Chest ACCP Career Connection
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Guest Access | Sign In via User Name/Password
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (8)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Gelb, A. F.
Right arrow Articles by Zamel, N.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Gelb, A. F.
Right arrow Articles by Zamel, N.
(Chest. 2004;126:1855-1860.)
© 2004 American College of Chest Physicians

Simplified Detection of Dynamic Hyperinflation*

Arthur F. Gelb, MD, FCCP; Carlos A. Gutierrez, MD, MSc; Idelle M. Weisman, MD; Randy Newsom, CPFT, RCP; Colleen Flynn Taylor, MA and Noe Zamel, MD, FCCP

* From the Pulmonary Division (Dr. Gelb), Department of Medicine Lakewood Regional Medical Center, Lakewood, CA; Faculty of Medicine (Drs. Gutierrez and Zamel), University of Toronto, Toronto, ON, Canada; Department of Clinical Investigation and Pulmonary Critical Care Medicine (Dr. Weisman), William Beaumont Army Medical Center, El Paso, TX; and independent contractors (Mr. Newsom and Ms. Taylor), Lakewood, CA.

Correspondence to: Arthur F. Gelb, MD, FCCP, 3650 E South St, Suite 308, Lakewood, CA 90712; e-mail: afgelb{at}msn.com


    Abstract
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 References
 
Study objective: To detect dynamic hyperinflation by comparing reduction in inspiratory capacity (IC) during both paced hyperventilation and cycle ergometry in patients with moderate-to-severe COPD, studied before and after acute bronchodilation.

Methods: IC and FEV1 were measured before and after metronome-paced hyperventilation at twice the resting respiratory rate for 20 s in 16 patients with COPD before and after 54 µg aerosolized ipratropium bromide (IB). We also studied the same 16 patients before and after administration of 54 µg aerosolized IB during symptom-limited incremental cycle ergometry when the final respiratory rate was also twice the resting rate.

Results: Resting IC was 2.23 ± 0.53 L (mean ± SD), and the mean decrease in IC from baseline was 0.36 ± 0.25 L after exercise (p < 0.001), and not significantly different (p = 0.64) from mean decrease in IC of 0.40 ± 0.29 L following hyperventilation. Results following hyperventilation and exercise were similar after bronchodilator. The mean difference for decrease of IC between hyperventilation and exercise was 0.138 L (95% confidence interval, – 0.347 to 0.622; r = 0.66, p = 0.006). The decrease in FEV1 was 0.01 ± 0.13 L after exercise and 0.06 ± 0.18 L after hyperventilation. Separately, baseline and peak end-expiratory and end-inspiratory lung volumes were similar with hyperventilation vs exercise both before and after bronchodilator.

Conclusion: Both metronome-paced hyperventilation and incremental cycle ergometry, when resting respiratory rate was doubled, provoked similar significant decrease in IC, even after administration of 54 µg aerosolized IB. The noninvasive simplicity of hyperventilation for 20 s provides a clinically useful screening surrogate to monitor changes in IC following exercise.

Key Words: COPD • dynamic hyperinflation • exercise • hyperventilation • lung function


    Introduction
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 References
 
Measurements of FEV1 have been widely accepted to classify the severity of COPD. Since FEV1 may be a poor predictor of clinical symptoms, exercise tolerance, and response to bronchodilators, additional signals have been sought. Alternatively, exercise testing with repeated measurements of inspiratory capacity (IC) has been used to detect dynamic hyperinflation and evaluate the response to bronchodilators.

O’Donnell et al1 have provided confirmatory evidence that "Borg dyspneic ratings and measurements of inspiratory capacity (IC) and endurance time during submaximal cycle exercise testing are highly reproducible and responsive to changes in severe COPD." Previous studies have demonstrated the utility of repeated measurements of IC during exercise to reflect changes in end-expiratory lung volume (EELV),234 since total lung capacity (TLC) remains constant after acute bronchodilation and during exercise.56 Progressive reduction in IC during exercise reflects dynamic hyperinflation, and is a good predictor of decreased exercise ability as well as increased exertional dyspnea.4 Additionally, peak values of inspiratory esophageal pressures used as a surrogate to estimate effort are relatively constant during multiple measurements of exercise IC.78

Changes in IC reflect dynamic hyperinflation and are correlated with breathing frequency in patients with COPD.4 We suspected increasing breathing frequency would produce changes in IC similar to the change in IC observed during exercise.

The present study was designed to compare metronome-paced hyperventilation, a relatively simple procedure, with incremental symptom-limited cycle ergometry to provoke respiratory rate-induced reduction of IC in patients with COPD. A similar comparison has not been previously reported. We also wanted to evaluate the role of inhaled ipratropium bromide (IB) to blunt the decrease in IC. O’Donnell et al1 have previously shown greater bronchodilation with IB when compared to inhaled albuterol sulfate in patients with COPD. In a preliminary study,9 we previously reported that hyperventilation in younger normal subjects resulted in insignificant changes in IC.


    Materials and Methods
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 References
 
We selected 16 patients with smoking history > 20 pack-years with documented moderate-to-severe COPD,10 who were in clinically stable condition for at least 6 weeks prior to the present study and were not receiving oxygen. A history of wheezing and/or responsiveness to aerosolized albuterol were not specific inclusion criteria. Patients were instructed to continue all their usual medications, but to withhold short-acting ß2-agonists and/or aerosolized IB for 6 h and long-acting ß2-agonists for 24 h prior to testing.

After obtaining informed consent, patients underwent lung function studies before and after administration of 270 µg aerosolized albuterol sulfate using techniques and predictive values previously described in detail.11 We used a pressure-compensated flow plethysmograph (Model 6200; SensorMedics; Viasys; Yorba Linda, CA).

Subsequently, on a separate day, metronome-paced hyperventilation was initiated in 16 patients with COPD to achieve a respiratory rate twice the baseline rate for 20 s, which was immediately followed by sequential measurement of IC and expiratory spirometry. While no attempt was made to control end-tidal carbon dioxide, patients were coached to maintain a respiratory rate synchronous with the metronome. Near-constant dynamic tidal volume during metronome-paced hyperventilation was achieved by having patients observe a graphic display of their breathing pattern. Patients were studied before and after administration of 54 µg aerosolized IB from a metered-dose inhaler. The technique for measuring IC has been previously described.14

On a separate day, symptom-limited incremental exercise was initiated in all of the 16 patients with COPD after baseline values were achieved, using an electronically braked cycle ergometer (Ergometrics 800; SensorMedics) before and 15 min after administration of 54 µg aerosolized IB. Following unloaded 2-min warm-up, a 10-W workload was increased at 1-min intervals during near-constant pedaling at 50 revolutions per minute. Immediately following symptom-limited peak exercise at constant workload for 1 min, measurements of IC and expiratory spirometry were obtained. During exercise, expired gases were collected and analyzed (max 29; SensorMedics). The protocol is described in Table 1 . We also studied 21 younger and 8 older nonsmoking healthy volunteers to evaluate the changes in IC during metronome-paced hyperventilation at twice the resting respiratory rate for 20 s.


View this table:
[in this window]
[in a new window]

 
Table 1.. Study Design for 16 Patients With Moderate-to-Severe COPD

 
Statistical Analysis
Repeated-measures analysis of variance was used. For the 16 patients, models with time (baseline and hyperventilation at twice the resting respiratory rate) and before and after bronchodilation were calculated for IC and FEV1. In the same 16 patients who performed both exercise and metronome-paced hyperventilation, repeated mixed-model analysis of variance was calculated for EELV, end-inspiratory lung volume (EILV), and FEV1 both before and after bronchodilation. The Bland-Altman method was used to measure the agreement between hyperventilation and exercise-induced changes in IC. Results are presented as mean ± SD. Analysis was done using statistical software (version 8.02 for Windows; SAS Institute; Cary, NC). Statistical significance was p < 0.05.


    Results
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 References
 
In the 21 younger healthy volunteers (13 women) aged 37 ± 13 years, baseline IC was 2.98 ± 0.73 L.9 After 20 s of metronome-paced hyperventilation at twice the resting respiratory rate, IC was 2.90 ± 0.70 L. The mean decrease in IC was 2%. In the eight older healthy volunteers (five women) aged 60 ± 2 years, baseline IC was 2.58 ± 0.68 L. Following 20 s of metronome-paced hyperventilation at twice the resting respiratory rate, IC was 2.56 ± 0.71 L. The mean decrease in IC was 0.8%. In both groups, intraclass correlation for the two measurements was 0.94 (95% confidence interval [CI], 0.77 to 0.94).

We studied 16 patients with COPD (10 men) aged 69 ± 6 years with a smoking history of 50 ± 15 pack-years. Results of routine lung function studies described in Table 2 are consistent with moderate-to-severe COPD. Following administration of 270 µg aerosolized albuterol sulfate, FEV1 increased 13 ± 12%.


View this table:
[in this window]
[in a new window]

 
Table 2.. Results of Lung Function Studies in 16 Patients With Moderate-to-Severe COPD*

 
Hyperventilation and Dynamic Hyperinflation
IC After Metronome-Paced Hyperventilation: In 16 patients, the metronome-paced respiratory rate at twice the resting rate for 20 s resulted in a significant decrease in IC from 2.23 ± 0.53 to 1.83 ± 0.53 L (p < 0.001). The resting respiratory rate was 16 ± 3 breaths/min, and the metronome-paced respiratory rate was 30 ± 5 breaths/min. The decrease in IC was 0.40 ± 0.29 L (Table 3 ).


View this table:
[in this window]
[in a new window]

 
Table 3.. Decrease in IC Following Hyperventilation and Exercise, Before and After Administration of 54 µg IB*

 
Following administration of 54 µg aerosolized IB in the 16 patients, there was a significant increase in baseline IC of 0.15 ± 0.35 L (p = 0.03). FEV1 increased from 1.42 ± 0.41 to 1.63 ± 0.53 L in the 16 patients with COPD who underwent both hyperventilation and exercise. However, with metronome-paced increased respiratory rate, the decrease in IC was similar to prebronchodilator decrease (p = 0.80) [Table 3; Fig 1 ].



View larger version (13K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 1.. Results of IC in 16 patients with COPD at rest and after metronome-paced hyperventilation (HV) and after exercise (EXE), when the resting respiratory rate was doubled, both before (left) and after (right) administration of 54 µg inhaled IB. The significant decrease in IC was similar after hyperventilation and exercise, and was not blunted by bronchodilator. Ins = inspiratory; BPM = breaths per minute; BD = bronchodilator.

 
Hyperventilation vs Exercise and Dynamic Hyperinflation
IC After Exercise vs Hyperventilation: In 16 patients, prebronchodilator symptom-limited cycle ergometry was achieved at 45 ± 6.0 W, with a respiratory rate of 29 ± 4 breaths/min, minute ventilation of 35 ± 4.5 L/min, and oxygen uptake of 1.1 ± 0.3 L/min. Following exercise, the decrease in IC was 0.36 ± 0.25 L from baseline, with an increase in respiratory rate 1.9 ± 0.3 times the resting respiratory rate. In these 16 patients, metronome-paced hyperventilation at twice the resting respiratory rate caused IC to decrease 0.40 ± 0.29 L similar to exercise (p = not significant [NS]) [Table 3, Fig 1].

After administration of 54 µg aerosolized IB, the decrease in IC was 0.32 ± 0.23 L following exercise to 43 ± 5 W, minute ventilation was 33 ± 4L/min, respiratory rate was 30 ± 4 breaths/min, and oxygen uptake was 0.9 ± 0.3 L/min. This was similar to the decrease in IC of 0.39 ± 0.29 L following metronome-paced hyperventilation at twice the resting respiratory rate for 20 s (p = NS) [Table 3, Fig 1].

Correlation Between Hyperventilation and Exercise: Before bronchodilator, the mean difference between the peak change in IC comparing hyperventilation and exercise was 0.04 L (95% CI, – 0.419 to 0.619 L; r = 0.41, p = 0.115). After bronchodilator, the mean difference between the peak change in IC comparing hyperventilation and exercise was 0.07 L (95% CI, – 0.347 to 0.622 L; r = 0.66, p = 0.006).

Bland and Altman Method for Assessing Agreement Between Hyperventilation vs Exercise
Results demonstrate bias for hyperventilation induced greater decrease in IC both before (100 mL) and after (130 mL) nebulized IB vs exercise (Fig 2 ). Furthermore, with the exception of one outlier, all values were within ± 2 SD and the majority of observations were within 1 SD of mean difference.



View larger version (17K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 2.. Bland and Altman method plotting the net difference in decrease in IC vs mean decrease in IC following hyperventilation and cycle ergometry both before (top) and after (bottom) aerosolized IB. There was a bias for greater decrease in IC following hyperventilation, and the majority of observations fell within 1 SD. BD = bronchodilator; see Figure 1 legend for expansion of abbreviation.

 
Effect of Hyperventilation and Exercise on FEV1
Prebronchodilator resting FEV1 was 1.47 ± 0.49 L, and with hyperventilation there was an insignificant (p = NS) decrease in FEV1 of 0.06 ± 0.02 L. Following cycle ergometry, the decrease in FEV1 was 0.01 ± 0.10 L (p = NS).

EELV and EILV
Prebronchodilator baseline EELV equivalent to functional residual capacity was 70 ± 7% of TLC in 16 patients, who underwent both hyperventilation and cycle ergometry testing (Table 4 ; Fig 3 ). EILV was 82 ± 7% of TLC. Following metronome-paced hyperventilation, EELV was 76 ± 6% of TLC and EILV was 88 ± 5% of TLC (p < 0.001). Following the administration of 54 µg IB, baseline EELV was 69 ± 6% TLC and EILV was 82 ± 6% TLC; after hyperventilation, EELV was 76 ± 6% TLC and EILV was 88 ± 6% TLC. After bronchodilator administration, these changes are similar to prebronchodilator changes (p = 0.37). Thus, acute bronchodilation failed to blunt the increase in EELV and EILV percentage of TLC following hyperventilation.


View this table:
[in this window]
[in a new window]

 
Table 4.. Increase in EELV and EILV Following Hyperventilation and Exercise in 16 Patients With COPD*

 


View larger version (15K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 3.. Increases in EELV and EILV in 16 patients with COPD from the resting respiratory rate to after exercise and after hyperventilation, both before (left) and after (right) administration of 54 µg aerosolized IB were similar. Obs = observed. See Figure 1, 2 legends for expansion of abbreviations.

 
Prebronchodilator and preexercise baseline EELV and EILV were 71 ± 6% of TLC and 82 ± 7% of TLC, respectively. Following exercise, EELV was 74 ± 6% of TLC and EILV was 92 ± 5% of TLC (p = 0.03). Following bronchodilation with 54 µg of IB, baseline EELV and EILV were 69 ± 6% of TLC and 81 ± 7% of TLC, respectively; after exercise, EELV and EILV were 73 ± 6% of TLC and 89 ± 6% of TLC, respectively. The increase in EELV from baseline was greater for hyperventilation compared to exercise (p = 0.01), and was not blunted by bronchodilation with 54 µg aerosolized IB (for EELV, p = 0.4) [Table 4; Fig 3]. Increases in EILV following hyperventilation vs exercise were similar (p = 0.6), and were significantly greater than baseline (p < 0.001) and not blunted by bronchodilator (p = 0.5).


    Discussion
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 References
 
The present study in 16 patients with COPD with moderate-to-severe expiratory airflow limitation demonstrated the provocative ability of metronome-paced hyperventilation at twice the resting respiratory rate for 20 s to induce a significant decrease in IC. This was similar as a group to the decrease in IC following symptom-limited incremental cycle ergometry when the baseline respiratory rate was also doubled. The noninvasive simplicity of hyperventilation for 20 s provided a clinically useful screening surrogate to monitor changes in IC following exercise. Furthermore, the significant decrease in IC following hyperventilation and exercise was not blunted by acute bronchodilation with 54 µg aerosolized IB.

The correlation between the decrease in IC following cycle ergometry vs hyperventilation was strengthened after aerosolized IB with reduction in vagal tone. However, a high correlation is not a good test of agreement since it reflects only the relationship between two variables. Agreement is better tested using the Bland and Altman method (Fig 2). Results suggest hyperventilation is more sensitive than cycle ergometry to detect decrease in IC both before and after aerosolized IB.

Additionally, increases in EELV and EILV were similar with hyperventilation and exercise, and not blunted with acute bronchodilation with inhaled IB. IB is presumably but not clearly proven a greater bronchodilator than inhaled albuterol sulfate in patients with COPD.1 This may also explain lack of improvement in dyspnea following use of albuterol and/or IB.

In normal subjects during exercise, minute ventilation increases primarily as a result of increased tidal volume rather than increased respiratory rate.12 In contrast, when patients with COPD with at least moderately severe expiratory airflow limitation and increased EELV undergo symptom limited exercise, we and others4713 have noted the increase in respiratory rate to approximately double baseline was primarily responsible for increased minute ventilation. Not surprisingly, O’Donnell et al4 reported respiratory frequency strongly correlated with Borg dyspnea scores, and dynamic hyperinflation strongly correlated with breathlessness during exercise in COPD.

It has been previously reported that following exercise, a decrease in EELV is associated with an improvement in breathlessness symptom score after acute bronchodilation in patients with COPD.1214 During exercise and hyperventilation, EELV increases at the same rate; but since the starting point for EELV is lower after acute bronchodilation, the peak is also lower when the same load is reached. A similar phenomenon occurs with EILV. However, no interaction has been shown for the effect of bronchodilators on either EELV or EILV.1214 The rationale to measure IC during exercise is that among all variables studied, changes in IC (baseline and peak exercise) not only showed good reproducibility, but correlated best with changes in Borg dyspnea scale.1214

In the present study, the failure to achieve reduction in EELV and EILV after administration of IB (or an equivalent increase in IC) at all levels (baseline and peak exercise and after hyperventilation) is in contrast to results previously reported.114 However, 500 µg nebulized IB was used, which is approximately 10 times the dose used in the present study.114

Attempts have been made to find simpler and less expensive tests to objectively assess exercise limitation and response to bronchodilator therapy.1516 Marin et al15 showed that IC decreased, using a standard 6-min walk test, as with standard exercise testing among patients with COPD. We are not aware of previously published studies using a standardized method of hyperventilation and comparison with exercise in patients with COPD. Metronome-paced hyperventilation at twice the resting breathing frequency is a simple maneuver that can be carried out by almost any patient and can be done in any respiratory laboratory. If the decrease in IC following metronome-paced hyperventilation is as reproducible as that following exercise,11417 it might be used as a screening surrogate for exercise-induced dynamic hyperinflation and to assess response to bronchodilators in patients with COPD. Normal age-matched control subjects in the present and previous study9 did not decrease IC with exercise, as do patients with COPD.17 Furthermore, since the extent of dynamic hyperinflation and decrease in IC with exercise is related to resting IC,17 hyperventilation testing should be provocative in patients with COPD with moderate-to-severe expiratory airflow limitation.

The response to bronchodilation with 270 µg aerosolized albuterol sulfate and to that with 54 µg aerosolized IB was not considered an inclusion, and the median increase in FEV1 percentage of predicted was 5%, and only 2 of 16 patients with COPD were responders according to American Thoracic Society criteria.18 In a previous study,14 16 of 29 patients with COPD (55%) met the American Thoracic Society criteria18 for responders following administration of 500 µg aerosolized IB, yet there was poor correlation with exercise endurance time. Lack of response or poor reproducibility of FEV1 are among the reasons that prompted the search for other objective markers of abnormal lung function that correlate with symptoms and are responsive and reproducible in patients with COPD.1214161920

If confirmed by other investigators, we suspect the decrease in IC obtained during metronome-paced hyperventilation at twice the resting respiratory rate for 20 s may be used as a screening surrogate for exercise testing in the evaluation and response to dynamic hyperinflation before and after aerosolized bronchodilator in patients with COPD with moderate-to-severe expiratory airflow limitation. The goal of the present study was not to usurp standardized cardiopulmonary testing with its inherent merits, but rather to simply the detection of dynamic hyperinflation using metronome-paced hyperventilation as a screening surrogate.


    Acknowledgements
 
The authors thank Christy Kirkendall for patient coordination, Aia White-Podue for manuscript preparation, Chris M. Shinar, PharmD, for computer assistance, and Michelle Curry for patient scheduling.


    Footnotes
 
Abbreviations: CI = confidence interval; EELV = end-expiratory lung volume; EILV = end-inspiratory lung volume; IB = ipratropium bromide; IC = inspiratory capacity; NS = not significant; TLC = total lung capacity

Received for publication September 29, 2003. Accepted for publication July 6, 2004.


    References
 TOP
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 References
 

  1. O’Donnell, DE, Lam, M, Webb, KA (1998) Measurement of symptoms, lung hyperinflation and endurance during exercise in chronic obstructive pulmonary disease. Am J Respir Crit Care Med 158,1557-1565[Abstract/Free Full Text]
  2. Belman, MJ, Botnick, WC, Shin, JC Inhaled bronchodilators reduce dynamic hyperinflation during exercise in patients with chronic obstructive pulmonary disease. Am J Respir Crit Care Med 1996;153,967-975[Abstract]
  3. Dodd, DS, Brancatisano, T, Engel, LA Chest wall mechanics during exercise in patients with severe chronic airflow obstruction. Am Rev Respir Dis 1984;12,33-38
  4. O’Donnell, DE, Webb, KA Exertional breathlessness in patients with chronic airflow limitation: the role of lung hyperinflation. Am Rev Respir Dis 1993;148,1351-1357[ISI][Medline]
  5. Stubbing, DG, Pengelly, LD, Morse, JLC, et al Pulmonary mechanics during exercise in subjects with chronic airflow obstruction. J Appl Physiol 1980;49,511-515[Abstract/Free Full Text]
  6. Duranti, R, Filipelli, M, Bianchi, R, et al Inspiratory capacity and decrease in lung hyperinflation with albuterol in COPD. Chest 2002;122,2009-2014[Abstract/Free Full Text]
  7. O’Donnell, DE, Chau, LKL, Bertley, JC, et al Qualitative aspects of exertional breathlessness in chronic airflow limitation: pathophysiologic mechanisms. Am J Respir Crit Care Med 1997;155,109-115[Abstract]
  8. Yan, S, Kaminski, D, Sliwinski, P Reliability of inspiratory capacity for estimating end-expiratory lung volume changes during exercise in patients with chronic obstructive pulmonary disease. Am J Respir Crit Care Med 1997;156,55-59[Abstract/Free Full Text]
  9. Gutierrez, CA, Chakravorty, S, Latty, P, et al Reproducibility of inspiratory capacity during baseline resting conditions and hyperventilation in normal subjects [abstract]. Am J Respir Crit Care Med 2002;165,A499
  10. Fabbri, LM, Hurd, SS Global strategy for the diagnosis, management and prevention of COPD: 2003 update. Eur Respir J 2003;22,1-2[Free Full Text]
  11. Gelb, AF, Zamel, N Pseudophysiologic emphysema in chronic persistent asthma. Am J Respir Crit Care Med 2000;162,1778-1782[Abstract/Free Full Text]
  12. Henke, KG, Sharratt, M, Pegelow, D, et al Regulation of end-expiratory lung volume during exercise. J Appl Physiol 1988;64,135-146[Abstract/Free Full Text]
  13. Hamilton, AL, Killian, KJ, Summers, E, et al Muscle strength, symptom intensity, and exercise capacity in patients with cardiorespiratory disorders. Am J Respir Crit Care Med 1995;152,2021-2031[Abstract]
  14. O’Donnell, DE, Lam, M, Webb, KA Spirometric correlates of improvement in exercise performance after anticholinergic therapy in chronic obstructive pulmonary disease. Am J Respir Crit Care Med 1999;160,542-549[Abstract/Free Full Text]
  15. Marin, JM, Carrizo, SJ, Gascon, M, et al Inspiratory capacity, dynamic hyperinflation, breathlessness, and exercise performance during the 6-minute walk test with chronic obstructive pulmonary disease. Am J Respir Crit Care Med 2001;163,1395-1399[Abstract/Free Full Text]
  16. Taube, C, Lehnigk, B, Paasch, K, et al Factor analysis of changes in dyspnea and lung function parameters after bronchodilation in chronic obstructive pulmonary disease. Am J Respir Crit Care Med 2000;162,216-220[Abstract/Free Full Text]
  17. O’Donnell, DE, Revill, SM, Webb, KA Dynamic hyperventilation and exercise intolerance in chronic obstructive pulmonary disease. Am J Respir Crit Care Med 2001;154,770-774
  18. American Thoracic Society. Lung function testing: selection of reference values and interpretative strategies. Am J Respir Crit Care Med 1991;144,1202-1218
  19. Guyatt, GH, Townsend, M, Nogradi, S, et al Acute response to bronchodilator: an imperfect guide for bronchodilator therapy in chronic airflow limitation. Arch Intern Med 1988;148,1949-1952[Abstract]
  20. Hatipoglu, U, Laghi, F, Tobin, MJ Does inhaled albuterol improve diaphragmatic contractility in patients with chronic obstructive pulmonary disease? Am J Respir Crit Care Med 1999;160,1916-1921[Abstract/Free Full Text]



This article has been cited by other articles:


Home page
ChestHome page
A. F. Gelb, C. F. Taylor, P. A. McClean, C. M. Shinar, M. T. Rodrigues, C. A. Gutierrez, K. R. Chapman, and N. Zamel
Tiotropium and Simplified Detection of Dynamic Hyperinflation
Chest, March 1, 2007; 131(3): 690 - 695.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (8)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Gelb, A. F.
Right arrow Articles by Zamel, N.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Gelb, A. F.
Right arrow Articles by Zamel, N.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS