|
|
||||||||
Guest Access | Sign In via User Name/Password |
|||||||||
* From the Department of Medicine (Drs. Wonisch, Lercher, Scherr, and Maier), Division of Cardiology, Medical University, Graz, Austria; Institute of Sports Physiology (Dr. Pokan), University of Vienna, Austria; Institute of Sport Sciences (Dr. Hofmann), Karl-Franzens University, Graz, Austria; and Human Performance Laboratory (Dr. von Duvillard), Department of Health, Kinesiology & Sports Studies, Texas A&M University, Commerce, TX.
Correspondence to: Manfred Wonisch, MD, PhD, Department of Medicine, University Hospital, Division of Cardiology, A-8036 Graz, Auenbruggerplatz 15; e-mail: manfred.wonisch{at}meduni-graz.at
| Abstract |
|---|
|
|
|---|
Design and setting: The purpose of this study was to show the impact of permanent right ventricular (RV) pacing on exercise capacity and related cardiorespiratory parameters in patients with chronic heart failure and ICDs.
Patients and interventions: Seventeen patients with chronic heart failure and a dual-chamber ICD performed cardiopulmonary exercise testing (CPX) on 3 different days. After CPX 1, patients were randomized either to back-up pacing or permanent RV pacing. After 3 months, CPX 2 was performed and patients changed groups (crossover design); CPX 3 was performed after 3 additional months.
Measurements and results: Maximal values for workload (108 ± 46 W vs 117 ± 48 W, p < 0.01), oxygen uptake (
O2) [21.0 ± 5.3 mL/min/kg vs 22.5 ± 6.4 mL/min/kg, p < 0.05], oxygen pulse (13 ± 3.7 mL vs 14 ± 4.0 mL, p < 0.05), and metabolic equivalent (6.0 ± 1.5 vs 6.4 ± 1.8, p < 0.05) were significantly lower with permanent RV pacing compared to back-up pacing. Workload,
O2, and oxygen pulse were significantly reduced at the ventilatory anaerobic threshold, while workload and
O2 were significantly lower at the respiratory compensation point. No differences were found for maximal heart rate, minute ventilation
E, and respiratory exchange ratio. The
E/carbon dioxide production slope was significantly steeper with permanent RV pacing compared to back-up pacing.
Conclusions: Permanent RV pacing significantly reduced maximal and submaximal measures of exercise. For patients with chronic heart failure and sufficient atrioventricular conduction, every effort should be made to minimize permanent right ventricular pacing.
Key Words: cardiopulmonary exercise testing dual-chamber implantable cardioverter defibrillator maximal oxygen consumption minute ventilation/carbon dioxide production slope respiratory compensation point ventilatory anaerobic threshold ventilatory efficiency
| Introduction |
|---|
|
|
|---|
Information from the atrial chamber allows better device programming and individualization of drug therapy for ventricular arrhythmias.4 Therefore, most currently implanted ICDs are dual-chamber devices.3 Moreover, positive short-term and long-term effects of dual-chamber RV pacing in the treatment of end-stage idiopathic dilated cardiomyopathy have been shown.567 However, 4 to 18% of ICD patients were found to need antibradycardia pacing.38 At the moment, however, it is unclear whether dual-chamber pacemakers (DDDs) should be implanted, or whether permanent DDD pacing should be activated in modern ICDs in every patient with chronic heart failure regardless of the intrinsic atrioventricular conduction delay.
The maximal oxygen uptake (
O2max) is used for risk stratification in chronic heart failure patients.9 In some cases,
O2max might be underestimated because of reduced patient motivation or premature termination of exercise on the examiners decision. Furthermore, patients with severe heart failure tend to perform activities of daily life that involve submaximal measures of exercise.10 Submaximal exercise parameters such as the ventilatory anaerobic threshold, the respiratory compensation point, and ventilatory efficiency (minute ventilation [
E]/carbon dioxide production [
CO2] slope) are less subject to these influences.10111213 The
E/
CO2 slope in particular was found to be a reliable predictor of prognosis in patients with chronic heart failure.111214 The purpose of this study was to show the impact of permanent RV pacing on exercise capacity and submaximal cardiorespiratory parameters in patients with chronic heart failure and dual-chamber ICDs.
| Materials and Methods |
|---|
|
|
|---|
Inclusion Criteria
Inclusion criteria included the following: a class I indication for an ICD1 (ie, history of ventricular tachycardia or survivor of sudden cardiac death), age > 18 years, left ventricular ejection fraction (LVEF) < 40%, permanent sinus rhythm, and medication not changed during the investigation period.
Exclusion Criteria
Exclusion criteria included the following: chronic atrial fibrillation, history of paroxysmal atrial fibrillation, sick sinus syndrome, need for permanent atrioventricular-synchronous RV pacing, an atrioventricular delay < 150 ms, female patients with child-bearing potential, age > 80 years, and hypertrophic obstructive cardiomyopathy.
Programming of the ICD
Antitachycardia Parameters:
The device was programmed individually according to the underlying ventricular arrhythmia. This programming was maintained during the 6-month follow-up period unless ventricular tachyarrhythmias developed for which the antitachycardic stimulation therapy was inappropriate or ineffective.
Antibradycardia Parameters: The lower rate was programmed to 40 to 50 beats/min for all patients. The upper rate of the pacemaker was programmed according to the age-predicted maximal heart rate, 130 to 160 beats/min.
Patients were randomized in a crossover design to either of the following: (1) intrinsic atrioventricular-sequential conduction without ventricular stimulation (DDD with a long atrioventricular delay or DDI) [back-up pacing group]; or (2) fixed atrioventricular delay of 100 to 120 ms ensuring permanent RV stimulation. At the 3-month follow-up, programming was switched to correspond to the other group so that the study patients served as their own control group.
CPX
Subjects completed CPX on an electronically braked cycle ergometer (ER 900; Jaeger; Wuerzburg, Germany) to volitional exhaustion on 3 different days: The first test (CPX 1) was intended to familiarize patients with the testing procedure and to determine their maximal workloads. To obtain baseline values, patients were asked to sit quietly on the cycle ergometer for 10 min. Resting values were collected in the last minute before the onset of the workload. Exercise was started with a workload of 10 W followed by an increase of 10 W for men and 7 W for women every minute thereafter. A 12-lead ECG (Jaeger) was used to monitor and record ECG tracings. A physician supervised each test. During CPX, subjects wore an airtight mask over their nose and mouth. Oxygen uptake (
O2) and
CO2 were analyzed continuously during the tests using a breath-by-breath system (Oxycon-Pro; Jaeger). Averaged values of these parameters were recorded every 20 s. To obtain an optimal exercise time of approximately 10 min, an individual approach for workload increments was chosen for the following two tests.15 The workload achieved from CPX 1 was divided by 10 to determine the incremental workload step for further testing.
In this randomized crossover study, CPX 2 and CPX 3 were done after 3 months and 6 months with either permanent atrioventricular synchronous ventricular pacing (permanent RV pacing) or permanent intrinsic atrioventricular conduction (back-up pacing) [Fig 1 ]. To keep the observer blinded, ECG tracings at CPX 2 and CPX 3 were stored but not shown on-line during the testing procedure.
|
O2 achieved during the test was defined as
O2max regardless of the establishment of a
O2 plateau. The respiratory equivalents for oxygen (
E/
O2) and carbon dioxide (
E/
CO2) were calculated off-line.
The
E/
CO2 slope11121416 was calculated off-line as an index of the ventilatory response to exercise. This slope represents the regression line relating
E to
CO2 using all unaveraged (ie, breath-by-breath) data points from rest to peak exercise. The slope computed from all data points for the entire range of an exercise test has been shown to have the highest prognostic value compared to submaximal measurement.14 The oxygen pulse, as the quotient of
O2 (in milliliters) divided by the heart rate, is determined by stroke volume and the arteriovenous oxygen difference.10
Submaximal Values:
Two submaximal values based on a three-phase model were analyzed with a computer-aided linear regression breakpoint analysis:1718 The ventilatory anaerobic threshold was defined as the point that corresponded to the initial deviation from linearity in
E and the onset of a systematic rise in the respiratory equivalent for
O2 without a simultaneous increase in the respiratory equivalent for
CO2. The respiratory compensation point was defined between the ventilatory anaerobic threshold and maximal workload as follows: the secondary rise in
E that corresponded to the secondary rise in the respiratory equivalent for
O2. This point is compatible to the nadir of the respiratory equivalent for
CO2 and is followed by a marked rise in respiratory equivalent for
CO2.131718
Statistics:
Continuous data were tested for normality using the Kolmogorov-Smirnov test. Data are reported as mean ± SD values. The paired Student t test was used to determine significant differences between permanent RV pacing and back-up pacing for all measured and calculated variables. The level of significance was set at
p < 0.05.
| Results |
|---|
|
|
|---|
Mean intrinsic atrioventricular delay for all patients at baseline was 191 ± 23 ms. Intrinsic left bundle-branch block was found in 7 patients, and normal QRS complex was found in the remaining 10 patients. A significant difference (p < 0.001) was found for mean QRS width during back-up pacing (125 ± 32 ms) and permanent RV pacing (176 ± 31 ms). During the back-up pacing period, 3.6 ± 5% of RV beats were paced. During the permanent RV pacing period, RV pacing was found in 99 ± 2% of beats.
CPX
Resting Values:
Resting values for heart rate,
O2,
E, and oxygen pulse (measured at CPX 2 and CPX 3) did not differ between back-up pacing and permanent RV pacing (Table 1
).
|
O2max, 21.5 ± 5.5 mL/kg/min; metabolic equivalent, 6.1 ± 1.6; maximum testing time; 708 ± 220 s; maximum heart rate, 135 ± 23 beats/min; maximum
E, 62 ± 14 L/min; maximal oxygen pulse, 12.8 ± 4.0 mL; and maximal respiratory exchange ratio, 1.03 ± 0.1. Permanent RV DDD pacing significantly reduced maximal workload by 8%, maximal
O2 by 7%, metabolic equivalent by 6%, and exercise time by 8% compared to back-up pacing (Table 1). Maximal oxygen pulse as surrogate parameter of stroke volume was also significantly reduced by 7% with permanent RV pacing. No differences were found for maximal heart rate, maximal
E, and maximal respiratory exchange ratio at peak exercise (Table 1).
Submaximal Parameters:
The ventilatory anaerobic threshold could be determined for every patient in all tests. Permanent RV pacing significantly reduced workload by 23%,
O2 by 10%, and oxygen pulse by 9% at the ventilatory anaerobic threshold compared to back-up pacing. No significant difference was found for heart rate at the ventilatory anaerobic threshold (Table 2
). The respiratory compensation point could be determined in 14 patients in the permanent RV pacing group and 15 patients in the back-up pacing group. Means ± SD were only calculated with the complete data set for every patient (n = 14). Permanent RV pacing significantly reduced workload by 18% and
O2 by 8% at the respiratory compensation point compared with back-up pacing. Heart rate at the respiratory compensation point also tended to be lower with permanent RV pacing. Although the oxygen pulse at the respiratory compensation point was found to be higher for back-up pacing, this difference did not reach statistical significance (Table 2). The
E/
CO2 slope was significantly steeper with permanent RV pacing than with back-up pacing (Table 1).
|
| Discussion |
|---|
|
|
|---|
Permanent RV pacing is uncommon in patients with pacemakers having adequate atrioventricular conduction. Cardiac stimulation in the apex of the right ventricle has unfavorable effects with induction of a left bundle-branch block and consequent disturbances in the contractility and hemodynamic parameters.1920 However, RV apical pacing has been used as an adjunctive treatment for heart failure regardless of intrinsic atrioventricular conduction. Improvement of hemodynamics was demonstrated using a short atrioventricular delay,5621 but results were not consistent.2223 All the above-mentioned studies used an atrioventricular delay between 90 ms and 130 ms. In our study, the atrioventricular delay for permanent pacing was programmed at 100 to 120 ms according to the studies that showed the greatest benefit.5621 Although an additional negative influence of a too-short atrioventricular interval cannot fully be excluded, an atrioventricular delay of at least 100 ms was found to be the shortest possible delay that does not significantly impair cardiac function in patients with DDD pacemaker therapy.24 Furthermore, in patients with the need for permanent RV pacing, it has been shown that small differences in atrioventricular delay have far less influence on left ventricular systolic function than pacing site.22
Maximal Measures of Exercise
O2max measured during CPX has been considered as a reference measurement in patients with heart failure.925 Patients with left bundle-branch block show lower exercise capacity and higher mortality. Biventricular pacing resulted in higher exercise capacity due to a normalization of the QRS width and normalization of contraction abnormalities.1026 Moreover, in patients with preexisting left bundle-branch block, a close relation has been found between the degree of QRS narrowing with biventricular pacing and clinical improvement.2627
The oxygen pulse was significantly reduced at maximal workload (Table 1). The
O2 per heart beat represents a substitute for the stroke volume, as the arteriovenous oxygen difference is constant. The decrease in
O2max was independent of heart rate response, as shown by the significantly higher oxygen pulse with back-up pacing. This would be consistent with the increase in stroke volume observed with biventricular pacing compared with an uncorrected left bundle-branch block.1028
A possible explanation could be impaired coronary blood flow due to RV pacing. Alterations in regional myocardial blood flow associated with permanent RV pacing were found in experimental animal29 and human studies.3031 An impairment of microvascular flow was found to be the underlying mechanism.3031 Reduced regional myocardial work,2932
O2,32 and free fatty acid metabolism33 have been observed in the early activated region during ventricular pacing.
Submaximal Measures of Exercise
An important aim of treatment for severe heart failure is to increase the patients ability to perform activities of daily life. Parameters that reflect submaximal activities may be more useful than maximal exercise parameters for the assessment of the efficacy of treatment. A significant improvement of workload and
O2 at the ventilatory anaerobic threshold and the respiratory compensation point was seen with back-up pacing compared to permanent RV pacing.
The
E/
CO2 slope provides a noninvasive assessment of the appropriateness or efficiency of ventilation during exercise and has been shown to be useful for risk stratification in chronic heart failure patients with intermediate exercise capacity.1116 We found a small but significantly steeper slope with permanent RV stimulation. In chronic heart failure, this slope is generally increased, implying increased ventilatory drive. Normalization of left bundle-branch block with biventricular pacing has been shown to increase cardiac output28 and to reduce left atrial pressure,34 leading to less pulmonary congestion, which in turn may improve ventilatory efficiency. Such changes might be expected to reduce ventilation-perfusion mismatch and to account for the dramatic improvement in ventilatory drive observed with biventricular pacing.
Methodologic Considerations and Limitations
This was a single-center study. Although a total of 198 ICD patients were available, only 17 could be analyzed. Differences were small but reached statistical significance.
The study duration was fairly short, with each intervention performed for 3 months. Further changes would have been seen between the two different pacing interventions over a longer period resulting in more pronounced differences in measured parameters.
A change in training status could influence maximal and submaximal parameters of exercise. Although we did not control for physical activity or exercise during the 3 months of back-up pacing and permanent RV pacing, due to the crossover design of the study an influence seems very unlikely.
Exercise parameters may be of importance for the explanation of previous results found by the Second Multicenter Automated Defibrillator Implantation Trial2 and the Dual Chamber and VVI Implantable Defibrillator trial.3 Parameters such as humoral factors or quality of life might be of interest in future studies.
| Conclusion |
|---|
|
|
|---|
| Footnotes |
|---|
CO2 = carbon dioxide production;
E = minute ventilation;
O2 = oxygen uptake;
O2max = maximal oxygen uptake Dr. Wonisch and Dr. Lercher contributed equally to the study.
Received for publication June 2, 2004. Accepted for publication October 7, 2004.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
Z. Goldberger and R. Lampert Implantable Cardioverter-Defibrillators: Expanding Indications and Technologies JAMA, February 15, 2006; 295(7): 809 - 818. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |