|
|
||||||||
Guest Access | Sign In via User Name/Password |
|||||||||
* From the Institute of Occupational and Environmental Medicine (Drs. Praml, Scharrer, de la Motte, and Nowak, and Radon), Ludwig-Maximilians-University, Munich; and Inamed GmbH (Drs. Scheuch and Sommerer), Institute for Aerosol Medicine, Gauting, Germany.
Correspondence to: Georg Praml, PhD, Institute of Occupational and Environmental Medicine, Ziemssenstr. 1, 80336 Munich, Germany; e-mail: Georg.Praml{at}med.uni-muenchen.de
| Abstract |
|---|
|
|
|---|
Method: The physical properties of both systems were characterized by the residual method indicating a similar particle size distribution and an average output of 6 µL/s for Mefar MB3 and 1.25 µL/s for APS Sidestream. In the comparison study, 34 subjects were included. Airway responsiveness was quantified by provocative dose of methacholine causing a 20% fall in FEV1.
Results: A significant difference was found between the two challenge systems (p =0.004, McNemar test). Nine subjects reached a 20% drop in FEV1 with the APS Sidestream only. The FEV1 dropped by > 20% using either system in eight subjects. In 17 subjects, none of the two systems caused a 20% decrease in FEV1.
Conclusion: Even if the physical dose is determined with elaborate methods, the biological dose may vary between two nebulizer systems, causing incomparable outcomes for subjects tested with different systems.
Key Words: bronchial hyperresponsiveness Lower Saxony Lung Study methacholine residual method
| Introduction |
|---|
|
|
|---|
Within the ECRHS, a standardized protocol for measurements of bronchial responsiveness to methacholine was developed.3 In the ECRHS, methacholine chloride was administered through an aerosol dosimeter (model MB3; Mefar; Bovezzo, Italy). However, during the study it found that the aerosol output of these nebulizers may vary largely.45
Since the start of the ECRHS in 1991, several new nebulizers have been developed, among them the APS Sidestream (Jaeger; Würzburg, Germany). Advantages of the APS Sidestream nebulizer include the use of a filter absorbing the methacholine in the exhalate, which reduces unknown extrainhalation of methacholine, and undesirable exposure to the technician handling the system. Furthermore, control of the dosimeter is an integral part of the lung function measurement software. To achieve comparable inhalation of methacholine for both protocols, we adapted the ECRHS protocol to the APS Sidestream device. Therefore, we compared output and particle size of methacholine droplets of the Mefar MB3 and the APS Sidestream dosimeters. Furthermore, we tested both systems in a sample of adult subjects with various degrees of bronchial hyperresponsiveness.
| Materials and Methods |
|---|
|
|
|---|
Particle Size: The particle size in terms of mass median aerodynamic diameter (MMAD) was determined with the residual method6: a 99mTc-tagged NaCl solution of known concentration was nebulized with the dosimeter and passed into a 300-L glass container that was continuously flushed with dry air (24 L/min, relative humidity < 10%), thus drying up the aerosol droplets. The particles of the solid residual were then size classified with an eight-stage cascade impactor (Andersen Mark II; Thermo Electron Corp; Minneapolis, MN). MMAD and geometric SD of the original droplets were obtained by measuring the count rates of the deposits on the respective impactor stages with a scintillation counter, taking into account the aerodynamic diameter and the density of the dry residuals as well as the NaCl concentration.
Three different Mefar MB3 nebulizers were each filled as prescribed by the manufacturer with 3 mL of 99mTc-tagged 5% NaCl solution and activated 30 times for 1 s (ECRHS setting; total nebulizing time, 30 s). Filling volume for the APS Sidestream was 1.03 mL. Since APS Sidestream software allows for precise timing of the nebulizing period, three cycles of 0.2 s, 0.4 s, and 1 s, respectively, were performed with a total nebulizing time of 30 s each. All of the cycles were followed by a 5-min impactor sampling period and were done twice.
Output: The same 99mTc-labeled 5% NaCl solution was nebulized and deposited directly on a filter (B50TE; Pall Corporation; East Hills, NY) without any further treatment; the output was then calculated from the results of a scintillation counter. Three Mefar MB3 nebulizers were filled with 3 mL of solution; the dosimeter was activated 10 times for 1 s. The APS Sidestream nebulizer was equally filled with 1.03 mL of solution; output was measured at 0.2-s, 0.4-s, and 1-s nebulizing periods (total nebulizing time, 10 s each). All of the tests were done twice.
Biological Measurements
Subjects:
A convenience sample of 34 subjects took part in the comparison study. Mean age was 35 ± 9 years (± SD); 12 of the subjects were male. The mean baseline FEV1 was 108.4 ± 13.6% of predicted7; eight subjects reported current wheeze. The study was approved by the Ethics Committee of the Faculty of Medicine at the Ludwig-Maximilians-University, Munich, Germany.
Baseline Lung Function Measurements: Prior to each methacholine challenge, lung function measurements including body plethysmography and spirometry were performed in all subjects according to American Thoracic Society criteria.8 All lung function measurements were done by two well-trained and experienced lung function technicians. For each subject, both challenges were done by the same technician.
Methacholine Testing: Methacholine challenge was done according to the protocol of the ECRHS.3 In short, baseline FEV1 was measured in the body plethysmograph according to American Thoracic Society recommendations.8 Participants were asked to avoid smoking for 1 h, refrain from using ß2-agonists or anticholinergic inhalers for 4 h, and not to use oral medications (ß2-agonists, theophylline, or antimuscarinics) for 8 h before each test.
Subjects who reported symptoms suggestive of asthma or chronic bronchitis or with a baseline FEV1 < 90% of predicted received doubling doses of methacholine (long protocol). For participants without such symptoms, fourfold increments of methacholine were administered until the FEV1 had dropped by > 10% from the control value. After that, doubling doses were used. The challenge was terminated if the FEV1 had dropped by > 20% from the control value or the maximum cumulative dose (1.2 mg) had been reached.
Prior to methacholine challenge, saline solution diluent (0.9%) was administered on both challenge days. The maximum post-saline solution FEV1 recorded 2 min after inhalation of the solution was used as the control value. In both methacholine challenge procedures, the subjects were asked to inhale deep and slowly from residual lung volume to near total lung capacity. FEV1 maneuvers were performed 2 min after inhalation and repeated until FEV1 maneuvers were technically acceptable.9 The participants were challenged twice on two different days within an 11-day interval using the Mefar MB3 and the APS Sidestream in random order. Both measurements were done at the same time of day.
Mefar MB3 Protocol
The challenge protocol was taken from the ECRHS manual. However, the first dose level had been omitted in order to shorten the duration of the procedure in the field study. Within the study, the Mefar MB3 dosimeter of the ECRHS center in Erfurt was used. The methacholine challenge doses for the short and long protocols are given in Table 1
. In addition to the dose assumed in the ECRHS protocol, the effective dose was calculated based on the mean output of the dosimeter calculated from the physical measurements. After expiration, subjects placed their lips around the mouthpiece, inspired to near total lung capacity, held their breath for at least 3 s, and exhaled into the challenge chamber. As in the ECRHS, the aerosol was delivered over a period of 1 s for each inhalation. If the FEV1 did not drop to 20% below the control FEV1, the next dose was administered immediately.
|
|
| Results |
|---|
|
|
|---|
|
Consequently, the APS Sidestream protocol was reliably adaptable to the ECRHS Mefar MB3 protocol assuming an output of 1.27 µL/s, provided the nebulization periods remain fairly close to 0.4 s. Additionally, the 6 µL/s Mefar MB3 output turned out to be substantially lower than the 10 µL/s output assumed during the ECRHS survey. Therefore, the 6 µL/s output found in the laboratory was used to establish the corresponding bronchial challenge steps for the biological measurements (Tables 1 , 2).
Biological Measurements
Technically satisfactory methacholine challenges with both methods were obtained in all 34 subjects. All subjects had baseline FEV1 > 80% predicted in both tests. The mean ± SD variation in baseline FEV1 between the two challenge days was 2.0 ± 1.7%. The maximum difference between the two baseline measurements was 8.5%. On average, the measurements were done within 5 days.
Using the McNemar test, the two methods resulted in significantly different results (p = 0.004): nine subjects reached a 20% drop in FEV1 with APS Sidestream only, and eight subjects did so with both systems (Table 4 ). Half of the subjects did not reach a 20% drop in FEV1 in either protocol. Thus, the overall agreement was 74%.
|
| Discussion |
|---|
|
|
|---|
In order to adapt the Mefar MB3 protocol to the APS Sidestream protocol, the physical properties of the two nebulizing systems had to be characterized in detail. This can only be done in a specialized laboratory for aerosol medicine, as the aerosol particles behave highly dynamically. Particle size and MMAD may change rapidly because of evaporation losses.6 Therefore, determining aerosol output by measuring weight lossas often recommended by manufacturersis inadequate.
Our study has shown that in spite of identical particle size and physical output, the dose producing a 20% fall in FEV1 differed significantly between the two nebulizer systems. Basically, a larger number of PD20-positive subjects was found with the APS Sidestream system.
In general, the intraindividual variation of methacholine challenge performed within a couple of days is low.1112 Likewise, a high level of agreement was shown for the Mefar MB3 dosimeter protocol using different methods and different challenge agents. Siersted and colleagues13 compared the ECRHS Mefar MB3 dosimeter protocol to the 2-min tidal breathing method, and found an overall agreement of 89% between the two methods and therefore much higher than in our study. Toelle et al14 compared the Mefar MB3 dosimeter to the protocol by Yan et al,15 and also showed an agreement of 89% between the two tests.
In these studies,1415 the variation between the two methods seemed to be at random, while in our study all subjects who were defined as bronchial hyperresponsive by the Mefar MB3 method were also bronchial hyperresponsive using the APS Sidestream method. Therefore, our result might indicate a systematic underestimation of the biological dose using the APS Sidestream protocol. One has to consider that not only particle size, fine particle mass, and nebulizer output determine the effective dose. In addition, the nebulization period and the triggering time might influence the amount of aerosols that is deposited in the lungs of the subjects as well as the site of aerosol deposition.
We did not control for caffeine consumption on the two challenge days.16 Likewise, we did not assess use of longer-acting bronchodilating agents prior to testing. However, due to the small intraindividual differences in baseline FEV1 between the two challenge days, we do not think that our results were largely affected. For the physical measurements, pure 5% NaCl solution was used, while the bronchial challenge was done with methacholine diluted in 0.9% NaCl solution. It would be desirable to use tagged methacholine solution for the physical measurements; however, this is unfortunately not feasible with the sophisticated method used to determine the output of the two nebulizers. Different filling volumes were used for the physical measurements in the APS Sidestream (1.03 mL) and Mefar MB3 (3 mL) nebulizers. This was done because these volumes are usually applied in the challenge protocols. The same filling volumes were used in our biological measurements because our physical measurements have shown that higher filling volumes, especially in the APS Sidestream system, might slightly increase the output (data not shown).
In conclusion, even if the physical dose of the nebulizer systems is determined with elaborate methods, the biological dose may vary largely. Therefore, in clinical practice results of bronchial challenge tests in a patient using an Mefar MB3 may not be comparable to results using the APS Sidestream.
| Acknowledgements |
|---|
| Footnotes |
|---|
The study was funded by the Lower Saxony Ministry for Social Affairs and the European Union.
Received for publication December 21, 2004. Accepted for publication June 1, 2005.
| References |
|---|
|
|
|---|
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |