* From the Departments of Medicine and Therapeutics (Drs. Hui and Sung), Anesthesia and Intensive Care (Drs. Chan and Joynt), and Architecture (Drs. Chow and Tsou), The Chinese University of Hong Kong, Hong Kong; the School of Mechanical Engineering (Dr. Hall), The University of New South Wales, Sydney, NSW, Australia; and the Department of Medicine (Dr. Sullivan), The University of Sydney, NSW, Australia.
Correspondence to: David S. Hui, MD, FCCP, Department of Medicine and Therapeutics, The Chinese University of Hong Kong, Prince of Wales Hospital, 3032 Ngan Shing St, Shatin, NT, Hong Kong; e-mail: dschui{at}cuhk.edu.hk
Abstract
Background: Health-care workers are concerned about the risk of acquiring contagious diseases such as severe acute respiratory syndrome and avian influenza after recent outbreaks. We studied exhaled air and particle dispersion through an oronasal mask attached to a human-patient simulator (HPS) during noninvasive positive-pressure ventilation (NPPV).
Methods: Airflow was marked with intrapulmonary smoke for visualization. Therapy with inspiratory positive airway pressure (IPAP) was started at 10 cm H2O and gradually increased to 18 cm H2O, whereas expiratory positive airway pressure was maintained at 4 cm H2O. A leakage jet plume was revealed by a laser light sheet and images captured by video. Smoke concentration in the plume was estimated from the light scattered by smoke particles.
Findings: A jet plume of air leaked through the mask exhaust holes to a radial distance of 0.25 m from the mask during the application of IPAP at 10 cm H2O with some leakage from the nasal bridge. The leakage plume exposure probability was highest about 60 to 80 mm lateral to the median sagittal plane of the HPS. Without nasal bridge leakage, the jet plume from the exhaust holes increased to a 0.40-m radius from the mask, whereas exposure probability was highest about 0.28 m above the patient. When IPAP was increased to 18 cm H2O, the vertical plume extended to 0.45 m above the patient with some horizontal spreading along the ward ceiling.
Conclusion: Substantial exposure to exhaled air occurs within a 0.5-m radius of patients receiving NPPV. Medical wards should be designed with an architectural aerodynamics approach and knowledge of air/particle dispersion from common mechanical ventilatory techniques.
Key Words: air and particle dispersion influenza noninvasive positive-pressure ventilation
Noninvasive positive-pressure ventilation (NPPV) has become an established method in the treatment of various causes of respiratory failure.1 A commonly used method of providing NPPV utilizes an orofacial mask and a single-arm delivery circuit driven by a high-gas flow-generating ventilator. This system requires a method to discharge excess gas flow and pressure to the environment through a bleeding port.2
NPPV is effective in the treatment of patients with respiratory failure due to COPD and acute cardiogenic pulmonary edema, and of immunocompromised patients with pneumonia.1 Anecdotal reports have indicated that NPPV applied via a single circuit was effective in treating patients with respiratory failure due to severe acute respiratory syndrome (SARS), and might reduce the need for intubation.345 An influenza pandemic will place extra demand on mechanical ventilatory support, and the safe use of NPPV would greatly enhance potential therapeutic options. However, deliberate gas leakage via exhalation ports may disperse infectious particles to the environment.
Viral pneumonia is potentially infectious to health-care workers (HCWs) and other patients. The outbreak several years ago of the SARS virus with high infectivity,67 and its potential to spread by airborne transmission891011 highlighted the risk of viral pneumonia. The reemergence of avian H5N1 infection with a high rate of respiratory failure in humans since 2004 in Asia has highlighted the potential need for providing mechanical ventilation in environments that are safe for HCWs.1213 The US pandemic influenza plan14 has recommended airborne precautions for HCWs when managing pandemic influenza with increased transmissibility and during procedures that may generate small aerosol particles of respiratory secretions.
It is not known how exhaled air and particles may disperse during NPPV in clinical settings because there is no reliable marker that can be safely introduced to the patients. We therefore assessed the risks of single-circuit NPPV in spreading infectious particles through the bleeding port and the orofacial mask interface using a high-fidelity human patient simulator (HPS). Knowledge about the extent and direction of mask leakage is essential in preparation for an influenza pandemic to reduce the risk of nosocomial transmission during the application of NPPV to high-risk patients.
Materials and Methods
The study was undertaken by a multidisciplinary team of investigators consisting of physicians, intensivists, anesthetists, architects, and an aeronautical engineer. The study was conducted in a quiet laboratory room, which measured 7.1 x 8.5 m with a height of 2.7 m. The room ventilation was temporarily suspended during the experiment to avoid potentially confounding environmental factors.
NPPV and the Lung Model
We studied the mask interface leakage and deliberate leakage from the exhaust holes of an oronasal mask (Ultra Mirage Medium; ResMed; North Ryde, NSW, Australia) firmly attached to a high-fidelity HPS (HPS 6.1; Medical Education Technologies Inc; Sarasota, FL) [Fig 1
]. The HPS is a realistic representation of human respiration. It contains a realistic airway and a lung model that undergoes gas exchange (ie, it removes oxygen and adds carbon dioxide to the system). The lung compliance and airway resistance also respond in a realistic manner to relevant challenges. In addition, the HPS produces an airflow pattern that is close to the in vivo human situation and has been applied in previous studies1516171819 to simulate human respiration.
|
Flow Visualization
The visualization of airflow around the oronasal mask was facilitated by marking the air with smoke particles produced by an M-6000 smoke generator (model N19; DS Electronics; Tempe, AZ). The oil-based smoke particles measured < 1 µm in diameter.20 The smoke was introduced continuously to the right main bronchus of the HPS, mixed with alveolar gas, and then exhaled through the airway. Sections through the leakage jet plume were then revealed by a thin laser light sheet created by an Nd:YVO4 Q-switched, frequency-doubled laser device (OEM T20-BL 10106Q; Spectra-Physic; Mountain View, CA), with custom cylindrical optics for two-dimensional laser light sheet generation (Fig 2
, top, A).
|
Image Analysis
We estimated the normalized smoke concentration of the mask leakage jet plume from the light scattered by the smoke particles. This extended the laser flow visualization and provided a useful quantitative method for understanding the range and shape of the leakage jet plumes. This analysis technique was previously developed for turbulent, two-phase air-particle flows and was shown to be reliable compared to isokinetic sampling21 and particle image velocimetry.22
The laser light sheet illuminated the smoke particles in the mask airflow leakage. We assumed the intensity of the laser light sheet was constant (Fig 2, middle, B) because the mask was positioned relatively far from the optics. Scattered laser light perpendicular to the light sheet was collected and integrated by the video camera charge-coupled device element (Fig 2, bottom, C). For a small region, the number of smoke particles within the region (or the particle concentration) is proportional to the total scattered light intensity in that small region if each particle scatters an equal intensity of light. This is true when particles are monodisperse, as scattering is dependent on particle size.23 For most types of smoke, the particles are essentially monodisperse with a tight size distribution.2024 This analysis was based on the scattered light intensity being proportional to the particle concentration under the special conditions of constant-intensity laser light sheet illumination and monodisperse small (submicrometer) particles.
Image Capture and Frame Extraction
The motion video of several breathing cycles for a given variable positive airway pressure (VPAP) setting was captured by a computer, and individual frames were extracted as gray-scale bitmap images for intensity analysis. Frames were extracted at times corresponding to the beginning of the HPS breath inhalation and the VPAP maximum pressure to generate an ensemble average for the corresponding instant of the respiratory cycle.
Intensity Averaging and Concentration Normalization
All gray-scale frames were read into a program that was specifically developed for this study (MathCad 8.0 2000; MathSoft Inc; Cambridge, MA)25 along with background intensity images that were obtained with the laser switched off. The background intensity was subtracted from each frame, pixel by pixel, to remove any stray background light, and the pixel intensity values were averaged over all frames to determine the ensemble-averaged intensity. The resulting image showed the total intensity of light scattered perpendicular to the light sheet by the smoke particles, which was directly proportional to the particle concentration. The image was normalized against the highest intensity found within the leakage jet plume to generate normalized particle concentration contours.
