* From the Pulmonary Division (Dr. Gelb), Department of Medicine Lakewood Regional Medical Center, Lakewood, CA; Faculty of Medicine (Drs. Gutierrez and Zamel), University of Toronto, Toronto, ON, Canada; Department of Clinical Investigation and Pulmonary Critical Care Medicine (Dr. Weisman), William Beaumont Army Medical Center, El Paso, TX; and independent contractors (Mr. Newsom and Ms. Taylor), Lakewood, CA.
Correspondence to: Arthur F. Gelb, MD, FCCP, 3650 E South St, Suite 308, Lakewood, CA 90712; e-mail: afgelb{at}msn.com
| Abstract |
|---|
|
|
|---|
Methods: IC and FEV1 were measured before and after metronome-paced hyperventilation at twice the resting respiratory rate for 20 s in 16 patients with COPD before and after 54 µg aerosolized ipratropium bromide (IB). We also studied the same 16 patients before and after administration of 54 µg aerosolized IB during symptom-limited incremental cycle ergometry when the final respiratory rate was also twice the resting rate.
Results: Resting IC was 2.23 ± 0.53 L (mean ± SD), and the mean decrease in IC from baseline was 0.36 ± 0.25 L after exercise (p < 0.001), and not significantly different (p = 0.64) from mean decrease in IC of 0.40 ± 0.29 L following hyperventilation. Results following hyperventilation and exercise were similar after bronchodilator. The mean difference for decrease of IC between hyperventilation and exercise was 0.138 L (95% confidence interval, 0.347 to 0.622; r = 0.66, p = 0.006). The decrease in FEV1 was 0.01 ± 0.13 L after exercise and 0.06 ± 0.18 L after hyperventilation. Separately, baseline and peak end-expiratory and end-inspiratory lung volumes were similar with hyperventilation vs exercise both before and after bronchodilator.
Conclusion: Both metronome-paced hyperventilation and incremental cycle ergometry, when resting respiratory rate was doubled, provoked similar significant decrease in IC, even after administration of 54 µg aerosolized IB. The noninvasive simplicity of hyperventilation for 20 s provides a clinically useful screening surrogate to monitor changes in IC following exercise.
Key Words: COPD dynamic hyperinflation exercise hyperventilation lung function
| Introduction |
|---|
|
|
|---|
ODonnell et al1 have provided confirmatory evidence that "Borg dyspneic ratings and measurements of inspiratory capacity (IC) and endurance time during submaximal cycle exercise testing are highly reproducible and responsive to changes in severe COPD." Previous studies have demonstrated the utility of repeated measurements of IC during exercise to reflect changes in end-expiratory lung volume (EELV),234 since total lung capacity (TLC) remains constant after acute bronchodilation and during exercise.56 Progressive reduction in IC during exercise reflects dynamic hyperinflation, and is a good predictor of decreased exercise ability as well as increased exertional dyspnea.4 Additionally, peak values of inspiratory esophageal pressures used as a surrogate to estimate effort are relatively constant during multiple measurements of exercise IC.78
Changes in IC reflect dynamic hyperinflation and are correlated with breathing frequency in patients with COPD.4 We suspected increasing breathing frequency would produce changes in IC similar to the change in IC observed during exercise.
The present study was designed to compare metronome-paced hyperventilation, a relatively simple procedure, with incremental symptom-limited cycle ergometry to provoke respiratory rate-induced reduction of IC in patients with COPD. A similar comparison has not been previously reported. We also wanted to evaluate the role of inhaled ipratropium bromide (IB) to blunt the decrease in IC. ODonnell et al1 have previously shown greater bronchodilation with IB when compared to inhaled albuterol sulfate in patients with COPD. In a preliminary study,9 we previously reported that hyperventilation in younger normal subjects resulted in insignificant changes in IC.
| Materials and Methods |
|---|
|
|
|---|
After obtaining informed consent, patients underwent lung function studies before and after administration of 270 µg aerosolized albuterol sulfate using techniques and predictive values previously described in detail.11 We used a pressure-compensated flow plethysmograph (Model 6200; SensorMedics; Viasys; Yorba Linda, CA).
Subsequently, on a separate day, metronome-paced hyperventilation was initiated in 16 patients with COPD to achieve a respiratory rate twice the baseline rate for 20 s, which was immediately followed by sequential measurement of IC and expiratory spirometry. While no attempt was made to control end-tidal carbon dioxide, patients were coached to maintain a respiratory rate synchronous with the metronome. Near-constant dynamic tidal volume during metronome-paced hyperventilation was achieved by having patients observe a graphic display of their breathing pattern. Patients were studied before and after administration of 54 µg aerosolized IB from a metered-dose inhaler. The technique for measuring IC has been previously described.14
On a separate day, symptom-limited incremental exercise was initiated in all of the 16 patients with COPD after baseline values were achieved, using an electronically braked cycle ergometer (Ergometrics 800; SensorMedics) before and 15 min after administration of 54 µg aerosolized IB. Following unloaded 2-min warm-up, a 10-W workload was increased at 1-min intervals during near-constant pedaling at 50 revolutions per minute. Immediately following symptom-limited peak exercise at constant workload for 1 min, measurements of IC and expiratory spirometry were obtained. During exercise, expired gases were collected and analyzed (
max 29; SensorMedics). The protocol is described in Table 1
. We also studied 21 younger and 8 older nonsmoking healthy volunteers to evaluate the changes in IC during metronome-paced hyperventilation at twice the resting respiratory rate for 20 s.
|
| Results |
|---|
|
|
|---|
We studied 16 patients with COPD (10 men) aged 69 ± 6 years with a smoking history of 50 ± 15 pack-years. Results of routine lung function studies described in Table 2 are consistent with moderate-to-severe COPD. Following administration of 270 µg aerosolized albuterol sulfate, FEV1 increased 13 ± 12%.
|
|
|
After administration of 54 µg aerosolized IB, the decrease in IC was 0.32 ± 0.23 L following exercise to 43 ± 5 W, minute ventilation was 33 ± 4L/min, respiratory rate was 30 ± 4 breaths/min, and oxygen uptake was 0.9 ± 0.3 L/min. This was similar to the decrease in IC of 0.39 ± 0.29 L following metronome-paced hyperventilation at twice the resting respiratory rate for 20 s (p = NS) [Table 3, Fig 1].
Correlation Between Hyperventilation and Exercise: Before bronchodilator, the mean difference between the peak change in IC comparing hyperventilation and exercise was 0.04 L (95% CI, 0.419 to 0.619 L; r = 0.41, p = 0.115). After bronchodilator, the mean difference between the peak change in IC comparing hyperventilation and exercise was 0.07 L (95% CI, 0.347 to 0.622 L; r = 0.66, p = 0.006).
Bland and Altman Method for Assessing Agreement Between Hyperventilation vs Exercise
Results demonstrate bias for hyperventilation induced greater decrease in IC both before (100 mL) and after (130 mL) nebulized IB vs exercise (Fig 2
). Furthermore, with the exception of one outlier, all values were within ± 2 SD and the majority of observations were within 1 SD of mean difference.
|
EELV and EILV
Prebronchodilator baseline EELV equivalent to functional residual capacity was 70 ± 7% of TLC in 16 patients, who underwent both hyperventilation and cycle ergometry testing (Table 4
; Fig 3
). EILV was 82 ± 7% of TLC. Following metronome-paced hyperventilation, EELV was 76 ± 6% of TLC and EILV was 88 ± 5% of TLC (p < 0.001). Following the administration of 54 µg IB, baseline EELV was 69 ± 6% TLC and EILV was 82 ± 6% TLC; after hyperventilation, EELV was 76 ± 6% TLC and EILV was 88 ± 6% TLC. After bronchodilator administration, these changes are similar to prebronchodilator changes (p = 0.37). Thus, acute bronchodilation failed to blunt the increase in EELV and EILV percentage of TLC following hyperventilation.
|
|
| Discussion |
|---|
|
|
|---|
The correlation between the decrease in IC following cycle ergometry vs hyperventilation was strengthened after aerosolized IB with reduction in vagal tone. However, a high correlation is not a good test of agreement since it reflects only the relationship between two variables. Agreement is better tested using the Bland and Altman method (Fig 2). Results suggest hyperventilation is more sensitive than cycle ergometry to detect decrease in IC both before and after aerosolized IB.
Additionally, increases in EELV and EILV were similar with hyperventilation and exercise, and not blunted with acute bronchodilation with inhaled IB. IB is presumably but not clearly proven a greater bronchodilator than inhaled albuterol sulfate in patients with COPD.1 This may also explain lack of improvement in dyspnea following use of albuterol and/or IB.
In normal subjects during exercise, minute ventilation increases primarily as a result of increased tidal volume rather than increased respiratory rate.12 In contrast, when patients with COPD with at least moderately severe expiratory airflow limitation and increased EELV undergo symptom limited exercise, we and others4713 have noted the increase in respiratory rate to approximately double baseline was primarily responsible for increased minute ventilation. Not surprisingly, ODonnell et al4 reported respiratory frequency strongly correlated with Borg dyspnea scores, and dynamic hyperinflation strongly correlated with breathlessness during exercise in COPD.
It has been previously reported that following exercise, a decrease in EELV is associated with an improvement in breathlessness symptom score after acute bronchodilation in patients with COPD.1214 During exercise and hyperventilation, EELV increases at the same rate; but since the starting point for EELV is lower after acute bronchodilation, the peak is also lower when the same load is reached. A similar phenomenon occurs with EILV. However, no interaction has been shown for the effect of bronchodilators on either EELV or EILV.1214 The rationale to measure IC during exercise is that among all variables studied, changes in IC (baseline and peak exercise) not only showed good reproducibility, but correlated best with changes in Borg dyspnea scale.1214
In the present study, the failure to achieve reduction in EELV and EILV after administration of IB (or an equivalent increase in IC) at all levels (baseline and peak exercise and after hyperventilation) is in contrast to results previously reported.114 However, 500 µg nebulized IB was used, which is approximately 10 times the dose used in the present study.114
Attempts have been made to find simpler and less expensive tests to objectively assess exercise limitation and response to bronchodilator therapy.1516 Marin et al15 showed that IC decreased, using a standard 6-min walk test, as with standard exercise testing among patients with COPD. We are not aware of previously published studies using a standardized method of hyperventilation and comparison with exercise in patients with COPD. Metronome-paced hyperventilation at twice the resting breathing frequency is a simple maneuver that can be carried out by almost any patient and can be done in any respiratory laboratory. If the decrease in IC following metronome-paced hyperventilation is as reproducible as that following exercise,11417 it might be used as a screening surrogate for exercise-induced dynamic hyperinflation and to assess response to bronchodilators in patients with COPD. Normal age-matched control subjects in the present and previous study9 did not decrease IC with exercise, as do patients with COPD.17 Furthermore, since the extent of dynamic hyperinflation and decrease in IC with exercise is related to resting IC,17 hyperventilation testing should be provocative in patients with COPD with moderate-to-severe expiratory airflow limitation.
The response to bronchodilation with 270 µg aerosolized albuterol sulfate and to that with 54 µg aerosolized IB was not considered an inclusion, and the median increase in FEV1 percentage of predicted was 5%, and only 2 of 16 patients with COPD were responders according to American Thoracic Society criteria.18 In a previous study,14 16 of 29 patients with COPD (55%) met the American Thoracic Society criteria18 for responders following administration of 500 µg aerosolized IB, yet there was poor correlation with exercise endurance time. Lack of response or poor reproducibility of FEV1 are among the reasons that prompted the search for other objective markers of abnormal lung function that correlate with symptoms and are responsive and reproducible in patients with COPD.1214161920
If confirmed by other investigators, we suspect the decrease in IC obtained during metronome-paced hyperventilation at twice the resting respiratory rate for 20 s may be used as a screening surrogate for exercise testing in the evaluation and response to dynamic hyperinflation before and after aerosolized bronchodilator in patients with COPD with moderate-to-severe expiratory airflow limitation. The goal of the present study was not to usurp standardized cardiopulmonary testing with its inherent merits, but rather to simply the detection of dynamic hyperinflation using metronome-paced hyperventilation as a screening surrogate.
| Acknowledgements |
|---|
| Footnotes |
|---|
Received for publication September 29, 2003. Accepted for publication July 6, 2004.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A. F. Gelb, C. F. Taylor, P. A. McClean, C. M. Shinar, M. T. Rodrigues, C. A. Gutierrez, K. R. Chapman, and N. Zamel Tiotropium and Simplified Detection of Dynamic Hyperinflation Chest, March 1, 2007; 131(3): 690 - 695. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |