. H. van der Veen, and J. G. Maessen Synchronously counterpulsating extracorporeal life support enhances myocardial working conditions regardless of systemic perfusion pressure Eur. J. Cardiothorac. Surg., December 1, 2005; 28(6): 790 - 796. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||