As smoke particles marked air that originated from the airways of the HPS before leaking from the mask, the concentration contours effectively represent the probability of encountering air around the patient that has come from within the mask and/or the patients respiratory system. The probability contours are made up of data collected from 10 breaths. We assigned a contour value of 1 to indicate a region that consisted of the entire amount of air leaked after exhalation by the patient, where there is a 100% chance of exposure to the exhaled air. A value near 0 indicates no measurable air leakage in the region and a negligible chance of exposure.
Magnification Factor and Image Distortion Correction
The final analysis step scaled the contour image to the dimensions of the measurement region (Fig 2, bottom, C) and corrected the optical distortions introduced by the camera and lens. Scaling and image distortion correction mapped the image in pixel coordinates back into the actual measurement region coordinates (in millimeters). Nonlinear mapping functions were determined by experimentally "testing" the camera and lens system. An accurate target grid placed in the measurement region was imaged, and image coordinates were compared to those on the grid to calculate third-order polynomial mapping functions using a least-squares method.2627 Previous work by Hall21 has shown this approach to be reliable, with a variance on probability contours not exceeding 5%, but often at < 3%. The maximum error on the x-y position from image distortion correction was ± 4 mm for this study.
Results
The results are presented predominantly with reference to the median sagittal plane, whereas the data taken from the paramedian planes are otherwise specified.
IPAP 10 cm H2O and EPAP 4 cm H2O With Leakage Around the Nasal Bridge
The first scenario was conducted with the VPAP machine set at an IPAP of 10 cm H2O and an EPAP of 4 cm H2O. The probability contours of the air leakage distribution around the patient and mask in the median sagittal plane are shown in Figure 3
, top, A. The vertical leakage plume jet from the mask exhaust holes was evident, along with a smaller leakage plume from the nasal bridge. The leakage plume from the six exhaust holes (each with a diameter of approximately 1.8 mm) was narrow and propagated well above the patient. Inclining the HPS to a 45° position oriented the exhaust holes such that the jet became almost vertical, potentially creating a maximum spread of particles.
|
We have identified several zones of safety with less than a given probability of the HCW being exposed to the patients exhaled air (Fig 3, top, A). Figure 3, bottom, B, shows the leakage jet plume from the right side of the mask, where the particles were illuminated along the median sagittal plane and five other paramedian planes. The median sagittal plane is located at Z = 0, where the patients head is shown (purple). The plume of the mask exhaust holes was directed vertically upward, and there was some leakage from around the nasal bridge. The leakage jet plume was actually most extensive about 60 to 80 mm lateral to the median sagittal plane. This is attributable to the fact that the six deliberate exhaust holes on the mask were positioned lateral to the median sagittal plane (Fig 1).
IPAP 10 cm H2O and EPAP 4 cm H2O With the Oronasal Mask Perfectly Sealed
After tightening the mask-strap on the HPS, air leakage from the nasal bridge was eliminated (Fig 4
). Again, the leakage jet plume from the mask exhaust holes was near vertical, but it had risen by 0.15 m because there was no longer a pressure drop from nasal bridge leakage. The plume was widened and pronounced at a distance between 0.1 and 0.4 m above the mask, but the jet directly from the exhaust holes was not strongly indicated. This is because the mask exhaust holes are not aligned on the sagittal plane. As the laser light sheet was focused on the sagittal plane, it did not detect the jet plume until it widened sufficiently to enter the light sheet, some distance above the mask. Based on both panels of Figure 4, we can identify the zones of safety for the probability of exposure to the patients exhaled air in this scenario.
|
|
|
This study shows that air marked with smoke particles can be emitted through deliberate mask leakage to a radial distance of approximately 0.25 m from an oronasal mask during the application of IPAP at 10 cm H2O and EPAP at 4 cm H2O. The leakage jet plume was most significant at approximately 60 to 80 mm lateral to the median sagittal plane. Despite what initially appeared to be a reasonably well-fitted mask, there was still leakage detected at the nasal bridge. With the elimination of this nasal bridge leakage, the jet plume radial distance from the mask increased to 0.4 m, whereas exposure probability was highest at approximately 0.28 m above the patient and the mask. When IPAP was increased to 14 cm H2O and then to 18 cm H2O, the vertical plume extended to 0.42 and 0.45 m, respectively, above the patient and the mask, with some horizontal spreading along the ceiling. These findings have important clinical implications for HCWs, who often nurse their patients at a close distance, especially during NPPV support for respiratory failure at a stage when the viral loads of the patients may reach peak levels. Our study emphasizes the importance of medical ward design in terms of a ventilated, aerodynamic space and the need for an "architectural aerodynamics" approach in order to minimize the risk of nosocomial infection. Air-conditioning or extraction systems will need to target the circular region that is above the mask rather than at the actual mask level.
It is controversial whether patients with respiratory failure due to severe community-acquired pneumonia (CAP) would benefit from NPPV. NPPV lowered the intubation rate, the length of stay in the ICU, and the 2-month mortality rate in patients with severe CAP, but these beneficial actions were confined to those patients with underlying COPD.28 In a prospective observational study29 that excluded patients with underlying COPD, NPPV improved oxygenation and respiratory rate in 22 of the 24 patients with CAP during the initial trial, but 16 patients subsequently needed intubation after an average time of 1.3 days, mainly because of worsening respiratory failure.
During the global outbreak of SARS, ARDS developed in 13 to 26% of patients, necessitating invasive ventilatory support at a time when a very high viral load was being reached; thus, HCWs were particularly prone to infection while caring for their patients.6303132 The relative risk of developing SARS was 13-fold for HCWs who were involved in intubating SARS patients compared to those who were not, whereas NPPV was not associated with a statistically significant risk for the HCWs.7 This was probably because tracheal suctioning was not generally performed for patients who were receiving ventilation with NPPV.7 However, a retrospective study by Xiao et al33 described NPPV exposure as being associated with clinical SARS infection in two HCWs.
Due to the fear of leakage from the mask interface with subsequent nosocomial transmission of SARS, the usage of NPPV was low. Nevertheless, individual reports345 have shown that single-circuit NPPV might provide life-saving treatment for patients who were in respiratory failure as a result of the infection. Treatment with NPPV was applied via oronasal masks to 20 SARS patients without COPD in whom severe hypoxemic respiratory failure developed in a hospital ward environment with efficient room air exchange, stringent infection control measures, full personal protective equipment, and the addition of a viral-bacterial filter to the exhalation port of the NPPV device.5 Endotracheal intubation was avoided in 14 patients (70%) who had a much shorter ICU LOS than those who were intubated. Clinical evidence of SARS did not develop in any of the 105 HCWs involved in the management of the 20 patients, whereas 102 HCWs (97%) had negative SARS serology findings.5 As there were still three HCWs who had refused to undergo blood testing, one cannot entirely eliminate the possibility of subclinical SARS infection related to use of NPPV, although it seemed highly unlikely.30 Subclinical SARS infection did occur in other HCWs who cared for SARS patients.3435
The reemergence of avian influenza H5N1 in humans in Asia and the recent spread to the Middle East is of grave concern.12133637 The infectivity potential for human-to-human transmission may increase with time after the mutation or genetic reassortment of viruses.38 Most of the human patients with H5N1 infection have presented with severe pneumonia complicated by ARDS and multiorgan dysfunction syndrome, and have a high mortality rate.12133637 The evidence for the use of NPPV as ventilatory support for patients with ARDS is weak, especially in the presence of multiorgan dysfunction syndrome and the lack of data from a controlled trial.3940 However, treatment with NPPV may play a role in reducing the need for endotracheal intubation among those patients with acute lung injury in an influenza pandemic, if it is applied with caution.
NPPV should be applied in patients with severe viral pneumonia only if there is adequate protection for HCWs because of the potential risk of viral transmission via deliberate or accidental mask interface leakage and flow compensation causing the dispersion of a contaminated aerosol.30 In patients with chronic respiratory failure who are receiving NPPV via nasal masks, air leakage through the mouth or other routes besides the exhalation valve can occur.2 In clinical practice, pressure necrosis often develops at the skin around the nasal bridge if the mask is applied tightly for a prolonged period of time. Many patients loosen the mask strap to relieve discomfort, and air leakage from the nasal bridge is definitely a potential source for the transmission of viral infections. Careful mask fitting is important for the successful and safe application of NPPV. The addition of a viral-bacterial filter to the NPPV breathing system between the mask and the exhalation port35 or the use of a dual-circuit NPPV device may reduce the risk of the nosocomial transmission of viral infection. However, the use of viral-bacterial filters was difficult due to blockage by moist secretions and could be applied only in some cases.35 In view of our observation that higher ventilator pressures result in a wider dispersion of exhaled air and a higher concentration of air leakage, we recommend starting NPPV with low IPAP level (ie, 8 to 10 cm H2O) and gradually increasing the level as necessary, instead of starting high and titrating downward if the patient is intolerant.41
There are several limitations to this study. First, smoke particles were used as markers for exhaled air, but they might not fully reflect the risk of droplet transmission. Our results indicate the spatial extent of the airflow leakage jet plumes but do not directly indicate the behavior of potentially infected droplets themselves. It is known that droplets generated by coughing, sneezing, talking, and breathing range from 0.5 to 12 µm in diameter4243 with a density similar to that of water. As these droplets are generally much larger than smoke particles (ie, they have a higher Stokes number), they will not follow the jet plume airflow perfectly but will slip relative to air streamlines under the influence of Stokes drag and gravitational forces. Stokes drag carries droplets with the airflow of the jet plume, but inertia causes droplets to lag behind the airflow and fall short of the full extent of the air jet plume. Conversely, gravitational force causes droplets to follow trajectories that are vertically below the air jet plume. We would therefore expect droplets to extend somewhat vertically below the contours given here but less than the contours given in directions vertically above and to the sides of the jet plume. The contours and safe distances suggested here are likely the "upper-bound" conservative estimates for droplet spread above and to the sides of the mask. Using the laser visualization method, we have shown that during the delivery of oxygen at a rate of 4 L/min via a Hudson mask to the HPS, the exhaled air at the peak of simulated exhalation could reach a distance of approximately 0.40 m.44 Last, we tested only one type of NPPV machine and one oronasal mask without supplemental oxygen, and did not test a nasal mask due to limited resources. Nevertheless, oronasal masks are generally preferred in the application of NPPV for acute respiratory failure as nasal masks are poorly tolerated in the acute setting due to mouth breathing.45 Further studies are needed to assess the safety of different respiratory therapies such as the addition of viral filters in the circuit, the use of supplemental oxygen, the use of a humidifier, and the use of different ward airflow settings, and to determine how the medical ward environment should be designed to protect HCWs from acquiring nosocomial pneumonia.
In conclusion, substantial exposure to exhaled air is likely to occur within 0.5 m from patients receiving NPPV. Higher ventilator pressures result in a wider distribution of exhaled air. HCWs should be aware of the potential risks of viral transmission during NPPV, and should take stringent contact and droplet precaution plus wear full personal protective equipment.
Footnotes
Abbreviations: CAP = community-acquired pneumonia; EPAP = expiratory positive airway pressure; HCW = health-care worker; HPS = human-patient simulator; IPAP = inspiratory positive airway pressure; NPPV = noninvasive positive-pressure ventilation; SARS = severe acute respiratory syndrome; VPAP = variable positive airway pressure
This research was supported by the Research Fund for the Control of Infectious Diseases (Health, Welfare and Food Bureau, Hong Kong Special Administrative Region).
The authors have reported to the ACCP that no significant conflicts of interest exist with any companies/organizations whose products or services may be discussed in this article.
Received for publication February 1, 2006. Accepted for publication April 6, 2006.
References
This article has been cited by other articles:
![]() |
L. Rubinson, J. L. Hick, J. R. Curtis, R. D. Branson, S. Burns, M. D. Christian, A. V. Devereaux, J. R. Dichter, D. Talmor, B. Erstad, et al. Definitive Care for the Critically Ill During a Disaster: Medical Resources for Surge Capacity: From a Task Force for Mass Critical Care Summit Meeting, January 26-27, 2007, Chicago, IL Chest, May 1, 2008; 133(5_suppl): 32S - 50S. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. S. Hui, S. D. Hall, M. T.V. Chan, B. K. Chow, S. S. Ng, T. Gin, and J. J.Y. Sung Exhaled Air Dispersion During Oxygen Delivery Via a Simple Oxygen Mask Chest, August 1, 2007; 132(2): 540 - 546. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |