| Abstract |
|---|
|
|
|---|
Design: The new monitoring technique was compared to established reference methods.
Setting: ICU and sleep laboratory of a university hospital.
Patients: Eighteen critically ill adult patients with acute respiratory failure or heart failure, and 12 patients with sleep apnea (mean [± SD] apnea/hypopnea index, 43 ± 24 events per hour).
Measurements: Continuous measurements were performed over several hours by the novel heated (temperature, 42°C) earlobe sensor (TOSCA; Linde Medical Sensors; Basel, Switzerland), incorporating electrochemical and optical elements for carbon dioxide measurement (PtcCO2) and pulse oximetry (SpO2), respectively. The data were compared to the results of repeated arterial blood gas analyses in critically ill patients and to simultaneous nocturnal pulse oximetry performed with different devices with earlobe or finger sensors in sleep apnea patients.
Results: In critically ill patients, the mean difference and limits of agreement (bias ± 2 SDs) of transcutaneous PtcCO2 vs arterial PaCO2 were 3 ± 7 mm Hg; the corresponding values for changes in PtcCO2 vs PaCO2 were 1 ± 6 mm Hg. The bias ± 2 SDs for pulse oximetric SpO2 vs arterial oxygen saturation (SaO2) were 1 ± 4%. In sleep apnea patients, the combined earlobe sensor identified more transient oxygen desaturations, and the rate of change in oxygen saturation during events was greater compared to those with other tested pulse oximeters, indicating a faster response.
Conclusions: Due to its ability to accurately assess both ventilation and oxygenation by a single transcutaneous sensor, the described noninvasive monitoring technique is a valuable tool for respiratory monitoring with potential applications in critical care and sleep medicine.
Key Words: capnometry carbon dioxide critical care oxygenation physiologic monitoring pulse oximetry sleep apnea sleep medicine ventilation
| Introduction |
|---|
|
|
|---|
Previous transcutaneous blood gas sensors, which were mainly used in pediatric care, have incorporated PtcCO2 in combination with transcutaneous partial pressure of oxygen measurement.89 However, in adults, transcutaneous partial pressure of oxygen depends heavily on local skin perfusion and does not reliably reflect systemic PaO2.10 Recently, a novel combined sensor for the measurement of both PtcCO2 and pulse oximetric saturation (SpO2) [TOSCA; Linde Medical Sensors; Basel, Switzerland] has been developed. It contains an electrochemical electrode (for PtcCO2 measurement), a light emitter/sensor (for SpO2 measurement), and a heating element (to increase local perfusion). The small size of the sensor allows convenient placement on the earlobe. Since the principles of the noninvasive monitoring of ventilation and oxygenation with a single cutaneous sensor are sound, and since the initial results obtained in healthy subjects and patients during anesthesia were promising,1112 we performed a clinical evaluation of the novel sensor in the following two settings: (1) in critically ill adult patients, the accuracy of PtcCO2 and SpO2 measurements by the novel sensor was compared to PaCO2 and SaO2 measurements from blood samples repeatedly drawn from indwelling arterial lines; and (2) in patients with obstructive sleep apnea syndrome, the response characteristics of SpO2 by the novel sensor were compared to those of several other pulse oximeters during rapid fluctuations in arterial oxygenation. In addition, the effect of sensor temperature on pulse oximeter performance was evaluated by comparing SpO2 measured simultaneously by a heated and a nonheated earlobe sensor.
| Materials and Methods |
|---|
|
|
|---|
Twelve patients with obstructive sleep apnea syndrome were studied during nocturnal polysomnography or limited sleep studies.13 Their mean age was 58 ± 8 years, mean body mass index was 35.4 ± 7.2 kg/m2, and mean apnea/hypopnea index was 43 ± 24 events per hour. The Hospital Ethics Committee approved the study.
Measurements
The novel sensor (TOSCA [software version MB1.03/DBGB1.04]; Linde Medical Sensors) has the shape of a flat cylinder (diameter x height, 14 x 7.5 mm) and is connected to the monitoring unit by a cable. It integrates a Stow-Severinghaus-type CO2 sensor,14 which is an optical sensor for measuring SpO2, and a heating element that is set to maintain a sensor temperature of 42°C.11 The calibration of the CO2 sensor is automatically performed between measurements when the sensor is placed in the storage chamber of the TOSCA unit. The sensor was fixed to the earlobe by a low-pressure clip or with double-sided adhesive tape in the shape of a ring. A reflective element was placed on the back of the earlobe. PtcCO2 and SpO2 values were displayed in real time on the TOSCA monitor.
In critically ill patients, PtcCO2 and SpO2 were continuously recorded by the earlobe sensor over a mean (± SD) duration of 160 ± 48 min. Repeated blood samples (three to six samples per patient) were drawn from the indwelling arterial line as clinically indicated to measure PaCO2 and SaO2 (RapidLab865 blood gas analysis and co-oximetry system; Bayer Health Care; Leverkusen, Germany). The time and corresponding values were recorded.
Twelve patients with obstructive sleep apnea were studied during nocturnal polysomnography or with limited sleep studies13 lasting over at least 6 h. SpO2 measurements made with the TOSCA sensor were compared with corresponding values obtained simultaneously by the following pulse oximeters: (1) Ohmeda Biox 3740 (Datex-Ohmeda; Madison, WI) with an earlobe sensor placed on the contralateral lobe; (2) Datex-Ohmeda 3740 with finger sensor; (3) Nellcor N-395 (Nellcor; Pleasanton, CA) with finger sensor; and (4) Masimo Radical (Masimo; Irvine, CA) with finger sensor. The Datex Ohmeda 3740, the Masimo Radical, and the TOSCA oximeters were operated in their fastest response mode (ie, 3, 2, and 2 s averaging time, respectively). The response mode of the Nellcor N-395 is not modifiable, and the averaging time is not stated in the user manual. The allocation of sensors of different devices to fingers 2 to 4 of the same hand was done at random, and the sensors were covered by opaque dressing to prevent cross-talk.
In seven patients with obstructive sleep apnea, the effect of sensor temperature on pulse oximeter performance was evaluated by simultaneously recording SpO2 during nocturnal polygraphy by two TOSCA sensors with one heated to 42°C, and the other, with the heating turned off, placed on the contralateral earlobe. In three of the patients, the ratio between the pulsatile and nonpulsatile components of infrared absorption, as measured by the heated and unheated sensors, respectively, was computed as an index of peripheral perfusion.15 To allow the comparison of data simultaneously monitored by the various devices, their analog outputs were digitally recorded at 50 Hz by a common recorder (RespitracePT16; NIMS; Miami Beach, FL).
Data Analysis
The data were summarized as the mean ± SD, and as medians and quartiles (for nonnormally distributed values). For critically ill patients, agreement between PtcCO2 and SpO2 by ear sensor with corresponding values of PaCO2 and SaO2 by arterial blood analysis, respectively, were evaluated by computing the mean difference and limits of agreement (ie, bias ± 2 SDs).16 In sleep apnea patients, 30 min of recordings with most pronounced repetitive oxygen desaturations were analyzed for each patient. The data were down-sampled to 1 Hz. The delay (phase lag) in SpO2 time series from different devices vs that from TOSCA was determined as the lag that provided the maximal coefficient of cross-correlation. A customized software identified desaturation events of > 3% from baseline, minimal SpO2 (nadir), and the slope (rate of fall) of SpO2 during events, as previously described.17 Briefly, the rolling baseline SpO2 was defined as the mean of the highest 15% of SpO2 values within the previous 5 min. A desaturation event was defined as a drop in SpO2 below baseline level by > 3% to a nadir, which was followed by a rapid increase in SpO2 of
3% within 10 s. Differences between values measured by different methods were compared by Student t test or Wilcoxon matched pairs tests, as appropriate. Significance was assumed at p < 0.05.
| Results |
|---|
|
|
|---|
|
Results in Sleep Apnea Patients
The sensor was well tolerated in all patients over the duration of the overnight sleep studies. The pulse oximeter response characteristics of the various devices and sensors applied in 12 patients with obstructive sleep apnea syndrome are summarized in Table 1
. The mean baseline SpO2 measured by the TOSCA sensor was 97 ± 2%. Pulse oximetry performed by the earlobe sensor of the TOSCA device, and the finger sensor of the Masimo Radical device revealed a significantly higher number of desaturation events (ie, SpO2 dips of > 3%), steeper desaturation slopes, and lower SpO2 nadirs during events compared to the other earlobe and finger sensors. SpO2 values measured with all different pulse oximeters and sensors were significantly correlated with the signal from the TOSCA sensor if the relative phase shift was compensated by cross-correlation analysis (Table 1). Visual and cross-correlation analysis revealed that the signal of the TOSCA sensor was leading, followed by the ear sensor of the Datex-Ohmeda 3740 with a modest delay of 5 s, and the finger sensor of the Datex-Ohmeda 3740, the Masimo Radical, and the Nellcor 395, with longer delays of 18 to 19 s (Fig 2
, Table 1).
|
|
| Discussion |
|---|
|
|
|---|
Several previous studies in adults with chronic respiratory failure or during acute critical illness or anesthesia,4518 and in mechanically ventilated children61920 revealed a clinically acceptable agreement between PtcCO2 values measured with a sensor at the trunk or extremity and PaCO2 values obtained from the analysis of arterial blood samples. The range of observed bias for PtcCO2 in the cited studies was 3 to + 3 mm Hg, and the range of limits of agreement was ± 4 to ± 7 mm Hg. The sensor used in the current investigation to monitor critically ill patients revealed a similar accuracy of PtcCO2 (ie, a bias of + 3 mm Hg that was statistically significant but clinically of minor relevance, and limits of agreement of ± 7 mm Hg). Of note, these results were achieved with a sensor heated to a temperature of only 42°C (rather than to 43 to 45°C as in the cited studies) in order to enhance patient comfort and to reduce the risk of skin trauma. In contrast to some earlier observations,21 the accuracy of PtcCO2 did not differ among patients with or without vasoactive drug therapy (Fig 1), suggesting that the accurate estimation of PtcCO2 is feasible in hemodynamically compromised patients. In terms of clinical decision making, the ability of a technique to detect changes in measured variables is important. Therefore, we evaluated the accuracy of the novel sensor in detecting changes in PaCO2 occurring spontaneously or in response to therapeutic interventions in the critically ill patients. The nonsignificant bias and the limits of agreement of changes in PtcCO2 vs changes in PaCO2 (ie, ±6 mm Hg) indicated that clinically relevant changes in PaCO2 of > 6 mm Hg can be identified by the transcutaneous sensor with a high level of confidence, and this performance of the noninvasive TOSCA sensor was similar to that of an intra-arterial probe for continuous blood gas analysis.22
Compared to SaO2 obtained by co-oximetry of arterial blood in critically ill patients, the SpO2 determined by the TOSCA sensor revealed a bias of 1% (p < 0.05), and limits of agreement of ± 4%. This is consistent with earlier results obtained with a prototype of the current sensor in healthy subjects and anesthetized patients,111223 and with studies evaluating other pulse oximeters.1
Since the SaO2 in critically ill patients was fairly stable and was maintained at > 88%, we further evaluated the pulse oximetry performance of the novel sensor in patients with obstructive sleep apnea. This condition is typically associated with rapid fluctuations in SpO2. We found that SpO2 measured by the novel ear sensor significantly correlated with corresponding values measured by the other tested pulse oximeters if the phase shift among devices and sensors was accounted for (Table 1). This analysis also demonstrated that the TOSCA sensor detected SpO2 desaturation events significantly earlier than the finger and earlobe sensors of the other devices. This was partly related to the greater circulation time for the finger compared to the earlobe, as evidenced by the greater delay of the Datex-Ohmeda 3740 signal when used with a finger sensor compared to an earlobe sensor (ie, a delay of 18 vs 5 s relative to the TOSCA signal) [Table 1]. Nevertheless, similar lag times of 18 to 19 s of various pulse oximeter signals with finger probes together with the 5-s lag of the Datex-Ohmeda 3740 signal with an earlobe probe indicate faster processing for the TOSCA signal. Rapid response characteristics of a pulse oximeter are desirable not only for the timely detection of deterioration in the status of a critically ill patient, but they also provide a greater sensitivity for the capture of transient drops in SpO2 such as desaturation events in sleep apnea patients. Thus, the TOSCA and Masimo Radical devices identified a significantly greater number of desaturation events that had steeper slopes and reached lower nadirs than corresponding events recorded by the Datex-Ohmeda 3740 and Nellcor N-395 devices (Table 1). These observations emphasize the impact of pulse oximetry signal processing on the outcome of diagnostic pulse oximetry for sleep apnea case finding and severity grading.24
We found an equivalent performance for heated vs unheated earlobe sensors in terms of SpO2 response characteristics, suggesting that a higher probe temperature was not essential for pulse oximetry in the studied setting. Yet, heating the probe is crucial for transcutaneous CO2 monitoring3 and might be advantageous for pulse oximetry in patients with impaired peripheral perfusion since it enhances pulsatile blood flow,11 as illustrated in three patients by an increase in the fraction of pulsatile infrared extinction by a factor of 1.8 to 2.6 in the heated vs the nonheated earlobe sensor.
In conclusion, we found that the novel combined SpO2 and PtcCO2 sensor accurately monitored SaO2, PaCO2, and their changes. Compared to other pulse oximeters, the novel sensor had favorable dynamic response characteristics. The current data corroborate previous results obtained with a prototype of the sensor in healthy subjects and patients during anesthesia,111223 and extends its validation to the application in critically ill patients, and for tracking rapid fluctuations in SpO2 in patients with sleep-disordered breathing. Due to its ability to noninvasively assess both ventilation and oxygenation in addition to pulse rate by a single transcutaneous sensor, the described noninvasive monitoring technique is a convenient and valuable tool for respiratory monitoring with potential applications in critical care, anesthesia, and sleep medicine.
| Acknowledgements |
|---|
| Footnotes |
|---|
The study was supported by an unconditional grant from Linde Medical Sensors, Switzerland.
Received for publication November 2, 2004. Accepted for publication December 23, 2004.